Monte Carlo characterization of the M-19 high dose rate Iridium-192 brachytherapy source

被引:25
|
作者
Medich, David C.
Munro, John J., III
机构
[1] Univ Massachusetts, Radiat Lab, Lowell, MA 01854 USA
[2] Source Prod & Equipment Co Inc, St Rose, LA 70087 USA
关键词
brachytherapy; iridium-192; radiation dosimetry; Monte Carlo; MCNP;
D O I
10.1118/1.2733809
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The MCNP5 Monte Carlo code was used to simulate the dosimetry of an M-19 iridium-192 high dose rate brachytherapy source in both air/vacuum and water environments with the in-air photon spectrum filtered to remove low-energy photons below delta= 10 keV. Dosimetric data was organized into an away-along table and was used to derive the updated AAPM Task Group Report No. 43 (TG-43U1) parameters including S-K, D(r, theta), Lambda, g(L)(r), F(r, theta), phi(an)(r), and (phi) over bar (an) and their respective statistical uncertainties. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1999 / 2006
页数:8
相关论文
共 50 条
  • [11] Comparison of calibration procedures of iridium-192 sources used in high dose rate brachytherapy system
    Kumar, A
    Singh, S
    Agrawal, S
    INDIAN JOURNAL OF PURE & APPLIED PHYSICS, 2003, 41 (04) : 253 - 257
  • [12] Dose rate effect: does it exist clinically in high-dose-rate intracavitary brachytherapy for cervical cancer using iridium-192 source?
    Wang, C
    Huang, E
    Sun, L
    Chen, H
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S237 - S237
  • [13] Erratum to: Monte Carlo dosimetry of the IRAsource high dose rate 192Ir brachytherapy source
    Akbar Sarabiasl
    Navid Ayoobian
    Hossein Poorbaygi
    Iraj Jabbari
    Mohammad Reza Javanshir
    Australasian Physical & Engineering Sciences in Medicine, 2016, 39 : 591 - 591
  • [14] Monte Carlo and TLD dosimetry of an 192Ir high dose-rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Sakelliou, L
    Sandilos, P
    Antypas, C
    Vlachos, L
    Koutsouveli, E
    MEDICAL PHYSICS, 1998, 25 (10) : 1975 - 1984
  • [15] High dose rate iridium-192 brachytherapy as a component of radical radiotherapy for the treatment of localised prostate cancer
    Chin, Y. S.
    Bullard, J.
    Bryant, L.
    Bownes, P.
    Ostler, P.
    Hoskin, P. J.
    CLINICAL ONCOLOGY, 2006, 18 (06) : 474 - 479
  • [16] Intraoperative and percutaneous iridium-192 high-dose-rate brachytherapy for previously irradiated lesions of the spine
    Folkert, Michael R.
    Bilsky, Mark H.
    Cohen, Gil'ad N.
    Zaider, Marco
    Lis, Eric
    Krol, George
    Laufer, Ilya
    Yamada, Yoshiya
    BRACHYTHERAPY, 2013, 12 (05) : 449 - 456
  • [17] TLD, diode and monte carlo dosimetry of an Ir-192 source for high dose-rate brachytherapy
    Kirov, AS
    Williamson, JF
    Meigooni, AS
    Zhu, Y
    PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12): : 2015 - 2036
  • [18] A Monte Carlo investigation of the dosimetric characteristics of the VariSource 192Ir high dose rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Baras, P
    Sakelliou, L
    Sandilos, P
    Dardoufas, K
    Vlachos, L
    MEDICAL PHYSICS, 1999, 26 (08) : 1498 - 1502
  • [19] A dosimetric comparison of Xoft Axxent Electronic Brachytherapy and iridium-192 high-dose-rate brachytherapy in the treatment of endometrial cancer
    Dickler, Adam
    Kirk, Michael C.
    Coon, Alan
    Bernard, Damian
    Zusag, Tom
    Rotmensch, Jacob
    Wazer, David E.
    BRACHYTHERAPY, 2008, 7 (04) : 351 - 354
  • [20] Monte Carlo dosimetry of a new 192Ir pulsed dose rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Pantelis, E
    Papagiannis, P
    Sakelliou, L
    Kouwenhoven, E
    Baltas, D
    MEDICAL PHYSICS, 2003, 30 (01) : 9 - 16