Normalized mean glandular dose computation from mammography using GATE: a validation study

被引:16
|
作者
Myronakis, Marios E. [1 ,2 ]
Zvelebil, Marketa [3 ]
Darambara, Dimitra G. [1 ,2 ]
机构
[1] Inst Canc Res, Div Radiotherapy & Imaging, London SW3 6JJ, England
[2] Royal Marsden NHS Fdn Trust, London SW3 6JJ, England
[3] Inst Canc Res, Breast Breakthrough Res Ctr, London SW3 6JB, England
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2013年 / 58卷 / 07期
关键词
SIMULATION TOOLKIT; BREAST MODEL; DOSIMETRY; CT; MOLYBDENUM; RHODIUM; PET;
D O I
10.1088/0031-9155/58/7/2247
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Mean glandular dose (MGD) is the figure of merit to assess breast dose after a mammographic acquisition. The use of normalized MGD obtained from Monte Carlo computations with measured incident air kerma determines the MGD delivered to patients. The Geant4 Application for Tomographic Emission (GATE) toolkit is a modern Monte Carlo application specifically designed for medical imaging systems modelling. Although there is an increasing number of publications using GATE worldwide for a wide range of medical imaging and therapeutic applications, there is currently no means to obtain normalized MGD. In this work, the GATE toolkit is extended, through the development of two new modules, to provide normalized MGD information for compressed breast phantoms based on simple geometries. The normalized MGD values were validated against published work and provided results at half value layers lower than 0.3 and greater than 0.6 mmAl. In addition, the skin thickness and composition were considered. Normalized MGD was computed after substitution of the adipose layer surrounding the standard breast phantom with skin tissue and the relative difference is reported. Spectrum generation was facilitated by further development of previously published work by other authors. Validation of the new GATE extension showed good agreement with published data and can be used to assess breast dose from mammographic as well as more complex x-ray imaging techniques. Changing skin thickness and composition revealed substantial changes in normalized MGD specifically for compressed breast thickness different than 5 cm and a possible revision of the structure of the standard breast model may be necessary.
引用
收藏
页码:2247 / 2265
页数:19
相关论文
共 50 条
  • [1] NORMALIZED AVERAGE GLANDULAR DOSE IN MAGNIFICATION MAMMOGRAPHY
    LIU, B
    GOODSITT, M
    CHAN, HP
    [J]. RADIOLOGY, 1995, 197 (01) : 27 - 32
  • [2] ASSESSMENT OF MEAN GLANDULAR DOSE IN MAMMOGRAPHY
    FAULKNER, K
    LAW, J
    ROBSON, KJ
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (812): : 877 - 881
  • [3] Estimating mean glandular dose using proprietary mammography phantoms
    Hartley, LD
    Cobb, BJ
    Hutchinson, DE
    [J]. APPLIED RADIATION AND ISOTOPES, 1999, 50 (01) : 205 - 213
  • [4] Parametrization of mammography normalized average glandular dose tables
    Sobol, WT
    Wu, XZ
    [J]. MEDICAL PHYSICS, 1997, 24 (04) : 547 - 554
  • [5] Computation and parameterization of normalized glandular dose using Geant4
    Kadri, Omrane
    Alnafea, Mohammed Ali
    Shamma, Khaled
    [J]. NUCLEAR SCIENCE AND TECHNIQUES, 2015, 26 (03)
  • [6] Computation and parameterization of normalized glandular dose using Geant4
    Omrane Kadri
    Mohammed Ali Alnafea
    Khaled Shamma
    [J]. Nuclear Science and Techniques, 2015, 26 (03) : 53 - 58
  • [7] Evaluation of mean glandular dose from mammography screening: A single-center study
    Alahmad, Haitham
    AlEnazi, Khaled
    Alshahrani, Abdulrhman
    Albariqi, Saleh
    Alnafea, Mohammed
    [J]. JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2023, 16 (04)
  • [8] Monte Carlo Evaluation of Normalized Glandular Dose Coefficients in Mammography
    Sarno, Antonio
    Mettivier, Giovanni
    Di Lillo, Francesca
    Russo, Paolo
    [J]. BREAST IMAGING, IWDM 2016, 2016, 9699 : 190 - 196
  • [9] PORTUGUESE STUDY OF MEAN GLANDULAR DOSE IN MAMMOGRAPHY AND COMPARISON WITH EUROPEAN REFERENCES
    dos Reis, Claudia Sa
    Fartaria, Mario Joao
    Garcia Alves, Joao H.
    Pascoal, Ana
    [J]. RADIATION PROTECTION DOSIMETRY, 2018, 179 (04) : 391 - 399
  • [10] Assessment of Mean Glandular Dose and Entrance Surface Dose in Mammography
    Abdallah, Yousif M.
    Abuhadi, Nouf H.
    Aldousari, Hanan A.
    [J]. JOURNAL OF RESEARCH IN MEDICAL AND DENTAL SCIENCE, 2021, 9 (02): : 8 - 11