Monte Carlo and TLD dosimetry of an 192Ir high dose-rate brachytherapy source

被引:87
|
作者
Karaiskos, P [1 ]
Angelopoulos, A
Sakelliou, L
Sandilos, P
Antypas, C
Vlachos, L
Koutsouveli, E
机构
[1] Univ Athens, Dept Phys, Nucl & Particle Phys Sect, GR-15771 Athens, Greece
[2] Univ Athens, Areteion Hosp, Sch Med, Dept Radiol, Athens 11528, Greece
[3] Hygeia Hosp, Dept Phys Med, Athens 15123, Greece
关键词
Ir-192; high dose rate brachytherapy; anisotropy; Monte Carlo; TLD;
D O I
10.1118/1.598371
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
An analytical Monte Carlo simulation code has been used to perform dosimetry calculations around an Ir-192 high dose-rate brachytherapy source utilized in the widely used microSelectron afterloaded system. Radial dose functions, dose rate constant and anisotropy functions, utilized in the AAPM Task Group 43 dose estimation formalism, have been calculated. In addition, measurements of anisotropy functions using LiF TLD-100 rods have been performed in a polystyrene phantom to support our Monte Carlo calculations. The energy dependence of LiF TLD response was investigated over the whole range of measurement distances and angles. TLD measurements and Monte Carlo calculations are in agreement to each other and agree with published data. The influence of phantom dimensions on calculations was also investigated. Radial dose functions were found to depend significantly on phantom dimensions at radial distances near phantom edges. Deviations of up to 25% are observed at these distances due to the lack of full scattering conditions, indicating that body dimensions should be taken into account in treatment planning when the absorbed dose is calculated near body edges. On the other hand, anisotropy functions do not demonstrate a strong dependence on phantom dimensions. However, these functions depend on radial distance at angles close to the longitudinal axis of the source, where deviations of up to 20% are observed. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)00310-1].
引用
收藏
页码:1975 / 1984
页数:10
相关论文
共 50 条
  • [1] TLD, diode and monte carlo dosimetry of an Ir-192 source for high dose-rate brachytherapy
    Kirov, AS
    Williamson, JF
    Meigooni, AS
    Zhu, Y
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12): : 2015 - 2036
  • [2] Monte Carlo dosimetry of the IRAsource high dose rate 192Ir brachytherapy source
    Sarabiasl, Akbar
    Ayoobian, Navid
    Jabbari, Iraj
    Poorbaygi, Hossein
    Javanshir, Mohammad Reza
    [J]. Australasian Physical and Engineering Sciences in Medicine, 2016, 39 (02): : 413 - 422
  • [3] Monte Carlo dosimetry of a new 192Ir high dose rate brachytherapy source
    Angelopoulos, A
    Baras, P
    Sakelliou, L
    Karaiskos, P
    Sandilos, P
    [J]. MEDICAL PHYSICS, 2000, 27 (11) : 2521 - 2527
  • [4] Erratum to: Monte Carlo dosimetry of the IRAsource high dose rate 192Ir brachytherapy source
    Akbar Sarabiasl
    Navid Ayoobian
    Hossein Poorbaygi
    Iraj Jabbari
    Mohammad Reza Javanshir
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2016, 39 : 591 - 591
  • [5] Monte Carlo dosimetry of a new 192Ir pulsed dose rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Pantelis, E
    Papagiannis, P
    Sakelliou, L
    Kouwenhoven, E
    Baltas, D
    [J]. MEDICAL PHYSICS, 2003, 30 (01) : 9 - 16
  • [6] TLD DOSIMETRY OF A NEW 192Ir PULSED DOSE RATE BRACHYTHERAPY SOURCE
    Kabiri, F.
    Sheibani, S.
    Ghozzati, B.
    Beigi, H. Pour
    Javanshir, M.
    [J]. INTERNAL MEDICINE JOURNAL, 2011, 41 : 45 - 45
  • [7] Monte Carlo dosimetry of the Buchler high dose rate 192Ir source
    Ballester, F
    Pérez-Calatayud, J
    Puchades, V
    Lluch, JL
    Serrano-Andrés, MA
    Limami, Y
    Lliso, F
    Casal, E
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (03): : N79 - N90
  • [8] Monte Carlo dosimetry of the VariSource high dose rate 192Ir source
    Wang, RQ
    Sloboda, RS
    [J]. MEDICAL PHYSICS, 1998, 25 (04) : 415 - 423
  • [9] Monte Carlo and experimental high dose rate 192Ir brachytherapy dosimetry with microDiamond detectors
    Rossi, Giulio
    Gainey, Mark
    Thomann, Benedikt
    Kollefrath, Michael
    Wuerfel, Jan
    Allgaier, Bernd
    Baltas, Dimos
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2019, 29 (03): : 272 - 281
  • [10] Experimental and Monte Carlo dosimetry of the Henschke applicator for high dose-rate 192Ir remote afterloading
    Watanabe, Y
    Roy, J
    Harrington, PJ
    Anderson, LL
    [J]. MEDICAL PHYSICS, 1998, 25 (05) : 736 - 745