Agreement and Repeatability of Corneal Thickness and Radius among Three Different Corneal Measurement Devices

被引:13
|
作者
Biswas, Sayantan [1 ,2 ]
Biswas, Partha [3 ]
机构
[1] Singapore Eye Res Inst, Visual Neurosci Res Grp, Singapore, Singapore
[2] MS Ramaiah Univ Appl Sci, Fac Life & Allied Hlth Sci, Dept Allied Hlth Sci, Bengaluru, Karnataka, India
[3] BB Eye Fdn, Kolkata, W Bengal, India
关键词
OPTICAL COHERENCE TOMOGRAPHY; ORBSCAN-II; TOPOGRAPHY; KERATOCONUS; PENTACAM; EYES; REPRODUCIBILITY; PACHYMETRY; CURVATURE; OCT;
D O I
10.1097/OPX.0000000000001785
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
SIGNIFICANCE Corneal refractive surgery screening, orthokeratology, and contact lens fitting require accurate estimation of the corneal thickness and curvature. We found that any of the imaging devices used in this study can be used to image a healthy myopic cornea, which is essential to screen for corneal ectasia. PURPOSE This study aimed to compare agreement and repeatability of corneal thickness and radius measurements obtained using Scheimpflug imaging (Pentacam HR), swept-source optical coherence tomography (Casia SS-1000), and Placido-Scheimpflug tomography (Topographic Modeling System, version 5 [TMS-5]). METHODS Sixty eyes of 60 subjects with myopia were measured with Casia, TMS-5, and Pentacam for their central corneal thickness (CCT), thinnest corneal thickness, central corneal radius of the steep meridian (Ks), central corneal radius of the flat meridian (Kf), mean central corneal radius (Km), and anterior and posterior best-fit spheres. Two readings were obtained from each instrument. Central corneal thickness was also measured with ultrasound pachymetry. Intraclass correlation coefficient, repeatability coefficient, and coefficient of variation were calculated. RESULTS Repeatability coefficient, coefficient of variation, and intraclass correlation coefficient of CCT among the instruments ranged between 2.48 and 12.79, 0.17 and 0.86, and 0.98 and 1.00, respectively. Mean CCT measurements were similar between Casia, TMS-5, Pentacam, and ultrasound pachymetry (P = .13). Intraclass range of correlation for agreement was 0.95 to 0.99 for CCT among instruments. The range of 95% limits of agreement (LoAs) for ultrasound CCT was narrowest with Casia (31.94 mu m). Casia-TMS-5 displayed a narrow range of 95% LoA for CCT (12.14 to -12.44), thinnest corneal thickness (12.31 to -11.31), and Ks (0.56 to -0.53), whereas Casia-Pentacam showed a narrow range of 95% LoA for Kf (0.56 to -0.42), Km (0.41 to -0.27), and anterior best-fit sphere (0.03 to -0.07). A significant difference in posterior best-fit sphere measurements was noted among all instruments (P = .01). CONCLUSIONS Casia, TMS-5, and Pentacam can be used interchangeably to measure corneal thickness and radius measurement in healthy eyes. Casia had the best agreement with ultrasound pachymeter CCT and exhibited the highest repeatability.
引用
收藏
页码:1196 / 1202
页数:7
相关论文
共 50 条
  • [1] Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices
    Kiraly, Laszlo
    Stange, Jana
    Kunert, Kathleen S.
    Sel, Saadettin
    JOURNAL OF OPHTHALMOLOGY, 2017, 2017
  • [2] Agreement and repeatability of central corneal thickness measurements by four different optical devices and an ultrasound pachymeter
    Nesrin Buyuktortop Gokcinar
    Erhan Yumusak
    Nurgul Ornek
    Serap Yorubulut
    Zafer Onaran
    International Ophthalmology, 2019, 39 : 1589 - 1598
  • [3] Agreement and repeatability of central corneal thickness measurements by four different optical devices and an ultrasound pachymeter
    Gokcinar, Nesrin Buyuktortop
    Yumusak, Erhan
    Ornek, Nurgul
    Yorubulut, Serap
    Onaran, Zafer
    INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (07) : 1589 - 1598
  • [4] Reproducibility and repeatability of central corneal thickness measurement in healthy eyes using four different optical devices
    Karadag, Remzi
    Unluzeybek, Murat
    Cakici, Ozgur
    Kanra, Ayse Yagmur
    Bayramlar, Huseyin
    INTERNATIONAL OPHTHALMOLOGY, 2017, 37 (04) : 1039 - 1045
  • [5] Reproducibility and repeatability of central corneal thickness measurement in healthy eyes using four different optical devices
    Remzi Karadag
    Murat Unluzeybek
    Ozgur Cakici
    Ayse Yagmur Kanra
    Huseyin Bayramlar
    International Ophthalmology, 2017, 37 : 1039 - 1045
  • [6] Central corneal thickness measurement with different devices in keratoconic patients
    De Bernardo, Maddalena
    Rosa, Nicola
    JOURNAL OF CURRENT OPHTHALMOLOGY, 2018, 30 (01): : 97 - 97
  • [7] Repeatability and agreement of central corneal thickness measurement with non-contact methods: a comparative study
    Tarannum Mansoori
    Nagalla Balakrishna
    International Ophthalmology, 2018, 38 : 959 - 966
  • [8] Repeatability and agreement of central corneal thickness measurement with non-contact methods: a comparative study
    Mansoori, Tarannum
    Balakrishna, Nagalla
    INTERNATIONAL OPHTHALMOLOGY, 2018, 38 (03) : 959 - 966
  • [9] The repeatability of corneal thickness measures
    Marsich, MM
    Bullimore, MA
    CORNEA, 2000, 19 (06) : 792 - 795
  • [10] A Comparison of Central Corneal Thickness Measurements and Measurement Repeatability Using Three Imaging Modalities
    Woo, Sang Earn
    Lee, Si Hyung
    JOURNAL OF THE KOREAN OPHTHALMOLOGICAL SOCIETY, 2021, 62 (02): : 184 - 192