Effect of range modulation on neutron dose equivalent exposures around a passive scattering proton therapy treatment nozzle

被引:0
|
作者
Polf, J [1 ]
Newhauser, W [1 ]
机构
[1] MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX USA
关键词
D O I
10.1118/1.1998668
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of our study was to examine the influence of range modulation on the neutron dose equivalent exposures outside the treatment volume around a large‐field passively scattered proton therapy treatment nozzle. Method and Materials: In this study, the neutron dose equivalent spectra per proton (H(E/p)) and total neutron dose equivalent per therapeutic absorbed dose (H/D) were calculated using Monte Carlo simulations of the neutron fluence and the energy dependent neutron fluence‐to‐dose equivalent conversion factor for NCRP 38. (H(E)/p) and H/D were calculated at 54 locations around a passively scattered proton therapy treatment nozzle for varying amounts of range modulation. Results: As the step thickness of the range modulator wheel increased from 1.0 to 12.0 cm, the peak values of H(E)/p increased from 1⋅10−17 to 2⋅10−17 mSv/Gy at 50 cm from isocenter along the beam's central axis. In general, H/D increased with increasing range modulation at all locations studied, and the maximum H/D exposures shifted upstream of isocenter and away from the end of the nozzle. Conclusion: Several important findings can be summarized from the presented work. First, with increasing thickness of the RMW step, the high‐energy peak in the H(E)/p spectra shifted to higher energies. Second, at 90° with respect to the proton beam axis, the high‐energy peak occurs at substantially lower neutron energies. Also, the H/D values around the treatment nozzle increased as the modulation of the beam increased. Finally, the H/D values change significantly with distance from isocenter and angle with respect to the incident beam axis, due in part to the effects of the nozzle components on the neutron fluence downstream of the end of the nozzle. © 2005, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2164 / 2164
页数:1
相关论文
共 50 条
  • [1] Patient neutron dose equivalent exposures outside of the proton therapy treatment field
    Polf, JC
    Newhauser, WD
    Titt, U
    RADIATION PROTECTION DOSIMETRY, 2005, 115 (1-4) : 154 - 158
  • [2] Monte Carlo study of neutron dose equivalent during passive scattering proton therapy
    Zheng, Yuanshui
    Newhauser, Wayne
    Fontenot, Jonas
    Taddei, Phil
    Mohan, Radhe
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (15): : 4481 - 4496
  • [3] Calculations of neutron dose equivalent exposures from range-modulated proton therapy beams
    Polf, JC
    Newhauser, WD
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (16): : 3859 - 3873
  • [4] Independent application of an analytical model for secondary neutron equivalent dose produced in a passive-scattering proton therapy treatment unit
    Gallagher, Kyle J.
    Taddei, Phillip J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (15):
  • [5] Measurement of neutron dose distribution for a passive scattering nozzle at the Proton Medical Research Center (PMRC)
    Tayama, Ryuichi
    Fujita, Yasuo
    Tadokoro, Masahiro
    Fujimaki, Hisataka
    Sakae, Takeji
    Terunuma, Toshiyuki
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 564 (01): : 532 - 536
  • [6] The effect of field size on neutron dose equivalent around the target in proton beams
    Romero-Exposito, M.
    Romero-Exposito, M.
    Dasu, A.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1553 - S1553
  • [7] MEASUREMENT OF NEUTRON DOSE EQUIVALENT AND ITS DEPENDENCE ON BEAM CONFIGURATION FOR A PASSIVE SCATTERING PROTON DELIVERY SYSTEM
    Wang, Xin
    Sahoo, Narayan
    Zhu, Ronald X.
    Zullo, John R.
    Gillin, Michael T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (05): : 1563 - 1570
  • [8] Measurement of Neutron Spectrum and Ambient Dose Equivalent Around a Mini-Phantom at a Proton Therapy Facility
    Wang, Z.
    Kry, S.
    Burgett, E.
    Howell, R.
    Tailor, R.
    Oliver, J.
    Followill, D.
    Smith, A.
    Salehpour, M.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [9] The Effect of Proton Therapy Beam Scanning Pattern Size On Secondary Neutron Dose Equivalent
    Hecksel, D.
    Anferov, V.
    Shahnazi, K.
    Fitzek, M.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [10] Experimental Study of Neutron Dose Equivalent in a Proton Treatment Facility
    Wang, X.
    Gillin, M.
    MEDICAL PHYSICS, 2008, 35 (06) : 2982 - +