High-pitch, high temporal resolution, multi-energy cardiac imaging on a dual-source photon-counting-detector CT

被引:9
|
作者
Ahmed, Zaki [1 ]
Campeau, David [2 ]
Gong, Hao [1 ]
Rajendran, Kishore [1 ]
Rajiah, Prabhakar [1 ]
McCollough, Cynthia
Leng, Shuai [3 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
[2] Mayo Clin, Dept Physiol & Biomed Engn, Rochester, MN USA
[3] 200 First St SW, Rochester, MN 55905 USA
基金
美国国家卫生研究院;
关键词
cardiac CT; computed tomography; dual energy CT; iodine quantification; material decomposition; motion artifact; photon counting detector; COMPUTED-TOMOGRAPHY; ENERGY CT; ANGIOGRAPHY; RECONSTRUCTION; QUALITY;
D O I
10.1002/mp.16124
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveTo measure the accuracy of material decomposition using a dual-source photon-counting-detector (DS-PCD) CT operated in the high-pitch helical scanning mode and compare the results against dual-source energy-integrating-detector (DS-EID) CT, which requires use of a low-pitch value in dual-energy mode. MethodsA DS-PCD CT and a DS-EID CT were used to scan a cardiac motion phantom consisting of a 3-mm diameter iodine cylinder. Iodine maps were reconstructed using DS-PCD in high-pitch mode and DS-EID in low-pitch mode. Image-based circularity, diameter, and iodine concentration of the iodine cylinder were calculated and compared between the two scanners. With institutional review board approval, in vivo exams were performed with the DS-PCD CT in high-pitch mode. Images were qualitatively compared against patients with similar heart rates that were scanned with DS-EID CT in low-pitch dual-energy mode. ResultsOn iodine maps, the mean circularity was 0.97 +/- 0.02 with DS-PCD in high-pitch mode and 0.95 +/- 0.06 with DS-EID in low-pitch mode. The mean diameter was 2.9 +/- 0.2 mm with DS-PCD and 3.1 +/- 0.2 mm with DS-EID, both of which are close to the 3 mm ground truth. For DS-PCD, the mean iodine concentration was 9.6 +/- 0.8 mg/ml and this was consistent with the 9.4 +/- 0.6 mg/ml value obtained with the cardiac motion disabled. For DS-EID, the concentration was 12.7 +/- 1.2 mg/ml with motion enabled and 11.7 +/- 0.5 mg/ml disabled. The background noise in the iodine maps was 15.1 HU with DS-PCD and 14.4 HU with DS-EID, whereas the volume CT dose index (CTDIvol) was 3 mGy with DS-PCD and 11 mGy with DS-EID. On comparison of six patients (three on PCD, three on EID) with similar heart rates, DS-PCD provided iodine maps with well-defined coronaries even at a high heart rate of 86 beats per minute. Meanwhile, there were substantial motion artifacts in iodine maps obtained with DS-EID for patients with similar heart rates. ConclusionIn a cardiac motion phantom, DS-PCD CT can perform accurate material decomposition in high-pitch mode, providing iodine maps with excellent geometric accuracy and robustness to motion at approximately 38% of the dose for similar noise as DS-EID CT.
引用
下载
收藏
页码:1428 / 1435
页数:8
相关论文
共 50 条
  • [1] Simultaneous high-pitch multi-energy CT pulmonary angiography using a dual-source photon-counting-detector CT: A phantom experiment
    Mihailovic, Jelena M.
    Bruesewitz, Michael R.
    Swicklik, Joseph R.
    Yalon, Mariana
    Rajiah, Prabhakar S.
    Fletcher, Joel G.
    McCollough, Cynthia H.
    Yu, Lifeng
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2024,
  • [2] Evaluation of Geometric Accuracy and Iodine Quantification of High-Pitch Multi-Energy Cardiac Imaging with Dual-Source Photon-Counting Detector CT
    Ahmed, Z.
    Campeau, D.
    Gong, H.
    Rajendran, K.
    Rajiah, P.
    McCollough, C.
    Leng, S.
    MEDICAL PHYSICS, 2022, 49 (06) : E592 - E592
  • [3] Multi-Contrast Imaging on Dual-Source Photon-Counting-Detector (PCD) CT
    Tao, Shengzhen
    Rajendran, Kishore
    McCollough, Cynthia H.
    Leng, Shuai
    MEDICAL IMAGING 2019: PHYSICS OF MEDICAL IMAGING, 2019, 10948
  • [4] Multi-energy CT imaging for large patients using dual-source photon-counting detector CT
    Tao, Shengzhen
    Marsh, Jeffrey F.
    Tao, Ashley
    Michalak, Greg J.
    Rajendran, Kishore
    McCollough, Cynthia H.
    Leng, Shuai
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (17):
  • [5] Radiation dose efficiency of multi-energy photon-counting-detector CT for dual-contrast imaging
    Ren, Liqiang
    Rajendran, Kishore
    McCollough, Cynthia H.
    Yu, Lifeng
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (24):
  • [6] High-Pitch Multienergy Coronary CT Angiography in Dual-Source Photon-Counting Detector CT Scanner at Low Iodinated Contrast Dose
    Rajiah, Prabhakar Shantha
    Dunning, Chelsea A. S.
    Rajendran, Kishore
    Tandon, Yasmeen K.
    Ahmed, Zaki
    Larson, Nicholas B.
    Collins, Jeremy D.
    Thorne, Jamison
    Williamson, Eric
    Fletcher, Joel G.
    McCollough, Cynthia
    Leng, Shuai
    INVESTIGATIVE RADIOLOGY, 2023, 58 (09) : 681 - 690
  • [7] Image quality of lung perfusion with photon-counting-detector CT: comparison with dual-source, dual-energy CT
    Remy-Jardin, Martine
    Guiffault, Lucas
    Oufriche, Idir
    Duhamel, Alain
    Flohr, Thomas
    Schmidt, Bernhard
    Remy, Jacques
    EUROPEAN RADIOLOGY, 2024, : 7831 - 7844
  • [8] Adaptive Multi-Energy Local Mean Filtering for Photon-Counting-Detector Based Multi-Energy CT
    Tao, S.
    Rajendran, K.
    Ferrero, A.
    Zhou, W.
    McCollough, C.
    Leng, S.
    MEDICAL PHYSICS, 2018, 45 (06) : E672 - E672
  • [9] Performance of iodine quantification through high-pitch dual-source photon-counting CT: a phantom study
    Peng Liu
    Shanshui Zhou
    Haipeng Dong
    Jiqiang Li
    Zhihan Xu
    Shushen Lin
    Wenjie Yang
    Fuhua Yan
    Le Qin
    Japanese Journal of Radiology, 2025, 43 (2) : 309 - 318
  • [10] The feasibility of high-pitch acquisition protocol for imaging of the pediatric abdomen by dual-source CT
    Eray Atlı
    Erhan Akpınar
    Emre Ünal
    Berna Oğuz Sayan
    Mithat Haliloğlu
    Japanese Journal of Radiology, 2018, 36 : 437 - 443