Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking

被引:73
|
作者
Farr, J. B. [1 ,2 ]
Mascia, A. E. [2 ]
Hsi, W. -C. [2 ]
Allgower, C. E. [2 ]
Jesseph, F. [2 ]
Schreuder, A. N. [2 ]
Wolanski, M. [2 ]
Nichiporov, D. F. [3 ]
Anferov, V. [3 ]
机构
[1] Indiana Univ, Dept Phys, Bloomington, IN 47405 USA
[2] Midwest Proton Radiotherapy Inst, Bloomington, IN 47408 USA
[3] Indiana Univ, Cyclotron Facil, Bloomington, IN 47408 USA
基金
美国国家卫生研究院;
关键词
proton; commissioning; uniform; scanning;
D O I
10.1118/1.2982248
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A proton beam delivery system on a gantry with continuous uniform scanning and dose layer stacking at the Midwest Proton Radiotherapy Institute has been commissioned and accepted for clinical use. This paper was motivated by a lack of guidance on the testing and characterization for clinical uniform scanning systems. As such, it describes how these tasks were performed with a uniform scanning beam delivery system. This paper reports the methods used and important dosimetric characteristics of radiation fields produced by the system. The commissioning data include the transverse and longitudinal dose distributions, penumbra, and absolute dose values. Using a 208 MeV cyclotron's proton beam, the system provides field sizes up to 20 and 30 cm in diameter for proton ranges in water up to 27 and 20 cm, respectively. The dose layer stacking method allows for the flexible construction of spread-out Bragg peaks with uniform modulation of up to 15 cm in water, at typical dose rates of 1-3 Gy/min. For measuring relative dose distributions, multielement ion chamber arrays, small-volume ion chambers, and radiographic films were employed. Measurements during the clinical commissioning of the system have shown that the lateral and longitudinal dose uniformity of 2.5% or better can be achieved for all clinically important field sizes and ranges. The measured transverse penumbra widths offer a slight improvement in comparison to those achieved with a double scattering beam spreading technique at the facility. Absolute dose measurements were done using calibrated ion chambers, thermoluminescent and alanine detectors. Dose intercomparisons conducted using various types of detectors traceable to a national standards laboratory indicate that the measured dosimetry data agree with each other within 5%. (C) 2008 American Association of Physicists in Medicine. [DOI: 10.1118/1.2982248]
引用
收藏
页码:4945 / 4954
页数:10
相关论文
共 50 条
  • [1] Uniform Scanning and Energy Stacking with Proton Beams
    Moyers, M.
    Anferov, V.
    Hsi, W.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [2] Forward / Backward Calibration and Clinical Application of a MLIC System for Uniform Scanning Proton Beam
    Lee, T.
    Mertens, T.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [3] The Effect of Scanning Pattern and Dose Rate on Output Factor for Uniform Scanning Proton Beam
    Zhao, Q.
    Wu, H.
    Pack, D.
    Cheng, C.
    Das, I.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [4] Reduction of the number of stacking layers in proton uniform scanning
    Fujitaka, Shinichiro
    Takayanagi, Taisuke
    Fujimoto, Rintaro
    Fujii, Yusuke
    Nishiuchi, Hideaki
    Ebina, Futaro
    Okazaki, Takashi
    Hiramoto, Kazuo
    Sakae, Takeji
    Terunuma, Toshiyuki
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (10): : 3101 - 3111
  • [5] Clinical commissioning of a therapeutic uniform proton beam scanning system with multi-element detectors
    Farr, J.
    Mascia, A.
    Hsi, W.
    Allgower, C.
    Jesseph, F.
    Nichiporov, D.
    Anferov, V.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2008, 184 : 127 - 127
  • [6] Therapeutic uniform scanning proton beam development and characterization: Longitudinal results
    Mascia, A.
    Farr, J.
    Hsi, W.
    Allgower, C.
    Nichiporov, D.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2632 - 2632
  • [7] Therapeutic uniform scanning proton beam development and characterization: Transverse results
    Farr, J.
    Mascia, A.
    Allgower, C.
    Hsi, W.
    Anferov, V.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2575 - 2575
  • [8] Dose monitoring and output correction for the effects of scanning field changes with uniform scanning proton beam
    Zhao, Qingya
    Wu, Huanmei
    Cheng, Chee-Wai
    Das, Indra J.
    [J]. MEDICAL PHYSICS, 2011, 38 (08) : 4655 - 4661
  • [9] Commissioning of Proton Beam Uniform Scanning for a Commercial Treatment Planning System
    Egan, A.
    Maes, D.
    Nevitt, T.
    Leuro, E.
    Fung, A.
    Regmi, R.
    Bloch, C.
    Saini, J.
    Taddei, P.
    Bowen, S.
    Janson, M.
    Wong, T.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E722 - E722
  • [10] MU Calculation for Uniform Scanning Proton Beam
    Zhao, Q.
    Wu, H.
    Wolanski, M.
    Hecksel, D.
    Pack, D.
    Das, I.
    [J]. MEDICAL PHYSICS, 2009, 36 (06) : 2795 - +