Calibration method and photon flux influences tiled flat-panel photon counting detector image uniformity in computed tomography

被引:3
|
作者
Inkinen, S. I. [1 ]
Juntunen, M. A. K. [1 ,2 ,3 ,4 ]
Kotiaho, A. [1 ,2 ,3 ,4 ]
Nieminen, M. T. [1 ,2 ,3 ,4 ]
机构
[1] Univ Oulu, Res Unit Med Imaging Phys & Technol, Aapistie 5A, FI-90220 Oulu, Finland
[2] Univ Oulu, Med Res Ctr, Aapistie 5A, FI-90220 Oulu, Finland
[3] Oulu Univ Hosp, Aapistie 5A, FI-90220 Oulu, Finland
[4] Oulu Univ Hosp, Dept Diagnost Radiol, POB 50, FI-90029 Oulu, Finland
关键词
Computerized Tomography (CT) and Computed Radiography (CR); Data processing methods; Image reconstruction in medical imaging; X-ray detectors; CT; REDUCTION; ARTIFACTS;
D O I
10.1088/1748-0221/15/08/T08005
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The aim of this study is to compare how different calibration methods influence the image quality of photon-counting detector computed tomography (PCD-CT) at high and low photon fluxes. We investigate the performance of flat-field correction, signal-to-equivalent thickness calibration (STC), and polynomial correction (PC) methods using polymethyl methacrylate (PMMA) and iron as calibration materials. Two different cylindrical imaging phantoms containing contrast targets were scanned: an agar phantom and a phantom consisting of titanium hip implant embedded in agar. The scans were acquired using 120 kVp, and the energy thresholds of the PCD were set at 10 keV and 60 keV to obtain low energy (10-60 keV), high energy (60-120 keV) and total energy images (10-120 keV). Additionally, virtual monochromatic images (VMIs) with energies between 60-180 keV with 20 keV increments were generated from PC data. The reconstructions were made using filtered back projection, and image quality was assessed by evaluating image noise, contrast-to-noise ratio (CNR), and image uniformity. Overall, STC with PMMA as calibration material yielded the best image quality in terms of CNR and uniformity. Flat-field correction produced uniform reconstruction at low photon flux, but the performance degraded substantially at high flux. STC with iron as calibration material did not improve the reconstructions of the titanium hip implant. The beam hardening effects arising from metal were reduced when the VMI energy was increased while the CNR evaluated from agar phantom decreased with increasing energy of the VMI. Over the methods investigated, STC with PMMA was the most optimal calibration method for PCD-CT, yielding excellent image uniformity with both photon flux conditions.
引用
收藏
页数:19
相关论文
共 50 条
  • [31] Comparison of cone-beam computed tomography with photon-counting detector computed tomography for dental implant surgery
    Husain, Adib Al-Haj
    Mergen, Victor
    Valdec, Silvio
    Husain, Nadin Al-Haj
    Stadlinger, Bernd
    Essig, Harald
    Frauenfelder, Thomas
    Kessler, Peter
    Lie, Suen An Nynke
    Alkadhi, Hatem
    Winklhofer, Sebastian
    INTERNATIONAL JOURNAL OF IMPLANT DENTISTRY, 2025, 11 (01)
  • [32] Innovative advances in pediatric radiology: computed tomography reconstruction techniques, photon-counting detector computed tomography, and beyond
    Ismail Mese
    Ceren Altintas Mese
    Ugur Demirsoy
    Yonca Anik
    Pediatric Radiology, 2024, 54 : 1 - 11
  • [33] Innovative advances in pediatric radiology: computed tomography reconstruction techniques, photon-counting detector computed tomography, and beyond
    Mese, Ismail
    Mese, Ceren Altintas
    Demirsoy, Ugur
    Anik, Yonca
    PEDIATRIC RADIOLOGY, 2024, 54 (01) : 1 - 11
  • [34] Photon-counting X-ray Computed Tomography Using a Cadmium Telluride Flat Panel Detector with High Spatial Resolutions and Dual-energy Selection
    Sato, Jiro
    Sato, Eiichi
    Ito, Kazuki
    Moriyama, Hodaka
    Hagiwara, Osahiko
    Enomoto, Toshiyuki
    Watanabe, Manabu
    Yoshida, Sohei
    Yoshioka, Kunihiro
    Nitta, Hiroyuki
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (03) : 441 - 447
  • [35] An Image Quality Study of Medipix2 Single Photon Counting Detector Based on Two Kinds of Flat Field Corrections for Breast Computed Tomography Application
    Frallicciardi, P. M.
    Mettivier, G.
    Montesi, M. C.
    Russo, P.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 685 - 688
  • [36] Flat-panel cone-beam computed tomography for image-guided radiation therapy
    Jaffray, DA
    Siewerdsen, JH
    Wong, JW
    Martinez, AA
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1337 - 1349
  • [37] X-ray flat-panel detector geometry correction to improve dimensional computed tomography measurements
    Luethi, Melina
    Bircher, Benjamin A.
    Meli, Felix
    Kueng, Alain
    Thalmann, Rudolf
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2020, 31 (03)
  • [38] View-dependent geometric calibration for offset flat-panel cone beam computed tomography systems
    Van-Giang Nguyen
    OPTICAL ENGINEERING, 2016, 55 (04)
  • [39] Aging of imaging properties of a CMOS flat-panel detector for dental cone-beam computed tomography
    Kim, D. W.
    Han, J. C.
    Yun, S.
    Kim, H. K.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [40] Development of an Integrated C-Arm Interventional Imaging System With a Strip Photon Counting Detector and a Flat Panel Detector
    Ji, Xu
    Feng, Mang
    Treb, Kevin
    Zhang, Ran
    Schafer, Sebastian
    Li, Ke
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2021, 40 (12) : 3674 - 3685