Agreement and Repeatability of Corneal Thickness and Radius among Three Different Corneal Measurement Devices

被引:13
|
作者
Biswas, Sayantan [1 ,2 ]
Biswas, Partha [3 ]
机构
[1] Singapore Eye Res Inst, Visual Neurosci Res Grp, Singapore, Singapore
[2] MS Ramaiah Univ Appl Sci, Fac Life & Allied Hlth Sci, Dept Allied Hlth Sci, Bengaluru, Karnataka, India
[3] BB Eye Fdn, Kolkata, W Bengal, India
关键词
OPTICAL COHERENCE TOMOGRAPHY; ORBSCAN-II; TOPOGRAPHY; KERATOCONUS; PENTACAM; EYES; REPRODUCIBILITY; PACHYMETRY; CURVATURE; OCT;
D O I
10.1097/OPX.0000000000001785
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
SIGNIFICANCE Corneal refractive surgery screening, orthokeratology, and contact lens fitting require accurate estimation of the corneal thickness and curvature. We found that any of the imaging devices used in this study can be used to image a healthy myopic cornea, which is essential to screen for corneal ectasia. PURPOSE This study aimed to compare agreement and repeatability of corneal thickness and radius measurements obtained using Scheimpflug imaging (Pentacam HR), swept-source optical coherence tomography (Casia SS-1000), and Placido-Scheimpflug tomography (Topographic Modeling System, version 5 [TMS-5]). METHODS Sixty eyes of 60 subjects with myopia were measured with Casia, TMS-5, and Pentacam for their central corneal thickness (CCT), thinnest corneal thickness, central corneal radius of the steep meridian (Ks), central corneal radius of the flat meridian (Kf), mean central corneal radius (Km), and anterior and posterior best-fit spheres. Two readings were obtained from each instrument. Central corneal thickness was also measured with ultrasound pachymetry. Intraclass correlation coefficient, repeatability coefficient, and coefficient of variation were calculated. RESULTS Repeatability coefficient, coefficient of variation, and intraclass correlation coefficient of CCT among the instruments ranged between 2.48 and 12.79, 0.17 and 0.86, and 0.98 and 1.00, respectively. Mean CCT measurements were similar between Casia, TMS-5, Pentacam, and ultrasound pachymetry (P = .13). Intraclass range of correlation for agreement was 0.95 to 0.99 for CCT among instruments. The range of 95% limits of agreement (LoAs) for ultrasound CCT was narrowest with Casia (31.94 mu m). Casia-TMS-5 displayed a narrow range of 95% LoA for CCT (12.14 to -12.44), thinnest corneal thickness (12.31 to -11.31), and Ks (0.56 to -0.53), whereas Casia-Pentacam showed a narrow range of 95% LoA for Kf (0.56 to -0.42), Km (0.41 to -0.27), and anterior best-fit sphere (0.03 to -0.07). A significant difference in posterior best-fit sphere measurements was noted among all instruments (P = .01). CONCLUSIONS Casia, TMS-5, and Pentacam can be used interchangeably to measure corneal thickness and radius measurement in healthy eyes. Casia had the best agreement with ultrasound pachymeter CCT and exhibited the highest repeatability.
引用
收藏
页码:1196 / 1202
页数:7
相关论文
共 50 条
  • [21] Agreement of Corneal Thickness Measurement Using Slitlamp and Ultrasound Pachymetry
    Agarwal, Tushar
    Bhartiya, Shibal
    Dada, Tanuj
    Panda, Anita
    Jhanji, Vishal
    Yu, Marco
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2012, 38 (04): : 231 - 233
  • [22] Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound
    Lackner, B
    Schmidinger, G
    Pieh, S
    Funovics, MA
    Skorpik, C
    OPTOMETRY AND VISION SCIENCE, 2005, 82 (10) : 892 - 899
  • [23] Repeatability, Reproducibility, and Agreement of Two Scheimpflug-Placido Anterior Corneal Analyzers for Posterior Corneal Surface Measurement
    Bao, Fangjun
    Savini, Giacomo
    Shu, Bao
    Zhu, Senmiao
    Gao, Rongrong
    Dang, Guanxin
    Yu, Ayong
    Wang, Qinmei
    Huang, Jinhai
    JOURNAL OF REFRACTIVE SURGERY, 2017, 33 (08) : 524 - U114
  • [24] Repeatability of a novel corneal indentation device for corneal biomechanical measurement
    Lam, Andrew K. C.
    Hon, Ying
    Leung, Leo K. K.
    Lam, David C. C.
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2015, 35 (04) : 455 - 461
  • [25] Repeatability, reproducibility, and agreement of corneal power measurements obtained with a new corneal topographer
    Mao, Xinjie
    Savini, Giacomo
    Zhuo, Zuopao
    Feng, Yifan
    Zhang, Jian
    Wang, Qinmei
    Chen, Hao
    Huang, Jinhai
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (10): : 1561 - 1569
  • [26] Ultrasound biomicroscopic corneal thickness measurement for corneal thickness mapping
    Wada, I
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 1997, 41 (01) : 12 - 18
  • [27] Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement
    Kawamorita, Takushi
    Uozato, Hiroshi
    Kamiya, Kazutaka
    Bax, Leon
    Tsutsui, Kenta
    Aizawa, Daisuke
    Shimizu, Kimiya
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (01): : 127 - 133
  • [28] Comparison of Central Corneal Thickness and Anterior Chamber Depth measured by Three Different Devices
    Shin, Minchul
    Chung, Se Yoon
    Lee, Hyo Kyung
    Hwang, Ho Sik
    Han, Kyung Eun
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (07)
  • [29] Repeatability of Pentacam peripheral corneal thickness measurements
    Martin, Raul
    Jonuscheit, Sven
    Rio-Cristobal, Ana
    Doughty, Michael J.
    CONTACT LENS & ANTERIOR EYE, 2015, 38 (06): : 424 - 429
  • [30] Reliability, repeatability, and accordance between three different corneal diagnostic imaging devices for evaluating the ocular surface
    Garcia-Terraza, Abril L.
    Jimenez-Collado, David
    Sanchez-Sanoja, Francisco
    Arteaga-Rivera, Jose Y.
    Flores, Norma Morales
    Perez-Solorzano, Sofia
    Garfias, Yonathan
    Graue-Hernandez, Enrique O.
    Navas, Alejandro
    FRONTIERS IN MEDICINE, 2022, 9