Monte Carlo dosimetric characterization of the Flexisource Co-60 high-dose-rate brachytherapy source using PENELOPE

被引:6
|
作者
Almansa, Julio F. [1 ]
Guerrero, Rafael [1 ]
Torres, Javier [2 ]
Lallena, Antonio M. [3 ]
机构
[1] Complejo Hosp Univ Granada, Serv Fis & Protecc Radiol, Granada, Spain
[2] Hosp Gen Univ Ciudad Real, Serv Radiofis & Protecc Radiol, Ciudad Real, Spain
[3] Univ Granada, Dept Fis Atom Mol & Nucl, E-18071 Granada, Spain
关键词
Monte Carlo dosimetry; Brachytherapy; (CO)-C-60; HDR; PENELOPE; TECHNICAL NOTE; AAPM; PARAMETERS;
D O I
10.1016/j.brachy.2017.04.245
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
PURPOSE: Co-60 sources have been commercialized as an alternative to Ir-192 sources for high dose-rate (HDR) brachytherapy. One of them is the Flexisource Co-60 HDR source manufactured by Elekta. The only available dosimetric characterization of this source is that of Vijande et al. [J Contemp Brachytherapy 2012; 4:34-44], whose results were not included in the AAPM/ESTRO consensus document. In that work, the dosimetric quantities were calculated as averages of the results obtained with the Geant4 and PENELOPE Monte Carlo (MC) codes, though for other sources, significant differences have been quoted between the values obtained with these two codes. The aim of this work is to perform the dosimetric characterization of the Flexisource Co-60 HDR source using PENELOPE. METHODS AND MATERIALS: The MC simulation code PENELOPE (v. 2014) has been used. Following the recommendations of the AAPM/ESTRO report, the radial dose function, the anisotropy function, the air-kerma strength, the dose rate constant, and the absorbed dose rate in water have been calculated. RESULTS: The results we have obtained exceed those of Vijande et al. In particular, the absorbed dose rate constant is similar to 0.85% larger. A similar difference is also found in the other dosimetric quantities. The effect of the electrons emitted in the decay of Co-60, usually neglected in this kind of simulations, is significant up to the distances of 0.25 cm from the source. CONCLUSIONS: The systematic and significant differences we have found between PENELOPE results and the average values found by Vijande et al. point out that the dosimetric characterizations carried out with the various MC codes should be provided independently. (C) 2017 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:1073 / 1080
页数:8
相关论文
共 50 条
  • [41] Characterization of Dose Distributions in High-Dose-Rate Surface Brachytherapy
    Buzurovic, I.
    Hansen, J.
    Bhagwat, M.
    O'Farrell, D.
    Damato, A.
    Friesen, S.
    Devlin, P.
    Cormack, R.
    MEDICAL PHYSICS, 2015, 42 (06) : 3369 - 3370
  • [42] HIGH DOSE-RATE BRACHYTHERAPY USING CO-60 SOURCES IN THE TREATMENT OF CANCER OF THE UTERINE CERVIX
    SUBANDONO, AJ
    KUIPERS, T
    THE, SK
    MEERWALDT, JH
    BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (704): : 829 - 829
  • [43] Dosimetric impact of source-positioning uncertainty in high-dose-rate balloon brachytherapy of breast cancer
    Kim, Yongbok
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2015, 7 (05) : 387 - 396
  • [44] Monte Carlo Characterization of a New Directional Pd-103 High Dose Rate Source for Brachytherapy Application
    Heredia, A.
    Robinson, A.
    Henderson, D.
    Thomadsen, B.
    MEDICAL PHYSICS, 2012, 39 (06) : 3777 - 3777
  • [45] Monte Carlo characterization of an ytterbium-169 high dose rate brachytherapy source with analysis of statistical uncertainty
    Medich, DC
    Tries, MA
    Munro, JJ
    MEDICAL PHYSICS, 2006, 33 (01) : 163 - 172
  • [46] Monte Carlo characterization of the M-19 high dose rate Iridium-192 brachytherapy source
    Medich, David C.
    Munro, John J., III
    MEDICAL PHYSICS, 2007, 34 (06) : 1999 - 2006
  • [47] Monte Carlo modeling of Co-60 HDR brachytherapy source in water and in different solid water phantom materials
    Sahoo, S.
    Selvam, T. Palani
    Vishwakarma, R. S.
    Chourasiya, G.
    JOURNAL OF MEDICAL PHYSICS, 2010, 35 (01) : 15 - 22
  • [48] Verification of SuperMC for Simulation of a High-Dose-Rate Brachytherapy Source
    Naeem, Hamza
    Wasaye, Muhammad Abdul
    Chen, Chaobin
    Zheng, Huaqing
    Hao, Lijuan
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2017, 70 (12) : 1077 - 1082
  • [49] A Simple Monte-Carlo Treatment Planning System for Episcleral Brachytherapy Using a Novel High-Dose-Rate Y-90 Disc Source
    Polanco, M.
    Chang, X.
    Chang, J.
    MEDICAL PHYSICS, 2024, 51 (10) : 7819 - 7819
  • [50] A Dosimetric Study by Using a Multichannel Cylinder in High-Dose-Rate Brachytherapy Treatment of the Vaginal Cuff
    Syh, J. H.
    Zhang, M.
    White, M.
    Patel, B.
    Syh, J.
    Song, X.
    Wu, H.
    Rosen, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S501 - S501