A system to track skin dose for neuro-interventional cone-beam computed tomography (CBCT)

被引:5
|
作者
Vijayan, Sarath [1 ]
Xiong, Zhenyu [1 ]
Rudin, Stephen [1 ]
Bednarek, Daniel R. [1 ]
机构
[1] Univ Buffalo, Toshiba Stroke & Vasc Res Ctr, Buffalo, NY USA
关键词
D O I
10.1117/12.2216931
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The skin-dose tracking system (DTS) provides a color-coded illustration of the cumulative skin-dose distribution on a closely-matching 3D graphic of the patient during fluoroscopic interventions in real-time for immediate feedback to the interventionist. The skin-dose tracking utility of DTS has been extended to include cone-beam computed tomography (CBCT) of neurointerventions. While the DTS was developed to track the entrance skin dose including backscatter, a significant part of the dose in CBCT is contributed by exit primary radiation and scatter due to the many overlapping projections during the rotational scan. The variation of backscatter inside and outside the collimated beam was measured with radiochromic film and a curve was fit to obtain a scatter spread function that could be applied in the DTS. Likewise, the exit dose distribution was measured with radiochromic film for a single projection and a correction factor was determined as a function of path length through the head. Both of these sources of skin dose are added for every projection in the CBCT scan to obtain a total dose mapping over the patient graphic. Results show the backscatter to follow a sigmoidal falloff near the edge of the beam, extending outside the beam as far as 8 cm. The exit dose measured for a cylindrical CTDI phantom was nearly 10 % of the entrance peak skin dose for the central ray. The dose mapping performed by the DTS for a CBCT scan was compared to that measured with radiochromic film and a CTDI-head phantom with good agreement.
引用
下载
收藏
页数:7
相关论文
共 50 条
  • [1] Measurement of skin dose from cone-beam computed tomography imaging
    Sercan Akyalcin
    Jeryl D English
    Kenneth M Abramovitch
    Xiujiang J Rong
    Head & Face Medicine, 9
  • [2] Measurement of skin dose from cone-beam computed tomography imaging
    Akyalcin, Sercan
    English, Jeryl D.
    Abramovitch, Kenneth M.
    Rong, Xiujiang J.
    HEAD & FACE MEDICINE, 2013, 9
  • [3] Cone-Beam Computed Tomography (CBCT) to Computed Tomography (CT) Image Translation with Perceptual Loss to Eliminate Cone-Beam Artifacts
    Yoo, S.
    Choi, B.
    Park, I.
    Kim, J.
    Kim, H.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [4] Cone-Beam Computed Tomography (CBCT) Dacryocystography for Imaging of the Nasolacrimal Duct System
    Wilhelm, Kai E.
    Rudrof, Heike
    Greschus, Susanne
    Garbe, Stefan
    Luessem, Marcel
    Lischka, Thomas
    Schild, Hans H.
    Gerstner, Andreas O. H.
    CLINICAL NEURORADIOLOGY-KLINISCHE NEURORADIOLOGIE, 2009, 19 (04): : 283 - 291
  • [5] Evaluation of enamel pearls by cone-beam computed tomography (CBCT)
    Akgul, Nilgun
    Caglayan, Fatma
    Durna, Nurhan
    Sumbullu, Muhammed-Akif
    Akgul, Hayati-Murat
    Durna, Dogan
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2012, 17 (02): : E218 - E222
  • [6] Cone Beam Computed Tomography (CBCT) in the Field of Interventional Oncology of the Liver
    Bapst, Blanche
    Lagadec, Matthieu
    Breguet, Romain
    Vilgrain, Valerie
    Ronot, Maxime
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2016, 39 (01) : 8 - 20
  • [7] Cone Beam Computed Tomography (CBCT) in the Field of Interventional Oncology of the Liver
    Blanche Bapst
    Matthieu Lagadec
    Romain Breguet
    Valérie Vilgrain
    Maxime Ronot
    CardioVascular and Interventional Radiology, 2016, 39 : 8 - 20
  • [8] Kilovoltage cone-beam computed tomography imaging dose estimation and optimization: Need of daily cone-beam computed tomography
    Trivedi, Gaurav
    Dixit, Chandra K.
    Oinam, Arun S.
    Kapoor, Rakesh
    Bahl, Amit
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2019, 15 (03) : 470 - 474
  • [9] A Method to Convert Cone-Beam Computed Tomography (CBCT) Image for Dose Calculation and the Phantom Evaluation
    Zhang, Guang-shun
    Huang, Shao-min
    Chen, Cui
    Zhang, Dan-dan
    Deng, Xiao-wu
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 372 - 372
  • [10] Cone-beam computed tomography (CBCT) dose optimization technique and image quality assessment scoring
    Trivedi, Gaurav
    Singh, Pushpendra P.
    Oinam, Arun S.
    Singh, Ranjit
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2024, 20 (01) : 71 - 78