Reliability, repeatability, and accordance between three different corneal diagnostic imaging devices for evaluating the ocular surface

被引:6
|
作者
Garcia-Terraza, Abril L. [1 ,2 ]
Jimenez-Collado, David [1 ,3 ]
Sanchez-Sanoja, Francisco [1 ,4 ]
Arteaga-Rivera, Jose Y. [1 ]
Flores, Norma Morales [1 ]
Perez-Solorzano, Sofia [1 ]
Garfias, Yonathan [1 ]
Graue-Hernandez, Enrique O. [1 ]
Navas, Alejandro [1 ]
机构
[1] Conde Valenciana Inst Ophthalmol, Dept Cornea & Refract Surg, Mexico City, Mexico
[2] Autonomous Univ Baja California, Fac Med, Mexicali, Baja California, Mexico
[3] Panamer Univ, Sch Med, Mexico City, Mexico
[4] Anahuac Univ, Fac Hlth Sci, North Campus, Mexico City, Mexico
关键词
dry eye disease; diagnostic imaging; ocular surface; topography; diagnosis; TEAR; KERATOGRAPH; MEIBOGRAPHY; AGREEMENT;
D O I
10.3389/fmed.2022.893688
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
PurposeTo evaluate repeatability, reproducibility, and accordance between ocular surface measurements within three different imaging devices. MethodsWe performed an observational study on 66 healthy eyes. Tear meniscus height, non-invasive tear break-up time (NITBUT) and meibography were measured using three corneal imaging devices: Keratograph 5M (Oculus, Wetzlar, Germany), Antares (Lumenis, Sidney, Australia), and LacryDiag (Quantel Medical, Cournon d'Auvergne, France). One-way ANOVAs with post hoc analyses were used to calculate accordance between the tear meniscus and NITBUT. Reproducibility was assessed through coefficients of variation and repeatability with intraclass correlation coefficients (ICC). Reliability of meibography classification was analyzed by calculating Fleiss' Kappa Index and presented in Venn diagrams. ResultsCoefficients of variation were high and differed greatly depending on the device and measurement. ICCs showed moderate reliability of NITBUT and tear meniscus height measurements. We observed discordance between measurements of tear meniscus height between the three devices, F2, 195 = 15.24, p < 0.01. Measurements performed with Antares were higher; 0.365 +/- 0.0851, than those with Keratograph 5M and LacryDiag; 0.293 +/- 0.0790 and 0.306 +/- 0.0731. NITBUT also showed discordance between devices, F2, 111 = 13.152, p < 0.01. Measurements performed with LacryDiag were lower (10.4 +/- 1.82) compared to those of Keratograph 5M (12.6 +/- 4.01) and Antares (12.6 +/- 4.21). Fleiss' Kappa showed a value of -0.00487 for upper lid and 0.128 for inferior lid Meibography classification, suggesting discrete to poor agreement between measurements. ConclusionDepending on the device used and parameter analyzed, measurements varied between each other, showing a difference in image processing.
引用
收藏
页数:7
相关论文
共 12 条
  • [1] Accordance between three corneal diagnostic imaging devices in measuring the ocular surface
    Navas, Alejandro
    Garcia-Terraza, Abril L.
    Jimenez-Collado, David
    Sanchez-Sanoja, Francisco
    Arteaga-Rivera, Jose Y.
    Morales, Norma
    Perez-Solorzano, Sofia
    Ramirez-Miranda, Arturo
    Graue-Hernandez, Enrique O.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [2] Agreement and Repeatability of Corneal Thickness and Radius among Three Different Corneal Measurement Devices
    Biswas, Sayantan
    Biswas, Partha
    OPTOMETRY AND VISION SCIENCE, 2021, 98 (10) : 1196 - 1202
  • [3] Comparing Repeatability and Agreement between Commonly Used Corneal Imaging Devices in Keratoconus
    Angelo, Lize
    Gokul, Akilesh
    Mcghee, Charles
    Ziaei, Mohammed
    OPTOMETRY AND VISION SCIENCE, 2023, 100 (11) : 761 - 769
  • [4] Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices
    Kiraly, Laszlo
    Stange, Jana
    Kunert, Kathleen S.
    Sel, Saadettin
    JOURNAL OF OPHTHALMOLOGY, 2017, 2017
  • [5] Comparison of central corneal thickness measurements using three different imaging devices
    Kan, E.
    Duran, M.
    Yakar, K.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2023, 46 (06): : 589 - 595
  • [6] Repeatability, reproducibility and agreement between three different diagnostic imaging platforms for tear film evaluation of normal and dry eye disease
    Singh, Swati
    Srivastav, Saumya
    Modiwala, Zarin
    Ali, Mohammed Hasnat
    Basu, Sayan
    EYE, 2023, 37 (10) : 2042 - 2047
  • [7] Repeatability, reproducibility and agreement between three different diagnostic imaging platforms for tear film evaluation of normal and dry eye disease
    Swati Singh
    Saumya Srivastav
    Zarin Modiwala
    Mohammed Hasnat Ali
    Sayan Basu
    Eye, 2023, 37 : 2042 - 2047
  • [8] Repeatability, Reproducibility, and Agreement Between Three Different Scheimpflug Systems in Measuring Corneal and Anterior Segment Biometry
    Hernandez-Camarena, Julio C.
    Chirinos-Saldana, Patricia
    Navas, Alejandro
    Ramirez-Miranda, Arturo
    de la Mota, Angie
    Jimenez-Corona, Aida
    Graue-Hernandez, Enrique O.
    JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (09) : 616 - U44
  • [9] Diffusion kurtosis imaging in evaluating gliomas: different region of interest selection methods on time efficiency, measurement repeatability, and diagnostic ability
    Jian-ping Chu
    Yu-kun Song
    Yi-su Tian
    Hai-shan Qiu
    Xia-hua Huang
    Yu-liang Wang
    Ying-qian Huang
    Jing Zhao
    European Radiology, 2021, 31 : 729 - 739
  • [10] Diffusion kurtosis imaging in evaluating gliomas: different region of interest selection methods on time efficiency, measurement repeatability, and diagnostic ability
    Chu, Jian-ping
    Song, Yu-kun
    Tian, Yi-su
    Qiu, Hai-shan
    Huang, Xia-hua
    Wang, Yu-liang
    Huang, Ying-qian
    Zhao, Jing
    EUROPEAN RADIOLOGY, 2021, 31 (02) : 729 - 739