Lens of the eye dose calculation for neuro-interventional procedures and CBCT scans of the head

被引:7
|
作者
Xiong, Zhenyu [1 ]
Vijayan, Sarath [1 ]
Rana, Vijay [1 ]
Jain, Amit [1 ]
Rudin, Stephen [1 ]
Bednarek, Daniel R. [1 ]
机构
[1] Univ Buffalo, Toshiba Stroke & Vasc Res Ctr, Buffalo, NY USA
关键词
lens dose; Monte Carlo; EGSnrc; DTS; CBCT; neuro-interventional procedure; RADIATION;
D O I
10.1117/12.2216504
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The aim of this work is to develop a method to calculate lens dose for fluoroscopically-guided neuro-interventional procedures and for CBCT scans of the head. EGSnrc Monte Carlo software is used to determine the dose to the lens of the eye for the projection geometry and exposure parameters used in these procedures. This information is provided by a digital CAN bus on the Toshiba Infinix C-Arm system which is saved in a log file by the real-time skin-dose tracking system (DTS) we previously developed. The x-ray beam spectra on this machine were simulated using BEAMnrc. These spectra were compared to those determined by SpekCalc and validated through measured percent-depth-dose (PDD) curves and half-value-layer (HVL) measurements. We simulated CBCT procedures in DOSXYZnrc for a CTDI head phantom and compared the surface dose distribution with that measured with Gafchromic film, and also for an SK150 head phantom and compared the lens dose with that measured with an ionization chamber. Both methods demonstrated good agreement. Organ dose calculated for a simulated neuro-interventional-procedure using DOSXYZnrc with the Zubal CT voxel phantom agreed within 10% with that calculated by PCXMC code for most organs. To calculate the lens dose in a neuro-interventional procedure, we developed a library of normalized lens dose values for different projection angles and kVp's. The total lens dose is then calculated by summing the values over all beam projections and can be included on the DTS report at the end of the procedure.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Lateral Head Shielding for Eye Lens Dose Reduction for Neuro-CBCT and Interventional Fluoroscopic Procedures
    Xiong, Z.
    Vijayan, S.
    Guo, C.
    Choudhary, P.
    Rudin, S.
    Bednarek, D.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [2] A method to reduce patient's eye lens dose in neuro-interventional radiology procedures
    Safari, M. J.
    Wong, J. H. D.
    Kadir, K. A. A.
    Sani, F. M.
    Ng, K. H.
    RADIATION PHYSICS AND CHEMISTRY, 2016, 125 : 75 - 81
  • [3] Variation of eye-lens dose with variation of the location of the beam isocenter in the head during neuro-interventional fluoroscopic procedures
    Guo, Chao
    Troville, Jonathan
    Sun, Sheng-Hsuan
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [4] A Method for Eye Lens Dose Estimation During Fluoroscopically-Guided Neuro-Interventional Procedures
    Xiong, Z.
    Kilian-Meneghin, J.
    Vijayan, S.
    Guo, C.
    Rudin, S.
    Bednarek, D.
    MEDICAL PHYSICS, 2017, 44 (06) : 3198 - 3198
  • [5] Multiparameter Dense Neural Network (DNN) to Estimate Patient Eye-Lens Dose During Neuro-Interventional Procedures
    Collins, J.
    Guo, C.
    Podgorsak, A.
    Rudin, S.
    Bednarek, D.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [6] Estimation of Patient Eye-Lens Dose During Neuro-Interventional Procedures using a Dense Neural Network (DNN)
    Collins, J.
    Sun, S.
    Guo, C.
    Podgorsak, A.
    Rudin, S.
    Bednarek, D. R.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [7] A system to track skin dose for neuro-interventional cone-beam computed tomography (CBCT)
    Vijayan, Sarath
    Xiong, Zhenyu
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [8] A Real-Time Skin Dose Tracking System for Biplane Neuro-Interventional Procedures
    Rana, Vijay K.
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [9] Real Time Resolution Enhanced and Dose-Reduced Display for Neuro-Interventional Procedures
    Nagesh, S. Setlur
    Ionita, C.
    Bednarek, D.
    Rudin, S.
    MEDICAL PHYSICS, 2017, 44 (06) : 3197 - 3197
  • [10] Dose estimation of eye lens for interventional procedures in diagnosis
    Liu, Yu-Rong
    Huang, Chia-Yu
    Hsu, Ching-Han
    Hsu, Fang-Yuh