Cone-beam computed tomography with a flat-panel imager: Initial performance characterization

被引:452
|
作者
Jaffray, DA [1 ]
Siewerdsen, JH [1 ]
机构
[1] William Beaumont Hosp, Dept Radiat Oncol, Royal Oak, MI 48073 USA
关键词
cone-beam computed tomography; flat-panel imager; amorphous silicon; volume imaging; image-guided therapy; radiotherapy;
D O I
10.1118/1.599009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The development and performance of a system for x-ray cone-beam computed tomography (CBCT) using an indirect-detection flat-panel imager (FPI) is presented. Developed as a bench-top prototype for initial investigation of FPI-based CBCT for bone and soft-tissue localization in radiotherapy, the system provides fully three-dimensional volumetric image data from projections acquired during a single rotation. The system employs a 512x512 active matrix of a-Si:H thin-film transistors and photodiodes in combination with a luminescent phosphor. Tomographic imaging performance is quantified in terms of response uniformity, response linearity, voxel noise, noise-power spectrum (NPS), and modulation transfer function (MTF), each in comparison to the performance measured on a conventional CT scanner. For the geometry employed and the objects considered, response is uniform to within 2% and linear within 1%. Voxel noise, at a level of similar to 20 HU, is comparable to the conventional CT scanner. NPS and MTF results highlight the frequency-dependent transfer characteristics, confirming that the CBCT system can provide high spatial resolution and does not suffer greatly from additive noise levels. For larger objects and/or low exposures, additive noise levels must be reduced to maintain high performance. Imaging studies of a low-contrast phantom and a small animal (a euthanized rat) qualitatively demonstrate excellent soft-tissue visibility and high spatial resolution. Image quality appears comparable or superior to that of the conventional scanner, These quantitative and qualitative results clearly demonstrate the potential of CBCT systems based upon flat-panel imagers. Advances in FPI technology (e.g., improved x-ray converters and enhanced electronics) are anticipated to allow high-performance FPI-based CBCT for medical imaging. General and specific requirements of kilovoltage CBCT systems are discussed, and the applicability of FPI-based CBCT systems to tomographic localization and image-guidance for radiotherapy is considered. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)01306-7].
引用
收藏
页码:1311 / 1323
页数:13
相关论文
共 50 条
  • [31] Noise, sampling, and the number of projections in cone-beam CT with a flat-panel detector
    Zhao, Z.
    Gang, G. J.
    Siewerdsen, J. H.
    [J]. MEDICAL PHYSICS, 2014, 41 (06)
  • [32] Iterative Reconstruction for Circular Cone-Beam CT with an Offset Flat-Panel Detector
    Hansis, Eberhard
    Bredno, Joerg
    Sowards-Emmerd, David
    Shao, Lingxiong
    [J]. 2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2228 - 2231
  • [33] A volumetric cone-beam CT system based on a 41x41 cm2 flat-panel imager
    Jaffray, DA
    Siewerdsen, JH
    [J]. MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 800 - 807
  • [34] Quality assurance of the on-board imager and cone-beam computed tomography systems
    Yoo, S.
    Kim, G.
    Hammoud, R.
    Elder, E.
    Pawlicki, T.
    Guan, H.
    Fox, T.
    Luxton, G.
    Yin, F.
    Munro, P.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2131 - 2132
  • [35] Comparison of flat-panel detector with image-intensifier detector for cone-beam CT
    Baba, R
    Konno, Y
    Ueda, M
    Ikeda, S
    [J]. RADIOLOGY, 2000, 217 : 405 - 405
  • [36] Comparison of flat-panel detector and image-intensifier detector for cone-beam CT
    Baba, R
    Konno, Y
    Ueda, K
    Ikeda, S
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2002, 26 (03) : 153 - 158
  • [37] Design and performance characteristics of a digital flat-panel computed tomography system
    Ross, William
    Cody, Dianna D.
    Hazle, John D.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 1888 - 1901
  • [38] Cone-beam computed tomography for trauma
    Gupta, Saurabh
    Martinson, James R.
    Ricaurte, Daniel
    Scalea, Thomas M.
    Morrison, Jonathan J.
    [J]. JOURNAL OF TRAUMA AND ACUTE CARE SURGERY, 2020, 89 (03): : E34 - E40
  • [39] Cone-Beam Computed Tomography in Orthodontics
    Abdelkarim, Ahmad
    [J]. DENTISTRY JOURNAL, 2019, 7 (03)
  • [40] Four-dimensional cone-beam computed tomography using an on-board imager
    Li, Tianfang
    Xing, Lei
    Munro, Peter
    McGuinness, Christopher
    Chao, Ming
    Yang, Yong
    Loo, Bill
    Koong, Albert
    [J]. MEDICAL PHYSICS, 2006, 33 (10) : 3825 - 3833