Dose monitoring and output correction for the effects of scanning field changes with uniform scanning proton beam

被引:1
|
作者
Zhao, Qingya [1 ,2 ]
Wu, Huanmei [3 ]
Cheng, Chee-Wai [1 ,4 ]
Das, Indra J. [1 ,4 ]
机构
[1] IUHPTC, Bloomington, IN 47408 USA
[2] Purdue Univ, Sch Hlth Sci, W Lafayette, IN 47907 USA
[3] IUPUI, Purdue Sch Engn & Technol, Indianapolis, IN 46202 USA
[4] Indiana Univ, Dept Radiat Oncol, Sch Med, Indianapolis, IN 46202 USA
关键词
uniform scanning proton beam; dose monitor; stem effect; proton output calculation; BRAGG PEAK FIELDS; UNIT CALCULATIONS; THERAPY; DOSIMETRY; RANGE;
D O I
10.1118/1.3609417
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The output of a proton beam is affected by proton energy, Spread-Out Bragg Peak (SOBP) width, aperture size, dose rate, and the point of measurement. In a uniform scanning proton beam (USPB), the scanning field size is adjusted (including the vertical length and the horizontal width) according to the treatment field size with appropriate margins to reduce secondary neutron production. Different scanning field settings result in beam output variations that are investigated in this study. Methods: The measurements are performed with a parallel plate Markus chamber at the center of SOBP under the reference condition with 16 cm range, 10 cm SOBP, and 5 cm air gap. The effect of dose rate on field output is studied by varying proton beam current from 0.5 to 7 nA. The effects of scanning field settings are studied by varying independently the field width and length from 12 x 12 to 30 x 30 cm(2). Results: The results demonstrate that scanning field variations can produce output variation up to 3.80%. In addition, larger output variation is observed with scanning field changes along the stem direction of the patient dose monitor (PDM). By investigating the underlying physics of incomplete charge collection and the stem effects of the PDM, an analytical model is proposed to calculate USPB output with consideration of the scanning field area and the PDM stem length that is irradiated. The average absolute difference between the measured output and calculated output using our new correction model are within 0.13 and 0.08% for the 20 and 30 cm snouts, respectively. Conclusions: This study proposes a correction model for accurate USPB output calculation, which takes account of scanning field settings and the PDM stem effects. This model may be used to extend the existing output calculation model from one snout size to other snout sizes with customized scanning field settings. The study is especially useful for calculating field output for treatment without individualized patient specific measurements. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3609417]
引用
收藏
页码:4655 / 4661
页数:7
相关论文
共 50 条
  • [41] Pencil Beam Dose Algorithm for Intensity Modulated Proton Therapy with Scanning Proton Beams
    Wang, H.
    Zheng, H.
    Cao, R.
    Jia, J.
    Song, J.
    He, T.
    Wu, Y.
    [J]. MEDICAL PHYSICS, 2017, 44 (06)
  • [42] BEAM SCANNING FOR DOSE UNIFORMITY
    MCKEOWN, J
    CRAIG, ST
    DREWELL, NH
    FRKETICH, G
    SMYTH, DL
    [J]. RADIATION PHYSICS AND CHEMISTRY, 1995, 46 (4-6): : 1363 - 1372
  • [43] Ion recombination correction factors and detector comparison in a very-high dose rate proton scanning beam
    Leite, A. M. M.
    Cavallone, M.
    Ronga, M. G.
    Trompier, F.
    Ristic, Y.
    Patriarca, A.
    De Marzi, L.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 106
  • [44] A beam monitoring and validation system for continuous line scanning in proton therapy
    Klimpki, G.
    Psoroulas, S.
    Bula, C.
    Rechsteiner, U.
    Eichin, M.
    Weber, D. C.
    Lomax, A.
    Meer, D.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (15): : 6126 - 6143
  • [45] Effects of anatomical changes on pencil beam scanning proton plans in locally advanced NSCLC patients
    Szeto, Yenny Z.
    Witte, Marnix G.
    van Kranen, Simon R.
    Sonke, Jan-Jakob
    Belderbos, Jose
    van Herk, Marcel
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 120 (02) : 286 - 292
  • [46] Intrafraction motion and proton beam scanning
    Knopf, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S13 - S13
  • [47] Proton Beam Scanning for Mediastinal Lymphoma
    Zeng, C.
    Plastaras, J. P.
    Tochner, Z. A.
    James, P.
    White, B. M.
    Hill-Kayser, C. E.
    Hahn, S. M.
    Both, S.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : S163 - S164
  • [48] Uniform Scanning and Energy Stacking with Proton Beams
    Moyers, M.
    Anferov, V.
    Hsi, W.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [49] Method for exposure dose monitoring and control in scanning beam interference lithography
    Song, Ying
    Liu, Yujuan
    Jiang, Shan
    Zhu, Yang
    Zhang, Liu
    Liu, Zhaowu
    [J]. APPLIED OPTICS, 2021, 60 (10) : 2767 - 2774
  • [50] Forward / Backward Calibration and Clinical Application of a MLIC System for Uniform Scanning Proton Beam
    Lee, T.
    Mertens, T.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)