DOSIMETRIC IMPACT OF INTERFRACTION CATHETER MOVEMENT IN HIGH-DOSE RATE PROSTATE BRACHYTHERAPY

被引:33
|
作者
Foster, William [1 ]
Cunha, J. Adam M. [1 ]
Hsu, I-Chow [1 ]
Weinberg, Vivan [1 ]
Krishnamurthy, Devan [1 ]
Pouliot, Jean [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA 94143 USA
关键词
HDR brachytherapy; Prostate; Interfraction catheter displacement; Dosimetry; EXTERNAL-BEAM RADIOTHERAPY; HDR-BRACHYTHERAPY; CANCER; OPTIMIZATION; IMPLANT; BOOST;
D O I
10.1016/j.ijrobp.2010.01.016
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate the impact of interfraction catheter movement on dosimetry in prostate high-dose-rate (HDR) brachytherapy. Methods and Materials: Fifteen patients were treated with fractionated HDR brachytherapy. Implants were performed on day 1 under transrectal ultrasound guidance. A computed tomography (CT) scan was performed. Inverse planning simulated annealing was used for treatment planning. The first fraction was delivered on day 1. A cone beam CT (CBCT) was performed on day 2 before the second fraction was given. A fusion of the CBCT and CT was performed using intraprostatic gold markers as landmarks. Initial prostate and urethra contours were transferred to the CBCT images. Bladder and rectum contours were drawn, and catheters were digitized on the CBCT. The planned treatment was applied to the CBCT dataset, and dosimetry was analyzed and compared to the initial dose distribution. This process was repeated after a reoptimization was performed, using the same constraints used on day 1. Results: Mean interfraction catheter displacement was 5.1 mm. When we used the initial plan on day 2, the mean prostate V100 (volume receiving 100 Gy or more) decreased from 93.8% to 76.2% (p < 0.01). Rectal V75 went from 0.75 cm(3) to 1.49 cm(3) (p < 0.01). A reoptimization resulted in a mean prostate V100 of 88.1%, closer to the initial plan (p = 0.05). Mean rectal V75 was also improved with a value of 0.59 cm(3). There was no significant change in bladder and urethra dose on day 2. Conclusions: A mean interfraction catheter displacement of 5.1 mm results in a significant decrease in prostate V100 And an increase in rectum dose. A reoptimization before the second treatment improves dose distribution. (C) 2011 Elsevier Inc.
引用
收藏
页码:85 / 90
页数:6
相关论文
共 50 条
  • [41] A sector-based dosimetric analysis of dose heterogeneity in high-dose-rate prostate brachytherapy
    Mesko, Shane
    Park, Sang-June
    Kishan, Amar U.
    Demanes, D. Jeffrey
    Kamrava, Mitchell
    BRACHYTHERAPY, 2015, 14 (02) : 173 - 178
  • [42] A Dosimetric Study Investigating the Impact of Catheter Displacement in HDR Prostate Brachytherapy
    Hunt, D.
    Soliman, H.
    Sankreacha, R.
    Morton, G.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [43] The impact of body mass index on dosimetric quality in low-dose-rate prostate brachytherapy
    Echevarria, Michelle I.
    Naghavi, Arash O.
    Venkat, Puja S.
    Abuodeh, Yazan A.
    Chevere, Carlos
    Yamoah, Kosj
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2016, 8 (05) : 386 - 390
  • [44] Evaluation of the dosimetric impact of loss and displacement of seeds in prostate low-dose-rate brachytherapy
    Wang, Yinkun
    Nasser, Nicola J.
    Borg, Jette
    Saibishkumar, Elantholi P.
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2015, 7 (03) : 203 - 210
  • [45] An Inter-Fraction Adaptive Strategy for High-Dose Rate Prostate Brachytherapy
    Pouliot, J.
    Foster, W.
    Cunha, A. J.
    Hsu, I-C
    MEDICAL PHYSICS, 2009, 36 (06)
  • [46] Dosimetric impact of intrafraction changes in MR-guided high-dose-rate (HDR) brachytherapy for prostate cancer
    Rink, Alexandra
    Borg, Jette
    Simeonov, Anna
    O'Leary, Gerald
    Helou, Joelle
    Menard, Cynthia
    Chung, Peter
    Jaffray, David A.
    Berlin, Alejandro
    BRACHYTHERAPY, 2018, 17 (01) : 59 - 67
  • [47] Image-Guided Endoluminal High-Dose Rate Brachytherapy for Rectal Cancer: Dosimetric Analysis
    Trommel, M.
    Scherf, C.
    Fleischmann, M.
    Diefenhardt, M.
    Chatzikonstantinou, G.
    Tselis, N.
    Fokas, E.
    Roedel, C.
    Licher, J.
    Ramm, U.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2022, 198 (SUPPL 1) : S172 - S172
  • [48] Dosimetric impact of intrafraction motion and interfraction changes in dose-escalated prostate SBRT
    Panizza, D.
    Faccenda, V.
    Daniotti, M. C.
    Trivellato, S.
    Caricato, P.
    Lucchini, R.
    Arcangeli, S.
    De Ponti, E.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1636 - S1636
  • [49] High-dose rate brachytherapy for high-risk prostate cancer [High-dose-rate-brachytherapie beim risikoprostatakarzinom]
    Schenck M.
    Krause K.
    Schwandtner R.
    Haase I.
    Fluehs D.
    Friedrich J.
    Jaeger T.
    Boergermann C.
    Ruebben H.
    Stuschke M.
    Der Urologe, 2006, 45 (6): : 715 - 722
  • [50] The Dosimetric Impact of Rotational Error With Multichannel Vaginal Cylinder for High Dose Rate Brachytherapy
    Kim, H.
    Glaser, S. M.
    Beriwal, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E677 - E678