Calibration method and photon flux influences tiled flat-panel photon counting detector image uniformity in computed tomography

被引:3
|
作者
Inkinen, S. I. [1 ]
Juntunen, M. A. K. [1 ,2 ,3 ,4 ]
Kotiaho, A. [1 ,2 ,3 ,4 ]
Nieminen, M. T. [1 ,2 ,3 ,4 ]
机构
[1] Univ Oulu, Res Unit Med Imaging Phys & Technol, Aapistie 5A, FI-90220 Oulu, Finland
[2] Univ Oulu, Med Res Ctr, Aapistie 5A, FI-90220 Oulu, Finland
[3] Oulu Univ Hosp, Aapistie 5A, FI-90220 Oulu, Finland
[4] Oulu Univ Hosp, Dept Diagnost Radiol, POB 50, FI-90029 Oulu, Finland
关键词
Computerized Tomography (CT) and Computed Radiography (CR); Data processing methods; Image reconstruction in medical imaging; X-ray detectors; CT; REDUCTION; ARTIFACTS;
D O I
10.1088/1748-0221/15/08/T08005
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The aim of this study is to compare how different calibration methods influence the image quality of photon-counting detector computed tomography (PCD-CT) at high and low photon fluxes. We investigate the performance of flat-field correction, signal-to-equivalent thickness calibration (STC), and polynomial correction (PC) methods using polymethyl methacrylate (PMMA) and iron as calibration materials. Two different cylindrical imaging phantoms containing contrast targets were scanned: an agar phantom and a phantom consisting of titanium hip implant embedded in agar. The scans were acquired using 120 kVp, and the energy thresholds of the PCD were set at 10 keV and 60 keV to obtain low energy (10-60 keV), high energy (60-120 keV) and total energy images (10-120 keV). Additionally, virtual monochromatic images (VMIs) with energies between 60-180 keV with 20 keV increments were generated from PC data. The reconstructions were made using filtered back projection, and image quality was assessed by evaluating image noise, contrast-to-noise ratio (CNR), and image uniformity. Overall, STC with PMMA as calibration material yielded the best image quality in terms of CNR and uniformity. Flat-field correction produced uniform reconstruction at low photon flux, but the performance degraded substantially at high flux. STC with iron as calibration material did not improve the reconstructions of the titanium hip implant. The beam hardening effects arising from metal were reduced when the VMI energy was increased while the CNR evaluated from agar phantom decreased with increasing energy of the VMI. Over the methods investigated, STC with PMMA was the most optimal calibration method for PCD-CT, yielding excellent image uniformity with both photon flux conditions.
引用
收藏
页数:19
相关论文
共 50 条
  • [41] Polychromatic Iterative Statistical Material Image Reconstruction for Photon-Counting Computed Tomography
    Weidinger, Thomas
    Buzug, Thorsten M.
    Flohr, Thomas
    Kappler, Steffen
    Stierstorfer, Karl
    INTERNATIONAL JOURNAL OF BIOMEDICAL IMAGING, 2016, 2016
  • [42] Spiral breast computed tomography with a photon-counting detector (SBCT): The future of breast imaging?
    Wetzl, Matthias
    Dietzel, Matthias
    Ohlmeyer, Sabine
    Uder, Michael
    Wenkel, Evelyn
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 157
  • [43] Computed tomography with a full FOV photon-counting detector in a clinical setting, the first experience
    Ferda, Jiri
    Vendis, Tomas
    Flohr, Thomas
    Schmidt, Bernhard
    Henning, Andre
    Ulzheimer, Stefan
    Pecen, Ladislav
    Ferdova, Eva
    Baxa, Jan
    Mirka, Hynek
    EUROPEAN JOURNAL OF RADIOLOGY, 2021, 137
  • [44] Design considerations to overcome cross talk in a photon counting silicon strip detector for computed tomography
    Bornefalk, Hans
    Xu, Cheng
    Svensson, Christer
    Danielsson, Mats
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3): : 371 - 378
  • [45] Quantitative cartilage imaging using spectral photon-counting detector based computed tomography
    Endran, Kishore Raj
    Tao, Shengzhen
    Benike, Amy
    Leng, Shuai
    McCollough, Cynthia
    MEDICAL IMAGING 2019: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2019, 10953
  • [46] Value of Ultrahigh-Resolution Photon-Counting Detector Computed Tomography in Cardiac Imaging
    Kravchenko, Dmitrij
    Hagar, Muhammad Taha
    Vecsey-Nagy, Milan
    Tremamunno, Giuseppe
    Szilveszter, Balint
    Vattay, Borbala
    Zsarnoczay, Emese
    Beke, Samuel
    Maurovich-Horvat, Pal
    Emrich, Tilman
    Varga-Szemes, Akos
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2025, 42 (02):
  • [47] Flat-Panel Detector Computed Tomography Imaging Observer Performance in Detecting Pulmonary Nodules in Comparison With Conventional Chest Radiography and Multidetector Computed Tomography
    Otani, Hideji
    Nitta, Norihisa
    Ikeda, Mitsuru
    Nagatani, Yukihiro
    Tanaka, Toyohiko
    Kitahara, Hitoshi
    Murakami, Yoko
    Takahashi, Masashi
    Murata, Kiyoshi
    JOURNAL OF THORACIC IMAGING, 2012, 27 (01) : 51 - 57
  • [48] Flat-panel detector computed tomography for the assessment of coronary artery stents -: Phantom study in comparison with 16-slice spiral computed tomography
    Mahnken, AH
    Seyfarth, T
    Flohr, T
    Herzog, C
    Stahl, J
    Stanzel, S
    Kuettner, A
    Wildberger, JE
    Günther, RW
    INVESTIGATIVE RADIOLOGY, 2005, 40 (01) : 8 - 13
  • [49] Investigation of the Lag Effect in X-Ray Flat-Panel Detector for Cone-Beam Computed Tomography
    A. K. Avakyan
    I. L. Dergacheva
    A. A. Elanchik
    T. A. Krylova
    T. K. Lobzhanidze
    S. A. Polikhov
    V. P. Smirnov
    Biomedical Engineering, 2020, 54 : 179 - 183
  • [50] Morphologic changes of mammary carcinomas in mice over time as monitored by flat-panel detector volume computed tomography
    Missbach-Guentner, Jeannine
    Dullin, Christian
    Kimmina, Sarah
    Zientkowska, Marta
    Domeyer-Missbach, Melanie
    Malz, Cordula
    Grabbe, Eckhardt
    Stuhmer, Walter
    Alves, Frauke
    NEOPLASIA, 2008, 10 (07): : 663 - 673