Simultaneous high-pitch multi-energy CT pulmonary angiography using a dual-source photon-counting-detector CT: A phantom experiment

被引:0
|
作者
Mihailovic, Jelena M. [1 ]
Bruesewitz, Michael R. [1 ]
Swicklik, Joseph R. [1 ]
Yalon, Mariana [1 ]
Rajiah, Prabhakar S. [1 ]
Fletcher, Joel G. [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
关键词
dual source computed tomography; photon-counting-detector computed tomography; pulmonary embolism; PERFUSED BLOOD-VOLUME; ENERGY CT; EMBOLISM;
D O I
10.1002/acm2.14496
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A dual-source CT system can be operated in a high-pitch helical mode to provide a temporal resolution of 66 ms, which reduces motion artifacts in CT pulmonary angiography (CTPA). It can also be operated in a multi-energy (ME) mode to provide iodine maps, beneficial in the evaluation of pulmonary embolism (PE). No energy-integrating detector (EID) CT can perform simultaneous ME and high-pitch acquisition. This phantom study aimed to evaluate the ability of a photon-counting-detector (PCD) CT to perform simultaneous high-pitch and ME imaging for CTPA. Methods: A motion phantom was used to mimic the respiratory motion. Two tubes filled with iodine with intravascular thrombus mimicked by injecting glue within the tubes were placed along with 5, 10, and 15 mg/mL iodine samples, on a motion phantom at 20 and 30 revolutions per minute. Separate high-pitch and ME EID-CT scans and a single high-pitch ME PCD scan were acquired and virtual monoenergetic images and iodine maps reconstructed. Percent thrombus occlusion was measured and compared between static and moving images. Results: When there was motion, EID-CT ME suffered from significant motion artifacts. The measured iodine concentrations with PCD-CT in high-pitch ME were more stable when there was a motion, with a lower standard deviation than EID-CT in ME mode. The estimated percent thrombus occlusion dropped significantly with applied motion on EID-CT, while PCD-CT high-pitch ME mode showed good agreement between measurements on static or moving images. Conclusion: PCD-CT with combined ME and high-pitch mode facilitates simultaneous accurate iodine quantification and assessment of intravascular occlusion.
引用
下载
收藏
页数:8
相关论文
共 50 条
  • [21] High-pitch dual-source CT coronary angiography: systolic data acquisition at high heart rates
    Robert Goetti
    Gudrun Feuchtner
    Paul Stolzmann
    Lotus Desbiolles
    Michael Alexander Fischer
    Christoph Karlo
    Stephan Baumueller
    Hans Scheffel
    Hatem Alkadhi
    Sebastian Leschka
    European Radiology, 2010, 20 : 2565 - 2571
  • [22] High-pitch dual-source CT coronary angiography: systolic data acquisition at high heart rates
    Goetti, Robert
    Feuchtner, Gudrun
    Stolzmann, Paul
    Desbiolles, Lotus
    Fischer, Michael Alexander
    Karlo, Christoph
    Baumueller, Stephan
    Scheffel, Hans
    Alkadhi, Hatem
    Leschka, Sebastian
    EUROPEAN RADIOLOGY, 2010, 20 (11) : 2565 - 2571
  • [23] Feasibility of multi-contrast imaging on dual-source photon counting detector (PCD) CT: An initial phantom study
    Tao, Shengzhen
    Rajendran, Kishore
    McCollough, Cynthia H.
    Leng, Shuai
    MEDICAL PHYSICS, 2019, 46 (09) : 4105 - 4115
  • [24] Performance of turbo high-pitch dual-source CT for coronary CT angiography: first ex vivo and patient experience
    Morsbach, Fabian
    Gordic, Sonja
    Desbiolles, Lotus
    Husarik, Daniela
    Frauenfelder, Thomas
    Schmidt, Bernhard
    Allmendinger, Thomas
    Wildermuth, Simon
    Alkadhi, Hatem
    Leschka, Sebastian
    EUROPEAN RADIOLOGY, 2014, 24 (08) : 1889 - 1895
  • [25] Multi-Energy CT with Triple X-ray Beams and Photon-Counting-Detector CT for Simultaneous Imaging of Two Contrast Agents: An Experimental Comparison
    Ren, Liqiang
    McCollough, Cynthia H.
    Yu, Lifeng
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [26] Ultra-low dose contrast CT pulmonary angiography in oncology patients using a high-pitch helical dual-source technology
    Rajiah, Prabhakar
    Ciancibello, Leslie
    Novak, Ronald
    Sposato, Jennifer
    Landeras, Luis
    Gilkeson, Robert
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2019, 25 (03) : 195 - 203
  • [27] High-Pitch 128-Slice Dual-Source CT for the Assessment of Coronary Stents in a Phantom Model
    Donati, Olivio F.
    Burg, Matthias C.
    Desbiolles, Lotus
    Karlo, Christoph
    Stolzmann, Paul
    Bunck, Alexander
    Baumueller, Stephan
    Marincek, Borut
    Seifarth, Harald
    Alkadhi, Hatem
    Leschka, Sebastian
    Maintz, David
    ACADEMIC RADIOLOGY, 2010, 17 (11) : 1366 - 1374
  • [28] Performance of turbo high-pitch dual-source CT for coronary CT angiography: first ex vivo and patient experience
    Fabian Morsbach
    Sonja Gordic
    Lotus Desbiolles
    Daniela Husarik
    Thomas Frauenfelder
    Bernhard Schmidt
    Thomas Allmendinger
    Simon Wildermuth
    Hatem Alkadhi
    Sebastian Leschka
    European Radiology, 2014, 24 : 1889 - 1895
  • [29] Dual-source photon counting detector CT with a tin filter: a phantom study on iodine quantification performance
    Tao, Ashley
    Huang, Richard
    Tao, Shengzhen
    Michalak, Gregory J.
    McCollough, Cynthia H.
    Leng, Shuai
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (11):
  • [30] Novel single-source high-pitch protocol for CT angiography of the aorta: comparison to high-pitch dual-source protocol in the context of TAVI planning
    Bernhard Bischoff
    Felix G. Meinel
    Maximilian Reiser
    Hans-Christoph Becker
    The International Journal of Cardiovascular Imaging, 2013, 29 : 1159 - 1165