Comparison of scattering experiments using synchrotron radiation with Monte Carlo simulations using Geant4

被引:6
|
作者
Gerlach, M. [1 ]
Krumrey, M. [1 ]
Cibik, L. [1 ]
Mueller, P. [1 ]
Ulm, G. [1 ]
机构
[1] Phys Tech Bundesanstalt, D-10587 Berlin, Germany
关键词
Geant4; Synchrotron radiation; CdTe detector; Calibration; Polarization; X-RAY RANGE; DETECTORS; PERFORMANCE; CALIBRATION; ACCURACY; BEAMLINE; COMPTON; TOOLKIT; PHYSICS;
D O I
10.1016/j.nima.2009.06.062
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Monte Carlo techniques are powerful tools to simulate the interaction of electromagnetic radiation with matter. One of the most widespread simulation program packages is Geant4. Almost all physical interaction processes can be included. However, it is not evident what accuracy can be obtained by a simulation. In this work, results of scattering experiments using monochromatized synchrotron radiation in the X-ray regime are quantitatively compared to the results of simulations using Geant4. Experiments were performed for various scattering foils made of different materials such as copper and gold. For energy-dispersive measurements of the scattered radiation, a cadmium telluride detector was used. The detector was fully characterized and calibrated with calculable undispersed as well as monochromatized synchrotron radiation. The obtained quantum efficiency and the response functions are in very good agreement with the corresponding Geant4 simulations. At the electron storage ring BESSY II the number of incident photons in the scattering experiments was measured with a photodiode that had been calibrated against a cryogenic radiometer, so that a direct comparison of scattering experiments with Monte Carlo simulations using Geant4 was possible. It was shown that Geant4 describes the photoeffect, including fluorescence as well as the Compton and Rayleigh scattering, with high accuracy, resulting in a deviation of typically less than 20%. Even polarization effects are widely covered by Geant4, and for Doppler broadening of Compton-scattered radiation the extension G4LECS can be included, but the fact that both features cannot be combined is a limitation. For most polarization-dependent simulations, good agreement with the experimental results was found, except for some orientations where Rayleigh scattering was overestimated in the simulation. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:339 / 343
页数:5
相关论文
共 50 条
  • [1] The Effect of Beam Polarization in Microbeam Radiation Therapy (MRT): Monte Carlo Simulations Using Geant4
    Spiga, J.
    Prezado, Y.
    Braeuer-Krisch, E.
    Fanti, V.
    Randaccio, P.
    Bravin, A.
    [J]. 2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 2170 - +
  • [2] Geant4 Monte Carlo simulations for the LPCTrap setup
    Rodriguez, D.
    Ban, G.
    Durand, D.
    Duval, F.
    Flechard, X.
    Lienard, E.
    Mauger, F.
    Mery, A.
    Naviliat-Cuncic, O.
    Thomas, J. -C.
    Velten, Ph.
    [J]. EUROPEAN PHYSICAL JOURNAL A, 2009, 42 (03): : 397 - 400
  • [3] Geant4 Monte Carlo simulations for the LPCTrap setup
    D. Rodríguez
    G. Ban
    D. Durand
    F. Duval
    X. Fléchard
    E. Liénard
    F. Mauger
    A. Méry
    O. Naviliat-Cuncic
    J. -C. Thomas
    Ph. Velten
    [J]. The European Physical Journal A, 2009, 42 : 397 - 400
  • [4] Monte Carlo simulations in proton dosimetry with Geant4
    Moravek, Z
    Bogner, L
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S62 - S62
  • [5] Assessment of a new multileaf collimator concept using GEANT4 Monte Carlo simulations
    Tacke, MB
    Szymanowski, H
    Oelfke, U
    Schulze, C
    Nuss, S
    Wehrwein, E
    Leidenberger, S
    [J]. MEDICAL PHYSICS, 2006, 33 (04) : 1125 - 1132
  • [6] Monte Carlo simulation of the ELIMED beamline using Geant4
    Pipek, J.
    Romano, F.
    Milluzzo, G.
    Cirrone, G. A. P.
    Cuttone, G.
    Amico, A. G.
    Margarone, D.
    Larosa, G.
    Leanza, R.
    Petringa, G.
    Schillaci, F.
    Scuderi, V.
    [J]. JOURNAL OF INSTRUMENTATION, 2017, 12
  • [7] Cellular Dosimetry Using the Geant4 Monte Carlo Toolkit
    Freudenberg, Robert
    Kotzerke, Jorg
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51 (09) : 1488 - 1489
  • [8] Proton Source Modeling for Geant4 Monte Carlo Simulations
    Barnes, S.
    McAuley, G.
    Wroe, A.
    Slater, J.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3756 - 3757
  • [9] PRELIMINARY STUDY OF PROTON RADIOGRAPHY IMAGING QUALITIES USING GEANT4 MONTE CARLO SIMULATIONS
    Depauw, Nicolas
    Danto, Sylvain
    Bednarz, Bryan
    Paganetti, Harald
    Fink, Yoel
    Seco, Joao
    [J]. NUCLEAR TECHNOLOGY, 2011, 175 (01) : 6 - 10
  • [10] The Geant4 Virtual Monte Carlo
    Hrivnacova, I.
    [J]. INTERNATIONAL CONFERENCE ON COMPUTING IN HIGH ENERGY AND NUCLEAR PHYSICS 2012 (CHEP2012), PTS 1-6, 2012, 396