Repeatability of a novel corneal indentation device for corneal biomechanical measurement

被引:31
|
作者
Lam, Andrew K. C. [1 ]
Hon, Ying [1 ]
Leung, Leo K. K. [2 ]
Lam, David C. C. [2 ]
机构
[1] Hong Kong Polytech Univ, Sch Optometry, Hong Kong, Hong Kong, Peoples R China
[2] Hong Kong Univ Sci & Technol, Dept Mech & Aerosp Engn, Hong Kong, Peoples R China
关键词
corneal biomechanics; elastic modulus; intraocular pressure; stiffness; COLLAGEN CROSS-LINKING; INTRAOCULAR-PRESSURE; TANGENT MODULUS; DEFORMATION; HYSTERESIS; PARAMETERS; TOMOGRAPHY; THICKNESS; STIFFNESS;
D O I
10.1111/opo.12219
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PurposeThe aim of the study was to evaluate the repeatability of a new device for measuring corneal biomechanics. MethodsTwenty-nine normal subjects aged 20-28years (23.41.7years) underwent measurements of corneal stiffness and tangent elastic modulus using a novel corneal indentation device. Corneal topography, axial biometry and Goldmann applanation tonometry were also performed during the visit. Subjects returned after about 1week, at approximately the same time, and with the corneal biomechanics, corneal topography and Goldmann applanation tonometry measured again. Both the intrasession and intersession repeatability was assessed. ResultsBoth the corneal stiffness and tangent elastic modulus demonstrated good intrasession repeatability (corneal stiffness: coefficient of variation=7.32%, intraclass correlation coefficient=0.75; tangent elastic modulus: coefficient of variation=7.34%, intraclass correlation coefficient=0.84). The mean modulus after normalised to normal intraocular pressure of 15.5mmHg was 0.755 +/- 0.159MPa. There was no significant difference between the two visits (paired t-tests: p>0.05). The repeatability [1.96 times the standard deviation (S.D.) of the intersession difference] of the corneal stiffness and the tangent elastic modulus was 0.0022Nmm(-1) and 0.197MPa, respectively. ConclusionThe corneal indentation device has good intrasession and intersession repeatability. It has good potential to measure corneal biomechanics clinically, even at different corneal regions.
引用
收藏
页码:455 / 461
页数:7
相关论文
共 50 条
  • [1] Comparative study of corneal tangent elastic modulus measurement using corneal indentation device
    Ko, Match W. L.
    Leung, Leo K. K.
    Lam, David C. C.
    MEDICAL ENGINEERING & PHYSICS, 2014, 36 (09) : 1115 - 1121
  • [2] Measurement repeatability of corneal aberrations
    Gobbe, M
    Guillon, M
    Maissa, C
    JOURNAL OF REFRACTIVE SURGERY, 2002, 18 (05) : S567 - S571
  • [3] A Novel Device for Clinical Measurement of Corneal Elasticity
    Leung, Christopher
    Ye, Cong
    Ko, Match
    Leung, Leo
    Lam, David
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [4] A Novel Indentation Assessment to Measure Corneal Biomechanical Properties in Glaucoma and Ocular Hypertension
    Xu, Yunzhi
    Ye, Yiming
    Chong, Iok Tong
    Chen, Zidong
    Xu, Jiangang
    Yang, Yangfan
    Yu, Keming
    Lam, David C. C.
    Yu, Minbin
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2021, 10 (09):
  • [5] A novel corneal indentation device for comparison of corneal tangent modulus before and after FS-LASIK in vivo
    Zhang, Yan
    Lin, Junyu
    Lu, Shu-Hao
    Chong, Jones Iok-Tong
    Yang, Cheng
    Lan, Jianqing
    Xie, Wenjuan
    Li, Juan
    Lam, David Chuen-Chun
    Cao, Dan
    Zeng, Jin
    BIOMEDICAL ENGINEERING ONLINE, 2025, 24 (01)
  • [6] Repeatability of corneal and conjunctival sensitivity measurement.
    Stapleton, F
    Tan, ME
    Vega, J
    Papas, EB
    Belmonte, C
    Holden, BA
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S338 - S338
  • [7] REPEATABILITY OF CORNEAL TOPOGRAPHY - THE CORNEAL FIELD
    ZADNIK, K
    FRIEDMAN, NE
    MUTTI, DO
    JOURNAL OF REFRACTIVE SURGERY, 1995, 11 (02) : 119 - 125
  • [8] OCT based air jet indentation for corneal biomechanical assessment
    Interdisciplinary Division of Biomedical Engineering, Hong Kong Polytechnic University, Hong Kong
    999077, Hong Kong
    不详
    100853, China
    Guangxue Jingmi Gongcheng, 2 (325-333):
  • [9] Agreement and Repeatability of Corneal Thickness and Radius among Three Different Corneal Measurement Devices
    Biswas, Sayantan
    Biswas, Partha
    OPTOMETRY AND VISION SCIENCE, 2021, 98 (10) : 1196 - 1202
  • [10] Repeatability of corneal biomechanical measurements in children wearing spectacles and orthokeratology lenses
    Hon, Ying
    Cheung, Sin W.
    Cho, Pauline
    Lam, Andrew K. C.
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2012, 32 (04) : 349 - 354