Evaluation of the dosimetric impact of loss and displacement of seeds in prostate low-dose-rate brachytherapy

被引:15
|
作者
Wang, Yinkun [1 ]
Nasser, Nicola J. [2 ]
Borg, Jette [1 ]
Saibishkumar, Elantholi P. [3 ,4 ]
机构
[1] Univ Hlth Network, Princess Margaret Canc Ctr, Dept Radiat Phys, Toronto, ON M5G 2M9, Canada
[2] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, New York, NY 10021 USA
[3] Univ Hlth Network, Princess Margaret Canc Ctr, Dept Radiat Oncol, Toronto, ON M5G 2M9, Canada
[4] Univ Toronto, Dept Radiat Oncol, Toronto, ON, Canada
关键词
brachytherapy; dosimetry; LDR; prostate cancer; seeds; AMERICAN-BRACHYTHERAPY; PERMANENT IMPLANT; MIGRATION; I-125; CANCER; DISTRIBUTIONS; PLACEMENT; CHEST;
D O I
10.5114/jcb.2015.52127
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To analyze the seed loss and displacement and their dosimetric impact in prostate low-dose-rate (LDR) brachytherapy while utilizing the combination of loose and stranded seeds. Material and methods: Two hundred and seventeen prostate cancer patients have been treated with LDR brachytherapy. Loose seeds were implanted in the prostate center and stranded seeds in the periphery of the gland. Patients were imaged with transrectal ultrasound before implant and with computerized tomography/magnetic resonance imaging (CT/MR) one month after implant. The seed loss and displacement had been analyzed. Their impact on prostate dosimetry had been examined. The seed distribution beyond the prostate inferior boundary had been studied. Results: The mean number of seeds per patient that were lost to lung, pelvis/abdomen, urine, or unknown destinations was 0.21, 0.13, 0.03, and 0.29, respectively. Overall, 40.1% of patients had seed loss. Seed migration to lung and pelvis/abdomen occurred in 15.5% and 10.5% of the patients, respectively. Documented seed loss to urine was found in 3% of the patients while 20% of patients had seed loss to unknown destinations. Prostate length difference between pre-plan and post-implant images was within 6 mm in more than 98% of cases. The difference in number of seeds inferior to prostate between pre-plan and post-implant dosimetry was within 7 seeds for 93% of patients. At time of implant, 98% of seeds, inferior to prostate, were within 5 mm and 100% within 15 mm, and in one month post-implant 83% within 9 mm and 96.3% within 15 mm. Prostate post-implant V-100, D-90, and rectal wall RV100 for patients without seed loss were 94.6%, 113.9%, and 0.98 cm(3), respectively, as compared to 95.0%, 114.8%, and 0.95 cm(3) for the group with seed loss. Conclusions: Seed loss and displacement have been observed to be frequent. No correlation between seed loss and displacement and post-plan dosimetry has been reported.
引用
收藏
页码:203 / 210
页数:8
相关论文
共 50 条
  • [1] The impact of body mass index on dosimetric quality in low-dose-rate prostate brachytherapy
    Echevarria, Michelle I.
    Naghavi, Arash O.
    Venkat, Puja S.
    Abuodeh, Yazan A.
    Chevere, Carlos
    Yamoah, Kosj
    [J]. JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2016, 8 (05) : 386 - 390
  • [2] Low-dose-rate Prostate Brachytherapy: Characterization of Seeds Dynamics within the Prostate
    Karius, A.
    Lotter, M.
    Kreppner, S.
    Lubgan, D.
    Grigo, J.
    Fietkau, R.
    Strnad, V.
    Bert, C.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2022, 198 (SUPPL 1) : S118 - S119
  • [3] Dosimetric study of a low-dose-rate brachytherapy source
    Rodríguez-Villafuerte, M
    Arzamendi, S
    Díaz-Perches, R
    [J]. ADVANCED MONTE CARLO FOR RADIATION PHYSICS, PARTICLE TRANSPORT SIMULATION AND APPLICATIONS, 2001, : 473 - 477
  • [4] Dosimetric evaluation of clinical target volume in the postimplant analysis of low-dose-rate brachytherapy for prostate cancer
    Nasser, Nicola J.
    Sappiatzer, Joshua
    Wang, Yinkun
    Borg, Jette
    Saibishkumar, Elantholi P.
    [J]. BRACHYTHERAPY, 2015, 14 (02) : 189 - 196
  • [5] The impact of trainee involvement on outcomes in low-dose-rate brachytherapy for prostate cancer
    Shaikh, Talha
    Wang, Lora
    Ruth, Karen
    Hallman, Mark
    Chen, David Y.
    Greenberg, Richard E.
    Li, Jinsheng
    Crawford, Kevin
    Horwitz, Eric M.
    [J]. BRACHYTHERAPY, 2016, 15 (02) : 156 - 162
  • [6] Comparison of Intraoperative and Postoperative Dosimetric Parameters for the Urethra and Prostate in Patients Undergoing Low-Dose-Rate Prostate Brachytherapy
    Algan, O.
    Nicholas, Z.
    Ali, I.
    Sindhwani, P.
    Ahmad, S.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S380 - S381
  • [7] Control charts for evaluation of quality of low-dose-rate brachytherapy for prostate cancer
    Nasser, Nicola J.
    Saibishkumar, Elantholi P.
    Wang, Yinkun
    Chung, Peter W. M.
    Breen, Stephen L.
    [J]. JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2022, 14 (04) : 354 - 363
  • [8] Dosimetric Characteristics for a Low-Dose-Rate Brachytherapy Stent for Esophageal Cancer
    Rivard, Mark J.
    Herskovic, Arnold M.
    Hingston, John H.
    Clerc, Claude
    Favreau, John T.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 3125 - 3125
  • [9] Dosimetric guidance on using brachytherapy (low-dose-rate or high-dose-rate) to offset a flawed permanent prostate implant
    Zaider, Marco
    [J]. BRACHYTHERAPY, 2009, 8 (01) : 40 - 44
  • [10] Dosimetric Predictors of Sexual Function Decline Following Low-Dose-Rate Brachytherapy for Prostate Cancer (PCa)
    Mian, O. Y.
    Robertson, S. P.
    Narang, A.
    Agarwal, S.
    Bae, H. J.
    McNutt, T. R.
    Tran, P. T.
    DeWeese, T. L.
    Song, D.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : S123 - S123