Ultrasound-based computer navigation of the acetabular component: a feasibility study

被引:4
|
作者
Wassilew, Georgi I. [1 ]
Heller, Markus O. [2 ,3 ]
Hasart, Olaf [1 ]
Perka, Carsten [1 ]
Suedhoff, Ingrid [4 ]
Janz, Viktor [1 ]
Duda, Georg N. [2 ,3 ]
Koenig, Christian [2 ]
机构
[1] Charite, Ctr Musculoskeletal Surg, Dept Orthoped, D-10117 Berlin, Germany
[2] Charite, Ctr Sports Sci & Sports Med Berlin CSSB, Ctr Musculoskeletal Surg, D-10117 Berlin, Germany
[3] Charite, Ctr Sports Sci & Sports Med Berlin CSSB, Julius Wolff Inst, D-10117 Berlin, Germany
[4] Aesculap AG, Tuttlingen, Germany
关键词
Ultrasound; Computer navigation; Acetabular component; Cadaver; Accuracy; TOTAL HIP-ARTHROPLASTY; CUP PLACEMENT; PELVIC PLANE; REPLACEMENT; ORIENTATION; DISLOCATION; ACCURACY; POSITION; ANTEVERSION; MOTION;
D O I
10.1007/s00402-011-1412-4
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction This feasibility study investigated the accuracy of anterior pelvic reference plane (APP) registration and acetabular cup orientation in two cadavers with different BMIs. Method Five observers each registered the APP five times in the 2 cadavers (BMIs: 32 kg/m(2) and 25 kg/m(2)) using an ultrasound-based navigation system. By comparison against the CT-derived reference landmarks, the errors in determining the individual landmarks defining the APP, as well as the resulting errors in the orientation of the APP and the acetabular cup orientation were determined. Results Across all measurements obtained with the ultrasound navigation system, the errors in rotation and version in determining the APP were 0.5 degrees +/- 1.0 degrees and -0.4 degrees +/- 2.0 degrees, respectively. The cup abduction and anteversion errors determined from all measurements of the five investigators for both cadavers together were -0.1 degrees +/- 1.0 degrees and -0.4 degrees +/- 2.7 degrees, respectively. The data further demonstrated a high reproducibility of the measurements for the resulting cup adduction and anteversion angle. Conclusion Our preliminary results confirm that ultrasound navigation is a highly accurate tool that allows a reproducible registration of the APP and thereby enables accurate and precise intraoperative determination of the acetabular cup orientation also in patients with increased BMI.
引用
收藏
页码:517 / 525
页数:9
相关论文
共 50 条
  • [41] Ultrasound-Based Characterization of Prostate Cancer Using Joint Independent Component Analysis
    Imani, Farhad
    Ramezani, Mahdi
    Nouranian, Saman
    Gibson, Eli
    Khojaste, Amir
    Gaed, Mena
    Moussa, Madeleine
    Gomez, Jose A.
    Romagnoli, Cesare
    Leveridge, Michael
    Chang, Silvia
    Fenster, Aaron
    Siemens, D. Robert
    Ward, Aaron D.
    Mousavi, Parvin
    Abolmaesumi, Purang
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2015, 62 (07) : 1796 - 1804
  • [42] Design and implementation of an electromagnetic ultrasound-based navigation technique for laparoscopic ablation of liver tumors
    Paolucci, Iwan
    Schwalbe, Marius
    Prevost, Gian Andrea
    Lachenmayer, Anja
    Candinas, Daniel
    Weber, Stefan
    Tinguely, Pascale
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2018, 32 (07): : 3410 - 3419
  • [43] A robust and automatic method for evaluating accuracy in 3-D ultrasound-based navigation
    Lindseth, F
    Bang, J
    Lango, T
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2003, 29 (10): : 1439 - 1452
  • [44] ULTRASOUND-BASED SURGICAL NAVIGATION FOR PERCUTANEOUS RENAL INTERVENTION: IN VIVO MEASUREMENTS AND IN VITRO ASSESSMEN
    Li, Zhicheng
    Gu, Jia
    Chakareski, Jacob
    Wang, Lei
    2011 18TH IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP), 2011,
  • [45] An Integrated System Using Ultrasound-Based Registration for Surgical Navigation of Iliosacral Screw Insertions
    Pandey, Prashant U.
    Guy, Pierre
    Bliven, Emily K.
    Lefavire, Kelly A.
    Hodgson, Antony J.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2024, 71 (12) : 3482 - 3493
  • [46] Design and implementation of an electromagnetic ultrasound-based navigation technique for laparoscopic ablation of liver tumors
    Iwan Paolucci
    Marius Schwalbe
    Gian Andrea Prevost
    Anja Lachenmayer
    Daniel Candinas
    Stefan Weber
    Pascale Tinguely
    Surgical Endoscopy, 2018, 32 : 3410 - 3419
  • [47] Ultrasound-Based Registration for the Computer-Assisted Navigated Percutaneous Scaphoid Fixation
    Broessner, Peter
    Hohlmann, Benjamin
    Welle, Kristian
    Radermacher, Klaus
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2023, 70 (09) : 1064 - 1072
  • [48] Quantitative error analysis for computer assisted navigation: A feasibility study
    Gueler, Oe
    Perwoeg, M.
    Kral, F.
    Schwarm, F.
    Bardosi, Z. R.
    Goebel, G.
    Freysinger, W.
    MEDICAL PHYSICS, 2013, 40 (02)
  • [49] A-mode ultrasound-based registration in computer-aided surgery of the skull
    Amstutz, C
    Caversaccio, M
    Kowal, J
    Bächler, R
    Nolte, LP
    Häusler, R
    Styner, M
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2003, 129 (12) : 1310 - 1316
  • [50] A novel combined acetabular and femoral computer navigation system
    Sugano, N
    Nishii, T
    Sakai, T
    Nishihara, S
    Ohzono, K
    Yonenobu, K
    Sato, Y
    Sasama, T
    Nakahodo, K
    Tamura, S
    Ochi, T
    ARTHROPLASTY 2000: RECENT ADVANCES IN TOTAL JOINT REPLACEMENT, 2001, : 129 - 135