Geometric Calibration of Cone-beam CT with a Flat-panel Detector

被引:0
|
作者
Wu, Dufan [1 ]
Li, Liang [1 ]
Zhang, Li [1 ]
Xing, Yuxiang [1 ]
Chen, Zhiqiang [1 ]
Xiao, Yongshun [1 ]
机构
[1] Tsinghua Univ, Dept Engn Phys, Beijing 100084, Peoples R China
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Here we proposed a method to calibrate several important geometric parameters for cone-beam CT (CBCT) systems. Only three point-like objects and their spatial distances are required for the calibration. Our method is based on the analysis of the ellipse trajectories of the points' projections. Our mathematical expressions are much simpler than similar methods and our method yields better stability. The calibration procedure has been tested on numerical simulations and an actual micro-CT system.
引用
收藏
页码:2952 / 2955
页数:4
相关论文
共 50 条
  • [1] Comparison of flat-panel detector with image-intensifier detector for cone-beam CT
    Baba, R
    Konno, Y
    Ueda, M
    Ikeda, S
    RADIOLOGY, 2000, 217 : 405 - 405
  • [2] Comparison of flat-panel detector and image-intensifier detector for cone-beam CT
    Baba, R
    Konno, Y
    Ueda, K
    Ikeda, S
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2002, 26 (03) : 153 - 158
  • [3] Noise, sampling, and the number of projections in cone-beam CT with a flat-panel detector
    Zhao, Z.
    Gang, G. J.
    Siewerdsen, J. H.
    MEDICAL PHYSICS, 2014, 41 (06)
  • [4] Iterative Reconstruction for Circular Cone-Beam CT with an Offset Flat-Panel Detector
    Hansis, Eberhard
    Bredno, Joerg
    Sowards-Emmerd, David
    Shao, Lingxiong
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2228 - 2231
  • [5] Planar cone-beam computed tomography with a flat-panel detector
    Kim, S. H.
    Kim, D. W.
    Youn, H.
    Kim, D.
    Kam, S.
    Jeon, H.
    Kim, H. K.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [6] Lag correction for flat-panel cone-beam CT in SPECT/CT
    Wang, Jiong
    Hansis, Eberhard
    van de Haar, Peter
    Sowards-Emmerd, David
    Bredno, Joerg
    Shao, Lingxiong
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [7] Iterative cone-beam CT reconstruction with an off-center flat-panel detector for SPECT/CT
    Hansis, Eberhard
    Sowards-Emmerd, David
    Hines, Horace
    Shao, Lingxiong
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [8] Simulation of Motion Artifacts in Offset Flat-Panel Cone-Beam CT
    Hansis, Eberhard
    Shao, Lingxiong
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2471 - 2475
  • [9] Image noise due to quantum fluctuations in flat-panel detector based cone-beam CT imaging
    Zhang, Y
    Ning, R
    Conover, D
    Yu, Y
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 656 - 663
  • [10] Cone-beam CT with a flat-panel detector: From image science to image-guided surgery
    Siewerdsen, Jeffrey H.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 648 : S241 - S250