Simultaneous high-pitch multi-energy CT pulmonary angiography using a dual-source photon-counting-detector CT: A phantom experiment

被引:0
|
作者
Mihailovic, Jelena M. [1 ]
Bruesewitz, Michael R. [1 ]
Swicklik, Joseph R. [1 ]
Yalon, Mariana [1 ]
Rajiah, Prabhakar S. [1 ]
Fletcher, Joel G. [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, 200 First St SW, Rochester, MN 55905 USA
关键词
dual source computed tomography; photon-counting-detector computed tomography; pulmonary embolism; PERFUSED BLOOD-VOLUME; ENERGY CT; EMBOLISM;
D O I
10.1002/acm2.14496
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A dual-source CT system can be operated in a high-pitch helical mode to provide a temporal resolution of 66 ms, which reduces motion artifacts in CT pulmonary angiography (CTPA). It can also be operated in a multi-energy (ME) mode to provide iodine maps, beneficial in the evaluation of pulmonary embolism (PE). No energy-integrating detector (EID) CT can perform simultaneous ME and high-pitch acquisition. This phantom study aimed to evaluate the ability of a photon-counting-detector (PCD) CT to perform simultaneous high-pitch and ME imaging for CTPA. Methods: A motion phantom was used to mimic the respiratory motion. Two tubes filled with iodine with intravascular thrombus mimicked by injecting glue within the tubes were placed along with 5, 10, and 15 mg/mL iodine samples, on a motion phantom at 20 and 30 revolutions per minute. Separate high-pitch and ME EID-CT scans and a single high-pitch ME PCD scan were acquired and virtual monoenergetic images and iodine maps reconstructed. Percent thrombus occlusion was measured and compared between static and moving images. Results: When there was motion, EID-CT ME suffered from significant motion artifacts. The measured iodine concentrations with PCD-CT in high-pitch ME were more stable when there was a motion, with a lower standard deviation than EID-CT in ME mode. The estimated percent thrombus occlusion dropped significantly with applied motion on EID-CT, while PCD-CT high-pitch ME mode showed good agreement between measurements on static or moving images. Conclusion: PCD-CT with combined ME and high-pitch mode facilitates simultaneous accurate iodine quantification and assessment of intravascular occlusion.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Thoraco-abdominal high-pitch dual-source CT angiography: Experimental evaluation of injection protocols with an anatomical human vascular phantom
    Puippe, Gilbert D.
    Winklehner, Anna
    Hasenclever, Peter
    Plass, Andre
    Frauenfelder, Thomas
    Baumueller, Stephan
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (10) : 2592 - 2596
  • [42] Reproducibility of coronary artery calcium quantification on dual-source CT and dual-source photon-counting CT: a dynamic phantom study
    van der Werf, Niels R.
    Booij, Ronald
    Greuter, Marcel J. W.
    Bos, Daniel
    van der Lugt, A.
    Budde, R. P. J.
    van Straten, Marcel
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2022, 38 (07): : 1613 - 1619
  • [43] Characterization of single- and multi-energy CT performance of an oral dark borosilicate contrast media using a clinical photon-counting-detector CT platform
    Ren, Liqiang
    Sun, Yuxin
    Yeh, Benjamin
    Marsh, Jeffrey F.
    Winfree, Timothy N.
    Burke, Kristin A.
    Rajendran, Kishore
    Mccollough, Cynthia H.
    Mileto, Achille
    Fletcher, Joel G.
    Leng, Shuai
    MEDICAL PHYSICS, 2023, 50 (11) : 6779 - 6788
  • [44] Performance of calcium quantifications on low-dose photon-counting detector CT with high-pitch: A phantom study
    Zhou, Shanshui
    Liu, Peng
    Dong, Haipeng
    Li, Jiqiang
    Xu, Zhihan
    Schmidt, Bernhard
    Lin, Shushen
    Yang, Wenjie
    Yan, Fuhua
    Qin, Le
    HELIYON, 2024, 10 (12)
  • [45] High-pitch Dual-source Computed Tomography Pulmonary Angiography in Freely Breathing Patients
    Bauer, Ralf W.
    Schell, Boris
    Beeres, Martin
    Wichmann, Julian L.
    Bodelle, Boris
    Vogl, Thomas J.
    Kerl, Josef Matthias
    JOURNAL OF THORACIC IMAGING, 2012, 27 (06) : 376 - 381
  • [46] Approaches, advantages, and challenges to photon counting detector and multi-energy CT
    Toia, Giuseppe V.
    Mileto, Achille
    Borhani, Amir A.
    Chen, Guang-Hong
    Ren, Liqiang
    Uyeda, Jennifer W.
    Marin, Daniele
    ABDOMINAL RADIOLOGY, 2024, 49 (09) : 3251 - 3260
  • [47] Iterative reconstruction improves detection of in-stent restenosis by high-pitch dual-source coronary CT angiography
    Yang, Junjie
    Yang, Xiaobo
    De Cecco, Carlo N.
    Duguay, Taylor M.
    Chen, Zhiye
    Tesche, Christian
    Schoepf, U. Joseph
    Chen, Yundai
    SCIENTIFIC REPORTS, 2017, 7
  • [48] The ascending aortic image quality and the whole aortic radiation dose of high-pitch dual-source CT angiography
    Ying Liu
    Jian Xu
    Jian Li
    Jing Ren
    Hongtao Liu
    Junqing Xu
    Mengqi Wei
    Yuewen Hao
    Minwen Zheng
    Journal of Cardiothoracic Surgery, 8
  • [49] Iterative reconstruction improves detection of in-stent restenosis by high-pitch dual-source coronary CT angiography
    Junjie Yang
    Xiaobo Yang
    Carlo N. De Cecco
    Taylor M. Duguay
    Zhiye Chen
    Christian Tesche
    U. Joseph Schoepf
    Yundai Chen
    Scientific Reports, 7
  • [50] The ascending aortic image quality and the whole aortic radiation dose of high-pitch dual-source CT angiography
    Liu, Ying
    Xu, Jian
    Li, Jian
    Ren, Jing
    Liu, Hongtao
    Xu, Junqing
    Wei, Mengqi
    Hao, Yuewen
    Zheng, Minwen
    JOURNAL OF CARDIOTHORACIC SURGERY, 2013, 8