Validation of a digital method for patient-specific verification of VMAT treatment using a 2D ionisation detector array

被引:1
|
作者
U'wais, Fatima Adel [1 ]
Radzi, Yasmin [2 ]
Rizan, Naslinda Noor [3 ]
Zin, Hafiz M. [1 ]
机构
[1] Univ Sains Malaysia, Adv Med & Dent Inst, Kepala Batas 13200, Pulau Pinang, Malaysia
[2] Univ Sains Malaysia, Sch Phys, George Town 11800, Malaysia
[3] Gleneagles Hosp, George Town, Malaysia
关键词
VMAT; Patient-specific QA; Log file; QUALITY-ASSURANCE; LOG FILES; MLC; THERAPY; ACCURACY; QA;
D O I
10.1016/j.radphyschem.2022.110536
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
VMAT treatment involves the delivery of complex beam parameters including the dynamic multi-leaf collimators (MLC) from a linac. The use of a detector array has become a standard method to verify the treatment accuracy before the treatment planned is delivered to the patient with some limitations. This study aims to evaluate the accuracy of a high-resolution, digital method using linac log data for VMAT treatment verification. The log data was obtained from a Versa HD (Elekta ltd., Crawley, UK) linac during VMAT treatment. All the dynamic treatment parameters including MLC position, and MLC speed, were recorded at the rate of 4 Hz in the log data. The treatment delivery was also measured simultaneously using an Octavius 1500 2D ionisation chamber array (PTW, Freiburg, Germany) during the log data measurements. The parameters were extracted and analysed using algorithms written in MATLAB (MathWorks, Natick, MA). The results demonstrated that the log data can track MLC leaves with an accuracy of 1 mm at speeds ranging from 3.04 to 3.40 cm/s. Evaluation of the fluence generated for the VMAT delivery using log data was shown to agree well with the planned dose distribution measured in the 2D detector array, with an average gamma pass rate of 97.5% at 3%/3 mm. Log data obtained a higher gamma pass rate of 98.5% compared to the 2D detector array of 97.5%. The digital linac log data provides the basis for an essential high-resolution real-time verification tool that may be used for routine VMAT verification.
引用
下载
收藏
页数:7
相关论文
共 50 条
  • [21] Silicon strip detector for a novel 2D dosimetric method for radiotherapy treatment verification
    Bocci, A.
    Cortes-Giraldo, M. A.
    Gallardo, M. I.
    Espino, J. M.
    Arrans, R.
    Alvarez, M. A. G.
    Abou-Haidar, Z.
    Quesada, J. M.
    Perez Vega-Leal, A.
    Perez Nieto, F. J.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 673 : 98 - 106
  • [22] Characterization of responses of 2d array seven29 detector and its combined use with octavius phantom for the patient-specific quality assurance in rapidarc treatment delivery
    Syamkumar, S. A.
    Padmanabhan, Sriram
    Sukumar, Prabakar
    Nagarajan, Vivekanandan
    MEDICAL DOSIMETRY, 2012, 37 (01) : 53 - 60
  • [23] Volumetric Patient-Specific Quality Assurance for VMAT Treatment Delivery Using the Dynalog Files
    Schreibmann, E.
    Dhabaan, A.
    Elder, E.
    Fox, T.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [24] Retrospective analysis of 2D patient-specific IMRT verifications
    Childress, NL
    White, RA
    Bloch, C
    Salehpour, M
    Dong, L
    Rosen, II
    MEDICAL PHYSICS, 2005, 32 (04) : 838 - 850
  • [25] Implementation of 2D Arrays for TomoTherapy Patient-Specific QA
    Winningham, T.
    Staton, R.
    Meeks, S.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [26] Characterization of a novel 2D array dosimeter for patient-specific quality assurance with volumetric arc therapy
    Stathakis, Sotirios
    Myers, Pamela
    Esquivel, Carlos
    Mavroidis, Panayiotis
    Papanikolaou, Nikos
    MEDICAL PHYSICS, 2013, 40 (07)
  • [27] TomoEQA: Dose verification for patient-specific quality assurance in helical tomotherapy using an exit detector
    Han, Min Cheol
    Chang, Kyung Hwan
    Kim, Jihun
    Han, Su Chul
    Park, Kwangwoo
    Kim, Dong Wook
    Kim, Hojin
    Kim, Jin Sung
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 82 : 1 - 6
  • [28] VMAT patient-specific QA with 2D-array seven29/Octavius system: a retrospective analysis on 1000 patients
    Cilla, S.
    Viola, P.
    Craus, M.
    Deodato, F.
    Macchia, G.
    Digesu, C.
    Valentini, V.
    Piermattei, A.
    Morganti, A. G.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S752 - S753
  • [29] VMAT Monthly Quality Assurance Using a 2D Ionization Chamber Array
    Myers, P.
    Buckey, C.
    Mihailidis, D.
    Mavroidis, P.
    Esquivel, C.
    Gutierrez, A.
    Papanikolaou, N.
    Stathakis, S.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [30] Electron Beam Dosimetry Using a 2D Detector Array
    Srivastava, S.
    Das, I.
    MEDICAL PHYSICS, 2009, 36 (06)