Simultaneous high-pitch multi-energy CT pulmonary angiography using a dual-source photon-counting-detector CT: A phantom experiment

被引:0
|
作者
Mihailovic, Jelena M. [1 ]
Bruesewitz, Michael R. [1 ]
Swicklik, Joseph R. [1 ]
Yalon, Mariana [1 ]
Rajiah, Prabhakar S. [1 ]
Fletcher, Joel G. [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
关键词
dual source computed tomography; photon-counting-detector computed tomography; pulmonary embolism; PERFUSED BLOOD-VOLUME; ENERGY CT; EMBOLISM;
D O I
10.1002/acm2.14496
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A dual-source CT system can be operated in a high-pitch helical mode to provide a temporal resolution of 66 ms, which reduces motion artifacts in CT pulmonary angiography (CTPA). It can also be operated in a multi-energy (ME) mode to provide iodine maps, beneficial in the evaluation of pulmonary embolism (PE). No energy-integrating detector (EID) CT can perform simultaneous ME and high-pitch acquisition. This phantom study aimed to evaluate the ability of a photon-counting-detector (PCD) CT to perform simultaneous high-pitch and ME imaging for CTPA. Methods: A motion phantom was used to mimic the respiratory motion. Two tubes filled with iodine with intravascular thrombus mimicked by injecting glue within the tubes were placed along with 5, 10, and 15 mg/mL iodine samples, on a motion phantom at 20 and 30 revolutions per minute. Separate high-pitch and ME EID-CT scans and a single high-pitch ME PCD scan were acquired and virtual monoenergetic images and iodine maps reconstructed. Percent thrombus occlusion was measured and compared between static and moving images. Results: When there was motion, EID-CT ME suffered from significant motion artifacts. The measured iodine concentrations with PCD-CT in high-pitch ME were more stable when there was a motion, with a lower standard deviation than EID-CT in ME mode. The estimated percent thrombus occlusion dropped significantly with applied motion on EID-CT, while PCD-CT high-pitch ME mode showed good agreement between measurements on static or moving images. Conclusion: PCD-CT with combined ME and high-pitch mode facilitates simultaneous accurate iodine quantification and assessment of intravascular occlusion.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] High-pitch, high temporal resolution, multi-energy cardiac imaging on a dual-source photon-counting-detector CT
    Ahmed, Zaki
    Campeau, David
    Gong, Hao
    Rajendran, Kishore
    Rajiah, Prabhakar
    McCollough, Cynthia
    Leng, Shuai
    [J]. MEDICAL PHYSICS, 2023, 50 (03) : 1428 - 1435
  • [2] Evaluation of Geometric Accuracy and Iodine Quantification of High-Pitch Multi-Energy Cardiac Imaging with Dual-Source Photon-Counting Detector CT
    Ahmed, Z.
    Campeau, D.
    Gong, H.
    Rajendran, K.
    Rajiah, P.
    McCollough, C.
    Leng, S.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E592 - E592
  • [3] Multi-energy CT imaging for large patients using dual-source photon-counting detector CT
    Tao, Shengzhen
    Marsh, Jeffrey F.
    Tao, Ashley
    Michalak, Greg J.
    Rajendran, Kishore
    McCollough, Cynthia H.
    Leng, Shuai
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (17):
  • [4] High-Pitch Multienergy Coronary CT Angiography in Dual-Source Photon-Counting Detector CT Scanner at Low Iodinated Contrast Dose
    Rajiah, Prabhakar Shantha
    Dunning, Chelsea A. S.
    Rajendran, Kishore
    Tandon, Yasmeen K.
    Ahmed, Zaki
    Larson, Nicholas B.
    Collins, Jeremy D.
    Thorne, Jamison
    Williamson, Eric
    Fletcher, Joel G.
    McCollough, Cynthia
    Leng, Shuai
    [J]. INVESTIGATIVE RADIOLOGY, 2023, 58 (09) : 681 - 690
  • [5] Multi-Contrast Imaging on Dual-Source Photon-Counting-Detector (PCD) CT
    Tao, Shengzhen
    Rajendran, Kishore
    McCollough, Cynthia H.
    Leng, Shuai
    [J]. MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [6] Image quality of lung perfusion with photon-counting-detector CT: comparison with dual-source, dual-energy CT
    Remy-Jardin, Martine
    Guiffault, Lucas
    Oufriche, Idir
    Duhamel, Alain
    Flohr, Thomas
    Schmidt, Bernhard
    Remy, Jacques
    [J]. EUROPEAN RADIOLOGY, 2024,
  • [7] Adaptive Multi-Energy Local Mean Filtering for Photon-Counting-Detector Based Multi-Energy CT
    Tao, S.
    Rajendran, K.
    Ferrero, A.
    Zhou, W.
    McCollough, C.
    Leng, S.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E672 - E672
  • [8] Radiation dose efficiency of multi-energy photon-counting-detector CT for dual-contrast imaging
    Ren, Liqiang
    Rajendran, Kishore
    McCollough, Cynthia H.
    Yu, Lifeng
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (24):
  • [9] High-Pitch CT Pulmonary Angiography in Third Generation Dual-Source CT: Image Quality in an Unselected Patient Population
    Sabel, Bastian O.
    Buric, Kristijan
    Karara, Nora
    Thierfelder, Kolja M.
    Dinkel, Julien
    Sommer, Wieland H.
    Meinel, Felix G.
    [J]. PLOS ONE, 2016, 11 (02):
  • [10] Effect of High-Pitch Dual-Source CT to Compensate Motion Artifacts: A Phantom Study
    Farshad-Amacker, Nadia A.
    Alkadhi, Hatem
    Leschka, Sebastian
    Frauenfelder, Thomas
    [J]. ACADEMIC RADIOLOGY, 2013, 20 (10) : 1234 - 1239