Calibration method and photon flux influences tiled flat-panel photon counting detector image uniformity in computed tomography

被引:3
|
作者
Inkinen, S. I. [1 ]
Juntunen, M. A. K. [1 ,2 ,3 ,4 ]
Kotiaho, A. [1 ,2 ,3 ,4 ]
Nieminen, M. T. [1 ,2 ,3 ,4 ]
机构
[1] Univ Oulu, Res Unit Med Imaging Phys & Technol, Aapistie 5A, FI-90220 Oulu, Finland
[2] Univ Oulu, Med Res Ctr, Aapistie 5A, FI-90220 Oulu, Finland
[3] Oulu Univ Hosp, Aapistie 5A, FI-90220 Oulu, Finland
[4] Oulu Univ Hosp, Dept Diagnost Radiol, POB 50, FI-90029 Oulu, Finland
关键词
Computerized Tomography (CT) and Computed Radiography (CR); Data processing methods; Image reconstruction in medical imaging; X-ray detectors; CT; REDUCTION; ARTIFACTS;
D O I
10.1088/1748-0221/15/08/T08005
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The aim of this study is to compare how different calibration methods influence the image quality of photon-counting detector computed tomography (PCD-CT) at high and low photon fluxes. We investigate the performance of flat-field correction, signal-to-equivalent thickness calibration (STC), and polynomial correction (PC) methods using polymethyl methacrylate (PMMA) and iron as calibration materials. Two different cylindrical imaging phantoms containing contrast targets were scanned: an agar phantom and a phantom consisting of titanium hip implant embedded in agar. The scans were acquired using 120 kVp, and the energy thresholds of the PCD were set at 10 keV and 60 keV to obtain low energy (10-60 keV), high energy (60-120 keV) and total energy images (10-120 keV). Additionally, virtual monochromatic images (VMIs) with energies between 60-180 keV with 20 keV increments were generated from PC data. The reconstructions were made using filtered back projection, and image quality was assessed by evaluating image noise, contrast-to-noise ratio (CNR), and image uniformity. Overall, STC with PMMA as calibration material yielded the best image quality in terms of CNR and uniformity. Flat-field correction produced uniform reconstruction at low photon flux, but the performance degraded substantially at high flux. STC with iron as calibration material did not improve the reconstructions of the titanium hip implant. The beam hardening effects arising from metal were reduced when the VMI energy was increased while the CNR evaluated from agar phantom decreased with increasing energy of the VMI. Over the methods investigated, STC with PMMA was the most optimal calibration method for PCD-CT, yielding excellent image uniformity with both photon flux conditions.
引用
收藏
页数:19
相关论文
共 50 条
  • [21] Assessment of image quality in photon-counting detector computed tomography of the wrist-An ex vivo study
    Kammerling, Nina
    Sandstedt, Marten
    Farnebo, Simon
    Persson, Anders
    Tesselaar, Erik
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 154
  • [22] Cone-beam computed tomography with a flat-panel imager: Effects of image lag
    Siewerdsen, JH
    Jaffray, DA
    MEDICAL PHYSICS, 1999, 26 (12) : 2635 - 2647
  • [23] An Improved Ring Artifact Removal Approach for Flat-panel Detector Based Computed Tomography Images
    Zeng, Dong
    Ma, Jianhua
    Zhang, Yunwan
    Bian, Zhaoying
    Huang, Jing
    Chen, Wufan
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [24] Validation of Photon-counting Computed Tomography Image Data for Radiation Therapy
    Richter, Anne
    Razinskas, Gary
    Paulus, Patrick
    Weick, Stefan
    Lutyj, Paul
    Kraft, Johannes
    Huflage, Henner
    Grunz, Jan-Peter
    Bley, Thorsten
    Flentje, Michael
    STRAHLENTHERAPIE UND ONKOLOGIE, 2023, 199 : S141 - S141
  • [25] Image Quality Assessment of Photon Counting Computed Tomography Angiography of the Head and Neck
    Song, Jae W.
    Colfer, Michael
    Pasyar, Pouyan
    Kasner, Scott E.
    Cucchiara, Brett L.
    Saba, Luca
    Noel, Peter B.
    Woo, John H.
    STROKE, 2024, 55
  • [26] K-edge spectral computed tomography with a photon counting detector and discrete reconstruction
    Brun, Francesco
    Di Trapani, Vittorio
    Dreossi, Diego
    Longo, Renata
    Delogu, Pasquale
    Rigon, Luigi
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 5245 - 5248
  • [27] Photon-counting detector computed tomography Paradigm shift in cardiac CT imaging
    Kravchenko, Dmitrij
    Layer, Yannik Christian
    Vecsey-Nagy, Milan
    Tremamunno, Giuseppe
    Varga-Szemes, Akos
    Schlett, Christopher L.
    Bamberg, Fabian
    Emrich, Tilman
    Hagar, Muhammad Taha
    RADIOLOGIE, 2024, 64 (12): : 928 - 934
  • [28] A hybrid photon counting and flat panel detector system for periprocedural hemorrhage monitoring in the angio suite
    Treb, Kevin
    Ji, Xu
    Feng, Mang
    Zhang, Ran
    Schafer, Sebastian
    Li, Ke
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [29] Photon-Counting Detector Computed Tomography Versus Energy-Integrating Detector Computed Tomography for Coronary Artery Calcium Quantitation
    McCollough, Cynthia H.
    Winfree, Tim N.
    Melka, Elnata F.
    Rajendran, Kishore
    Carter, Rickey E.
    Leng, Shuai
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2024, 48 (02) : 212 - 216
  • [30] Optimizing iterative reconstruction for quantification of calcium hydroxyapatite with photon counting flat-detector computed tomography: a cardiac phantom study
    Juntunen, Mikael A. K.
    Kotiaho, Antti O.
    Nieminen, Miika T.
    Inkinen, Satu I.
    JOURNAL OF MEDICAL IMAGING, 2021, 8 (05)