Agreement and Repeatability of Corneal Thickness and Radius among Three Different Corneal Measurement Devices

被引:13
|
作者
Biswas, Sayantan [1 ,2 ]
Biswas, Partha [3 ]
机构
[1] Singapore Eye Res Inst, Visual Neurosci Res Grp, Singapore, Singapore
[2] MS Ramaiah Univ Appl Sci, Fac Life & Allied Hlth Sci, Dept Allied Hlth Sci, Bengaluru, Karnataka, India
[3] BB Eye Fdn, Kolkata, W Bengal, India
关键词
OPTICAL COHERENCE TOMOGRAPHY; ORBSCAN-II; TOPOGRAPHY; KERATOCONUS; PENTACAM; EYES; REPRODUCIBILITY; PACHYMETRY; CURVATURE; OCT;
D O I
10.1097/OPX.0000000000001785
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
SIGNIFICANCE Corneal refractive surgery screening, orthokeratology, and contact lens fitting require accurate estimation of the corneal thickness and curvature. We found that any of the imaging devices used in this study can be used to image a healthy myopic cornea, which is essential to screen for corneal ectasia. PURPOSE This study aimed to compare agreement and repeatability of corneal thickness and radius measurements obtained using Scheimpflug imaging (Pentacam HR), swept-source optical coherence tomography (Casia SS-1000), and Placido-Scheimpflug tomography (Topographic Modeling System, version 5 [TMS-5]). METHODS Sixty eyes of 60 subjects with myopia were measured with Casia, TMS-5, and Pentacam for their central corneal thickness (CCT), thinnest corneal thickness, central corneal radius of the steep meridian (Ks), central corneal radius of the flat meridian (Kf), mean central corneal radius (Km), and anterior and posterior best-fit spheres. Two readings were obtained from each instrument. Central corneal thickness was also measured with ultrasound pachymetry. Intraclass correlation coefficient, repeatability coefficient, and coefficient of variation were calculated. RESULTS Repeatability coefficient, coefficient of variation, and intraclass correlation coefficient of CCT among the instruments ranged between 2.48 and 12.79, 0.17 and 0.86, and 0.98 and 1.00, respectively. Mean CCT measurements were similar between Casia, TMS-5, Pentacam, and ultrasound pachymetry (P = .13). Intraclass range of correlation for agreement was 0.95 to 0.99 for CCT among instruments. The range of 95% limits of agreement (LoAs) for ultrasound CCT was narrowest with Casia (31.94 mu m). Casia-TMS-5 displayed a narrow range of 95% LoA for CCT (12.14 to -12.44), thinnest corneal thickness (12.31 to -11.31), and Ks (0.56 to -0.53), whereas Casia-Pentacam showed a narrow range of 95% LoA for Kf (0.56 to -0.42), Km (0.41 to -0.27), and anterior best-fit sphere (0.03 to -0.07). A significant difference in posterior best-fit sphere measurements was noted among all instruments (P = .01). CONCLUSIONS Casia, TMS-5, and Pentacam can be used interchangeably to measure corneal thickness and radius measurement in healthy eyes. Casia had the best agreement with ultrasound pachymeter CCT and exhibited the highest repeatability.
引用
收藏
页码:1196 / 1202
页数:7
相关论文
共 50 条
  • [31] The repeatability of corneal and corneal epithelial thickness measurements using optical coherence tomography
    Sin, Sokpheaktra
    Simpson, Trefford L.
    OPTOMETRY AND VISION SCIENCE, 2006, 83 (06) : 360 - 365
  • [32] Comparability and repeatability of corneal astigmatism measurements using different measurement technologies
    Visser, Nienke
    Berendschot, Tos T. J. M.
    Verbakel, Frenne
    de Brabander, John
    Nuijts, Rudy M. M. A.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (10): : 1764 - 1770
  • [33] Repeatability and agreement of central corneal thickness measurements with a new handheld non-contact pachymeter
    Lawrenson, John G. G.
    Gill, Simranjit
    Masuid, Isra
    Rashid, Fardip
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2023, 43 (06) : 1524 - 1530
  • [34] Agreement and repeatability of the assessment of central corneal thickness using a novel optical coherence based biometer
    Leube, Alexander
    Agarwala, Rajat
    Wahl, Siegfried
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [35] Comparison of central corneal thickness with four different optical devices
    Kuddusi Teberik
    Mehmet Tahir Eski
    Murat Kaya
    Handan Ankaralı
    International Ophthalmology, 2018, 38 : 2363 - 2369
  • [36] Comparison of central corneal thickness with four different optical devices
    Teberik, Kuddusi
    Eski, Mehmet Tahir
    Kaya, Murat
    Ankarali, Handan
    INTERNATIONAL OPHTHALMOLOGY, 2018, 38 (06) : 2363 - 2369
  • [37] Repeatability and agreement of intraocular pressure measurement among three tonometers
    Sanchez Pavon, Irene
    Canadas, Pilar
    Martin, Raul
    CLINICAL AND EXPERIMENTAL OPTOMETRY, 2020, 103 (06) : 808 - 812
  • [38] Repeatability and agreement of total corneal astigmatism measured in keratoconic eyes using four current devices
    Feng, Yue
    Nitter, Tore
    Bertelsen, Geir
    Stojanovic, Aleksandar
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2024, 52 (08): : 800 - 810
  • [39] Repeatability and Inter-device Agreement for Three Different Methods of Keratometry: Placido, Scheimpflug, and Color LED Corneal Topography
    Hidalgo, Irene Ruiz
    Rozema, Jos J.
    Dhubhghaill, Sorcha Ni
    Zakaria, Nadia
    Koppen, Carina
    Tassignon, Marie-Jose
    JOURNAL OF REFRACTIVE SURGERY, 2015, 31 (03) : 176 - U104
  • [40] Central Corneal Thickness Measurement
    Rao, Harsha Laxmana
    Pahuja, Shivani
    Murthy, Somasheila I.
    Senthil, Sirisha
    OPHTHALMOLOGY, 2011, 118 (05) : 1010 - 1010