Cone-beam computed tomography with a flat-panel imager: Initial performance characterization

被引:452
|
作者
Jaffray, DA [1 ]
Siewerdsen, JH [1 ]
机构
[1] William Beaumont Hosp, Dept Radiat Oncol, Royal Oak, MI 48073 USA
关键词
cone-beam computed tomography; flat-panel imager; amorphous silicon; volume imaging; image-guided therapy; radiotherapy;
D O I
10.1118/1.599009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The development and performance of a system for x-ray cone-beam computed tomography (CBCT) using an indirect-detection flat-panel imager (FPI) is presented. Developed as a bench-top prototype for initial investigation of FPI-based CBCT for bone and soft-tissue localization in radiotherapy, the system provides fully three-dimensional volumetric image data from projections acquired during a single rotation. The system employs a 512x512 active matrix of a-Si:H thin-film transistors and photodiodes in combination with a luminescent phosphor. Tomographic imaging performance is quantified in terms of response uniformity, response linearity, voxel noise, noise-power spectrum (NPS), and modulation transfer function (MTF), each in comparison to the performance measured on a conventional CT scanner. For the geometry employed and the objects considered, response is uniform to within 2% and linear within 1%. Voxel noise, at a level of similar to 20 HU, is comparable to the conventional CT scanner. NPS and MTF results highlight the frequency-dependent transfer characteristics, confirming that the CBCT system can provide high spatial resolution and does not suffer greatly from additive noise levels. For larger objects and/or low exposures, additive noise levels must be reduced to maintain high performance. Imaging studies of a low-contrast phantom and a small animal (a euthanized rat) qualitatively demonstrate excellent soft-tissue visibility and high spatial resolution. Image quality appears comparable or superior to that of the conventional scanner, These quantitative and qualitative results clearly demonstrate the potential of CBCT systems based upon flat-panel imagers. Advances in FPI technology (e.g., improved x-ray converters and enhanced electronics) are anticipated to allow high-performance FPI-based CBCT for medical imaging. General and specific requirements of kilovoltage CBCT systems are discussed, and the applicability of FPI-based CBCT systems to tomographic localization and image-guidance for radiotherapy is considered. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)01306-7].
引用
收藏
页码:1311 / 1323
页数:13
相关论文
共 50 条
  • [1] Cone-beam computed tomography with a flat-panel imager: Effects of image lag
    Siewerdsen, JH
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 1999, 26 (12) : 2635 - 2647
  • [2] Cone-beam computed tomography on a medical linear accelerator using a flat-panel imager
    Jaffray, DA
    Siewerdsen, JH
    Groh, B
    Drake, DG
    Wong, JW
    Martinez, AA
    [J]. USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 558 - 560
  • [3] Planar cone-beam computed tomography with a flat-panel detector
    Kim, S. H.
    Kim, D. W.
    Youn, H.
    Kim, D.
    Kam, S.
    Jeon, H.
    Kim, H. K.
    [J]. JOURNAL OF INSTRUMENTATION, 2015, 10
  • [4] Noise Characterization of a Clinical Flat-Panel Cone-Beam Computed Tomography (CBCT) System
    Vedantham, S.
    Gounis, M.
    van der Bom, M.
    Karellas, A.
    Wakhloo, A.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [5] Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter
    Siewerdsen, JH
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 2001, 28 (02) : 220 - 231
  • [6] Cone-beam CT with a flat-panel imager: Noise considerations for fully 3-D computed tomography
    Siewerdsen, JH
    Jaffray, DA
    [J]. MEDICAL IMAGING 2000: PHYSICS OF MEDICAL IMAGING, 2000, 3977 : 408 - 416
  • [7] Low-dose 2.5 MV cone-beam computed tomography with thick CsI flat-panel imager
    Tang, Grace
    Moussot, Christopher
    Morf, Daniel
    Seppi, Edward
    Amols, Howard
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (04): : 235 - 245
  • [8] Flat-panel cone-beam computed tomography for image-guided radiation therapy
    Jaffray, DA
    Siewerdsen, JH
    Wong, JW
    Martinez, AA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1337 - 1349
  • [9] Performance of cone beam CT using a flat-panel imager
    Endo, M
    Tsunoo, T
    Satoh, K
    Matsusita, S
    Kusakabe, M
    Fukuda, Y
    [J]. MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 815 - 821
  • [10] A performance comparison of flat-panel imager-based MV and kV cone-beam CT
    Groh, BA
    Siewerdsen, JH
    Drake, DG
    Wong, JW
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 967 - 975