TomoEQA: Dose verification for patient-specific quality assurance in helical tomotherapy using an exit detector

被引:7
|
作者
Han, Min Cheol [1 ]
Chang, Kyung Hwan [1 ]
Kim, Jihun [1 ]
Han, Su Chul [1 ]
Park, Kwangwoo [1 ]
Kim, Dong Wook [1 ]
Kim, Hojin [1 ]
Kim, Jin Sung [1 ]
机构
[1] Yonsei Univ, Coll Med, Dept Radiat Oncol, Seoul, South Korea
关键词
Helical tomotherapy; Patient-specific quality assurance; Phantom-less quality assurance; Leaf open time; Mobius3D; CLINICAL IMPLEMENTATION; DELIVERY; IMRT; QA; DOSIMETRY;
D O I
10.1016/j.ejmp.2020.12.021
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Existing phantom-less quality assurance (QA) platforms does not provide patient-specific QA for helical tomotherapy (HT). A new system, called TomoEQA, is presented to facilitate this using the leaf open time (LOT) of a binary multi-leaf collimator, as measured by an exit detector. Methods: TomoEQA was designed to provide measurement-based LOTs based on detector data and to generate a new digital imaging and communication in medicine (DICOM) dataset that includes the measured LOTs for use by secondary check platforms. To evaluate the system, 20 patient-specific QAs were performed using the program in Mobius3D software, and the results were compared to conventional phantom-based QA results. Results: From our assessment, most of the differences between the planned and measured (or calculated) data, excluding one case, were within the acceptance criteria comparing with those of conventional QA. Regarding the gamma analysis, all results considered in this study were within the acceptance criteria. In addition, the developed system was performed for a failed case and showed approximately the same trends as the conventional approach. Conclusions: TomoEQA could perform patient-specific QAs of HT using Mobius3D and provide reliable patientspecific QAs results by evaluating point dose errors and 3D gamma passing rates. TomoEQA could also distinguish whether an intensity-modulated radiation therapy plan failed or not.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 50 条
  • [1] Comprehensive clinical evaluation of TomoEQA for patient-specific pre-treatment quality assurance in helical tomotherapy
    Changhwan Kim
    Min Cheol Han
    Young Kyu Lee
    Han-Back Shin
    Hojin Kim
    Jin Sung Kim
    [J]. Radiation Oncology, 17
  • [2] Comprehensive clinical evaluation of TomoEQA for patient-specific pre-treatment quality assurance in helical tomotherapy
    Kim, Changhwan
    Han, Min Cheol
    Lee, Young Kyu
    Shin, Han-Back
    Kim, Hojin
    Kim, Jin Sung
    [J]. RADIATION ONCOLOGY, 2022, 17 (01)
  • [3] Patient-specific dose verification and synchronization quality assurance tests for helical tornotherapy using film
    James, J
    Hui, S
    Mackie, T
    Fenwick, J
    [J]. MEDICAL PHYSICS, 2004, 31 (06) : 1845 - 1845
  • [4] Automated patient-specific IMRT quality assurance using exit detector data
    Seibert, R.
    Ramsey, C.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2168 - 2168
  • [5] Tomotherapy Patient-Specific Delivery QA Using Exit Detector Array
    Chen, Y.
    Reeher, M.
    Mo, X.
    Chen, M.
    Olivera, G.
    Key, S.
    Parnell, D. W.
    Galmarini, D.
    Lu, W.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S740 - S740
  • [6] Patient specific quality assurance for helical tomotherapy
    Thomas, S
    MacKenzie, M
    Field, G
    Syme, A
    Fallone, B
    [J]. MEDICAL PHYSICS, 2005, 32 (07) : 2414 - 2414
  • [7] Comparative Evaluation of Four Commercial Patient-Specific Quality Assurance Devices for Helical Tomotherapy
    Chang, K.
    Lee, J.
    Choi, J.
    Hong, C.
    Jung, D.
    Kim, J.
    Kim, J.
    Han, M.
    Lee, H.
    Kim, D.
    Kim, J.
    Park, K.
    Cho, J.
    Kim, Y.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E611 - E611
  • [8] Dosimetric Comparison of Four Commercial Patient-Specific Quality Assurance Devices for Helical Tomotherapy
    Chang, Kyung Hwan
    Kim, Dong Wook
    Choi, Jung Hoan
    Shin, Han-Back
    Hong, Chae-Seon
    Jung, Dong Min
    Kim, Joo Ho
    Kim, Jihun
    Han, Mincheol
    Lee, Ho
    Park, Kwangwoo
    Kim, Jin Sung
    Cho, Jaeho
    Kim, Yong Bae
    Lee, Joo Ah
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2020, 76 (03) : 257 - 263
  • [9] Dosimetric Comparison of Four Commercial Patient-Specific Quality Assurance Devices for Helical Tomotherapy
    Kyung Hwan Chang
    Dong Wook Kim
    Jung Hoan Choi
    Han-Back Shin
    Chae-Seon Hong
    Dong Min Jung
    Joo Ho Kim
    Jihun Kim
    Mincheol Han
    Ho Lee
    Kwangwoo Park
    Jin Sung Kim
    Jaeho Cho
    Yong Bae Kim
    Joo Ah Lee
    [J]. Journal of the Korean Physical Society, 2020, 76 : 257 - 263
  • [10] Clinical implementation of an exit detector-based dose reconstruction tool for helical tomotherapy delivery quality assurance
    Deshpande, Shrikant
    Xing, Aitang
    Metcalfe, Peter
    Holloway, Lois
    Vial, Philip
    Geurts, Mark
    [J]. MEDICAL PHYSICS, 2017, 44 (10) : 5457 - 5466