Lens of the eye dose calculation for neuro-interventional procedures and CBCT scans of the head

被引:7
|
作者
Xiong, Zhenyu [1 ]
Vijayan, Sarath [1 ]
Rana, Vijay [1 ]
Jain, Amit [1 ]
Rudin, Stephen [1 ]
Bednarek, Daniel R. [1 ]
机构
[1] Univ Buffalo, Toshiba Stroke & Vasc Res Ctr, Buffalo, NY USA
关键词
lens dose; Monte Carlo; EGSnrc; DTS; CBCT; neuro-interventional procedure; RADIATION;
D O I
10.1117/12.2216504
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The aim of this work is to develop a method to calculate lens dose for fluoroscopically-guided neuro-interventional procedures and for CBCT scans of the head. EGSnrc Monte Carlo software is used to determine the dose to the lens of the eye for the projection geometry and exposure parameters used in these procedures. This information is provided by a digital CAN bus on the Toshiba Infinix C-Arm system which is saved in a log file by the real-time skin-dose tracking system (DTS) we previously developed. The x-ray beam spectra on this machine were simulated using BEAMnrc. These spectra were compared to those determined by SpekCalc and validated through measured percent-depth-dose (PDD) curves and half-value-layer (HVL) measurements. We simulated CBCT procedures in DOSXYZnrc for a CTDI head phantom and compared the surface dose distribution with that measured with Gafchromic film, and also for an SK150 head phantom and compared the lens dose with that measured with an ionization chamber. Both methods demonstrated good agreement. Organ dose calculated for a simulated neuro-interventional-procedure using DOSXYZnrc with the Zubal CT voxel phantom agreed within 10% with that calculated by PCXMC code for most organs. To calculate the lens dose in a neuro-interventional procedure, we developed a library of normalized lens dose values for different projection angles and kVp's. The total lens dose is then calculated by summing the values over all beam projections and can be included on the DTS report at the end of the procedure.
引用
收藏
页数:15
相关论文
共 50 条
  • [11] Patient Brain-Dose Estimation for Fluoroscopically-Guided Neuro-Interventional Procedures
    Xiong, Z.
    Guo, C.
    Troville, J.
    Rudin, S.
    Bednarek, D. R.
    MEDICAL PHYSICS, 2018, 45 (06) : E694 - E694
  • [12] Automatic Calculation of Organ and Effective Dose for CBCT and Interventional Fluoroscopic Procedures
    Xiong, Z.
    Vijayan, S.
    Oines, A.
    Rudin, S.
    Bednarek, D.
    MEDICAL PHYSICS, 2016, 43 (06) : 3749 - 3749
  • [13] Predictors of Clopidogrel Hyper-Responsiveness in Neuro-Interventional Procedures
    Suzuki, Ryotaro
    Nagaishi, Masaya
    Takano, Issei
    Takigawa, Tomoji
    Hyodo, Akio
    Suzuki, Kensuke
    JOURNAL OF STROKE & CEREBROVASCULAR DISEASES, 2021, 30 (01):
  • [14] Investigation of organ dose variation with adult head size and pediatric age for neuro-interventional projections
    Xiong, Zhenyu
    Vijayan, Sarath
    Guo, Chao
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [15] The effect of underlying bone on the beam angular correction in calculating the skin dose of the head in neuro-interventional imaging
    Sun, Sheng-Hsuan
    Rudin, Stephen
    Bednarek, Daniel R.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [16] Eye Lens Dose Reduction of Radiologists in Interventional Procedures with Lead Glasses
    Hu, P.
    Zhuo, W.
    Liu, H.
    MEDICAL PHYSICS, 2015, 42 (06) : 3719 - 3719
  • [17] Real-time Detection of Patient Head Position and Cephalometric Landmarks from Neuro-Interventional Procedure Images Using Machine Learning for Patient Eye-Lens Dose Prediction
    Collins, J.
    Troville, J.
    Williams, K.
    Rudin, S.
    Bednarek, D. R.
    MEDICAL IMAGING 2022: PHYSICS OF MEDICAL IMAGING, 2022, 12031
  • [18] Eye lens dosimetry for interventional procedures - Relation between the absorbed dose to the lens and dose at measurement positions
    Geber, Therese
    Gunnarsson, Mikael
    Mattsson, Soren
    RADIATION MEASUREMENTS, 2011, 46 (11) : 1248 - 1251
  • [19] Risk Factors of Ischemic Lesions Related to Cerebral Angiography and Neuro-interventional Procedures
    Sato, Masayuki
    Nakai, Yasunobu
    Tsurushima, Hideo
    Shiigai, Masanari
    Masumoto, Tomohiko
    Matsumura, Akira
    NEUROLOGIA MEDICO-CHIRURGICA, 2013, 53 (06) : 381 - 387
  • [20] Surgeon eye lens dose monitoring in interventional neuroradiology, cardiovascular and radiology procedures
    Busoni, S.
    Bruzzi, M.
    Giomi, S.
    Poggiali, C.
    Quattrocchi, M.
    Betti, M.
    D'Urso, D.
    Fedeli, L.
    Mazzoni, L. N.
    Paolucci, M.
    Rossi, F.
    Taddeucci, A.
    Bettarini, S.
    Tortoli, P.
    Belli, G.
    Bernardi, L.
    Gasperi, C.
    Campanella, F.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 104 : 123 - 128