Multi-Energy CT with Triple X-ray Beams and Photon-Counting-Detector CT for Simultaneous Imaging of Two Contrast Agents: An Experimental Comparison

被引:0
|
作者
Ren, Liqiang [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN 55902 USA
基金
美国国家卫生研究院;
关键词
Multi-energy CT (MECT); Photon-counting-detector CT (PCD-CT); Dual-energy CT (DECT); Material Decomposition; Multiple Contrast Agents; DUAL-ENERGY CT; IODINE;
D O I
10.1117/12.2513460
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Multi-energy CT (MECT) enabled by energy-resolved photon-counting-detector CT (PCD-CT) is promising for material specific imaging with multiple contrast agents. However, non-idealities of the PCD such as pulse pileup, K-edge escape, and charge sharing may degrade the spectral performance. To perform MECT, an alternative approach was proposed by extending a "Twin Beam" design to a dual-source CT scanner with energy-integrating-detector (EID) by operating one or both sources in the "Twin Beam" mode to acquire three (triple beam configuration) or four (quadruple-beam configuration) distinct X-ray beam measurements. Previous computer simulation studies demonstrated that the image quality and dose efficiency of the triple-beam configuration were comparable to that in PCD-CT for a three-material decomposition task involving iodine, bismuth, and water. The purpose of this work is to experimentally validate the proposed triple-beam MECT technique in comparison with PCD-CT. To mimic the dual-source triple-beam acquisition, two separate scans, one at 80 kV and the other at 120 kV operated in the "Twin Beam" mode, were performed on a single-source CT scanner. Two potential clinical applications of MECT for multiple contrast agents were investigated: iodine/gadolinium for biphasic liver imaging and iodine/bismuth for small bowel imaging. The results indicate that the imaging performance of the EID-based MECT may be comparable to that on the current PCD-CT platform for both the iodine/gadolinium and the iodine/bismuth material decomposition tasks.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Radiation dose efficiency of multi-energy photon-counting-detector CT for dual-contrast imaging
    Ren, Liqiang
    Rajendran, Kishore
    McCollough, Cynthia H.
    Yu, Lifeng
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (24):
  • [2] Adaptive Multi-Energy Local Mean Filtering for Photon-Counting-Detector Based Multi-Energy CT
    Tao, S.
    Rajendran, K.
    Ferrero, A.
    Zhou, W.
    McCollough, C.
    Leng, S.
    MEDICAL PHYSICS, 2018, 45 (06) : E672 - E672
  • [3] Multi-energy CT with Triple X-ray Beams: A Feasibility Animal Study
    Ren, Liqiang
    Rajendran, Kishore
    Fletcher, Joel G.
    Allmendinger, Thomas
    Halaweish, Ahmed
    Schmidt, Bernhard
    Flohr, Thomas
    McCollough, Cynthia H.
    Yu, Lifeng
    MEDICAL IMAGING 2020: PHYSICS OF MEDICAL IMAGING, 2020, 11312
  • [4] Detector technologies for photon counting/energy resolving x-ray and CT imaging
    Shikhaliev, P. M.
    MEDICAL PHYSICS, 2007, 34 (06) : 2554 - 2555
  • [5] Multi-Contrast Imaging on Dual-Source Photon-Counting-Detector (PCD) CT
    Tao, Shengzhen
    Rajendran, Kishore
    McCollough, Cynthia H.
    Leng, Shuai
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [6] Comparison of a Photon-Counting-Detector CT with an Energy-Integrating-Detector CT for Temporal Bone Imaging: A Cadaveric Study
    Zhou, W.
    Lane, J. I.
    Carlson, M. L.
    Bruesewitz, M. R.
    Witte, R. J.
    Koeller, K. K.
    Eckel, L. J.
    Carter, R. E.
    McCollough, C. H.
    Leng, S.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2018, 39 (09) : 1733 - 1738
  • [7] Characterization of single- and multi-energy CT performance of an oral dark borosilicate contrast media using a clinical photon-counting-detector CT platform
    Ren, Liqiang
    Sun, Yuxin
    Yeh, Benjamin
    Marsh, Jeffrey F.
    Winfree, Timothy N.
    Burke, Kristin A.
    Rajendran, Kishore
    Mccollough, Cynthia H.
    Mileto, Achille
    Fletcher, Joel G.
    Leng, Shuai
    MEDICAL PHYSICS, 2023, 50 (11) : 6779 - 6788
  • [8] Dual-Source Multi-Energy CT with Triple or Quadruple X-ray Beams
    Yu, Lifeng
    Li, Zhoubo
    Leng, Shuai
    McCollough, Cynthia H.
    MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
  • [9] Simultaneous high-pitch multi-energy CT pulmonary angiography using a dual-source photon-counting-detector CT: A phantom experiment
    Mihailovic, Jelena M.
    Bruesewitz, Michael R.
    Swicklik, Joseph R.
    Yalon, Mariana
    Rajiah, Prabhakar S.
    Fletcher, Joel G.
    McCollough, Cynthia H.
    Yu, Lifeng
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024,
  • [10] The Effects of Spectral Distortion on Multi-Energy X-ray Imaging Based on Photon Counting Detector
    Cho, Hyo-Min
    Choi, Yu-Na
    Lee, Seung-Wan
    Ryu, Hyun-Ju
    Lee, Young-Jin
    Kim, Hee-Joung
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3001 - 3003