View-dependent geometric calibration for offset flat-panel cone beam computed tomography systems

被引:3
|
作者
Van-Giang Nguyen [1 ]
机构
[1] Le Quy Don Tech Univ, Dept Informat Syst, 236 Hoang Quoc Viet, Hanoi, Vietnam
关键词
geometric calibration; perspective geometry; computed tomography; offset flat-panel cone beam computed tomography; projection matrix; MISALIGNMENT CORRECTION; CT;
D O I
10.1117/1.OE.55.4.043102
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Geometric parameters that define the geometry of imaging systems are crucial for image reconstruction and image quality in x-ray computed tomography (CT). The problem of determining geometric parameters for an offset flat-panel cone beam CT (CBCT) system, a recently introduced modality with a large field of view, with the assumption of an unstable mechanism and geometric parameters that vary in each view, is considered. To accurately and rapidly find the geometric parameters for each projection view, we use the projection matrix method and design a dedicated phantom that is partially visible in all projection views. The phantom consists of balls distributed symmetrically in a cylinder to ensure the inclusion of the phantom in all views, and a large portion of the phantom is covered in the projection image. To efficiently use calibrated geometric information in the reconstruction process and get rid of approximation errors, instead of decomposing the projection matrix into actual geometric parameters that are manually corrected before being used in reconstruction, as in conventional methods, we directly use the projection matrix and its pseudo-inverse in projection and backprojection operations of reconstruction algorithms. The experiments illustrate the efficacy of the proposed method with a real offset flat-panel CBCT system in dental imaging. (C) 2016 Society of Photo-Optical Instrumentation Engineers (SPIE)
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Geometric Calibration of Cone-beam CT with a Flat-panel Detector
    Wu, Dufan
    Li, Liang
    Zhang, Li
    Xing, Yuxiang
    Chen, Zhiqiang
    Xiao, Yongshun
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2952 - 2955
  • [2] Planar cone-beam computed tomography with a flat-panel detector
    Kim, S. H.
    Kim, D. W.
    Youn, H.
    Kim, D.
    Kam, S.
    Jeon, H.
    Kim, H. K.
    [J]. JOURNAL OF INSTRUMENTATION, 2015, 10
  • [3] Geometric Calibration for Offset Flat-panel CBCT Systems using Projection Matrix
    Van-Giang Nguyen
    [J]. 2016 INTERNATIONAL CONFERENCE ON ELECTRONICS, INFORMATION, AND COMMUNICATIONS (ICEIC), 2016,
  • [4] Cone-beam computed tomography with a flat-panel imager: Initial performance characterization
    Jaffray, DA
    Siewerdsen, JH
    [J]. MEDICAL PHYSICS, 2000, 27 (06) : 1311 - 1323
  • [5] Cone-beam computed tomography with a flat-panel imager: Effects of image lag
    Siewerdsen, JH
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 1999, 26 (12) : 2635 - 2647
  • [6] Noise Characterization of a Clinical Flat-Panel Cone-Beam Computed Tomography (CBCT) System
    Vedantham, S.
    Gounis, M.
    van der Bom, M.
    Karellas, A.
    Wakhloo, A.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [7] Flat panel cone beam computed tomography of the sinuses
    Zoumalan, Richard A.
    Lebowitz, Richard A.
    Wang, Edwin
    Yung, Kathryn
    Babb, James S.
    Jacobs, Joseph B.
    [J]. OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2009, 140 (06) : 841 - 844
  • [8] Flat-panel cone-beam computed tomography for image-guided radiation therapy
    Jaffray, DA
    Siewerdsen, JH
    Wong, JW
    Martinez, AA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1337 - 1349
  • [9] Cone-beam computed tomography on a medical linear accelerator using a flat-panel imager
    Jaffray, DA
    Siewerdsen, JH
    Groh, B
    Drake, DG
    Wong, JW
    Martinez, AA
    [J]. USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 558 - 560
  • [10] Simulation of Motion Artifacts in Offset Flat-Panel Cone-Beam CT