Monte Carlo characterization of the M-19 high dose rate Iridium-192 brachytherapy source

被引:25
|
作者
Medich, David C.
Munro, John J., III
机构
[1] Univ Massachusetts, Radiat Lab, Lowell, MA 01854 USA
[2] Source Prod & Equipment Co Inc, St Rose, LA 70087 USA
关键词
brachytherapy; iridium-192; radiation dosimetry; Monte Carlo; MCNP;
D O I
10.1118/1.2733809
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The MCNP5 Monte Carlo code was used to simulate the dosimetry of an M-19 iridium-192 high dose rate brachytherapy source in both air/vacuum and water environments with the in-air photon spectrum filtered to remove low-energy photons below delta= 10 keV. Dosimetric data was organized into an away-along table and was used to derive the updated AAPM Task Group Report No. 43 (TG-43U1) parameters including S-K, D(r, theta), Lambda, g(L)(r), F(r, theta), phi(an)(r), and (phi) over bar (an) and their respective statistical uncertainties. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1999 / 2006
页数:8
相关论文
共 50 条
  • [1] Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy
    Gadhi, M. A.
    Buzdar, S. A.
    Afzal, M.
    Fatmi, Sh.
    Akhtar, M. S.
    Nizamani, A. H.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2013, 11 (03): : 189 - 193
  • [2] Determination of the tissue inhomogeneity correction in high dose rate Brachytherapy for Iridium-192 source
    Ravikumar, Barlanka
    Lakshminarayana, S.
    JOURNAL OF MEDICAL PHYSICS, 2012, 37 (01) : 27 - 31
  • [3] Dosimetric characterization of the M-15 high-dose-rate Iridium-192 brachytherapy source using the AAPM and ESTRO formalism
    Minh-Tri Ho Thanh
    Munro, John J., III
    Medich, David C.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 305 - 317
  • [4] The Dose Distribution from Iridium-192 Source on Cervical Cancer Brachytherapy by Manchester System Using Monte Carlo Simulation
    Kurniati, F.
    Krisna, F. P.
    Junios, J.
    Haryanto, F.
    ATOM INDONESIA, 2021, 47 (03) : 205 - 211
  • [5] Monte Carlo and experimental relative dose determination for an Iridium-192 source in water phantom
    Mowlavi, A. A.
    Cupardo, F.
    Severgnini, M.
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2008, 6 (01): : 37 - 41
  • [6] Four year clinical statistics of iridium-192 high dose rate brachytherapy
    Shigehara, K
    Mizokami, A
    Komatsu, K
    Koshida, K
    Namiki, M
    INTERNATIONAL JOURNAL OF UROLOGY, 2006, 13 (02) : 116 - 121
  • [7] Clinical Results of Iridium-192 High Dose Rate Brachytherapy with External Beam Radiotherapy
    Nohara, Takahiro
    Mizokami, Atsushi
    Kumano, Tomoyasu
    Shigehara, Kazuyoshi
    Konaka, Hiroyuki
    Yoshifumi, Kadono
    Yasuhide, Kitagawa
    Izumi, Kouji
    Narimoto, Kazutaka
    Namiki, Mikio
    JAPANESE JOURNAL OF CLINICAL ONCOLOGY, 2010, 40 (07) : 677 - 683
  • [8] Monte Carlo dosimetry of the IRAsource high dose rate 192Ir brachytherapy source
    Sarabiasl, Akbar
    Ayoobian, Navid
    Jabbari, Iraj
    Poorbaygi, Hossein
    Javanshir, Mohammad Reza
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2016, 39 (02) : 413 - 422
  • [9] Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source
    Angelopoulos, A
    Baras, P
    Sakelliou, L
    Karaiskos, P
    Sandilos, P
    MEDICAL PHYSICS, 2000, 27 (11) : 2521 - 2527
  • [10] New fluoroscopy system for monitoring source position in brachytherapy by a high-dose-rate iridium-192 remote afterloader
    Ishii, M
    Fukuda, H
    Miyamoto, T
    Hirokawa, K
    Nagano, K
    Fouriner-Bidoz, N
    RADIOLOGY, 1998, 209P : 158 - 158