Performance of calcium quantifications on low-dose photon-counting detector CT with high-pitch: A phantom study

被引:0
|
作者
Zhou, Shanshui [1 ,2 ]
Liu, Peng [1 ]
Dong, Haipeng [1 ]
Li, Jiqiang [1 ]
Xu, Zhihan [3 ]
Schmidt, Bernhard [4 ]
Lin, Shushen [3 ]
Yang, Wenjie [1 ]
Yan, Fuhua [1 ,2 ]
Qin, Le [1 ]
机构
[1] Shanghai Jiao Tong Univ, Sch Med, Ruijin Hosp, Dept Radiol, 197 Ruijin Er Rd, Shanghai 200025, Peoples R China
[2] Shanghai Jiao Tong Univ, Sch Med, Coll Hlth Sci & Technol, Fac Med Imaging Technol, 150 Ruijin Er Rd, Shanghai 200025, Peoples R China
[3] Siemens Healthineers, 399 West Haiyang Rd, Shanghai 200126, Peoples R China
[4] Siemens Healthineers, Siemensstr 3, D-91301 Erlangen, Germany
关键词
Computed tomography; Photon-counting CT; High-pitch; Low dose; Calcium quantification; DUAL-ENERGY CT; CORONARY-ARTERY CALCIUM; BONE-MINERAL DENSITY; COMPUTED-TOMOGRAPHY; ATTENUATION; ACCURACY; OSTEOPOROSIS; ANGIOGRAPHY; DXA;
D O I
10.1016/j.heliyon.2024.e32819
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: To evaluate the performance of calcium quantification on photon-counting detector CT (PCD-CT) with high-pitch at low radiation doses compared to third-generation dual-source energy-integrating detector CT (EID-CT). Materials and methods: The phantom with three calcium inserts (50, 100, and 300 mg of calcium per milliliter), with and without the elliptical outer layer, was evaluated using high-pitch (3.2) and standard pitch (0.8) on PCD-CT, and standard pitch on EID-CT. Scans were performed with different tube voltages (PCD-CT: 120 and 140 kilo-voltage peak [kVp]; EID-CT: 70/Sn150 and 100/Sn150 kVp) and four radiation doses (1, 3, 5, and, 10 milli-Gray [mGy]). Utilizing the true calcium concentrations (CCtrue) of the phantom as the gold standard references, regression equations for each kVp setting were formulated to convert CT attenuations (CaCT) into measured calcium concentrations (CCm). The correlation analysis between CaCT and CCtrue was performed. The percentage absolute bias (PAB) was calculated from the differences between CCm and CCtrue and used to analyze the effects of scanning parameters on calcium quantification accuracy. Results: A strong correlation was found between CaCT and CCtrue on PCD-CT (r > 0.99) and EID-CT (r > 0.98). For high- and standard-pitch scans on PCD-CT, the accuracy of calcium quantification is comparable (p = 0.615): the median (interquartile range [IQR]) of PAB was 5.59% (2.79%-8.31%) and 4.87 % (2.62%-8.01%), respectively. The PAB median (IQR) was 7.43% (3.77%-11.75%) for EID-CT. The calcium quantification accuracy of PCD-CT is superior to EID-CT at the large phantom (5.46% [2.68%-9.55%] versus 9.01% [6.22%-12.74%]), and at the radiation dose of 1 mGy (4.43% [2.08%-8.59%] versus 13.89% [8.93%-23.09%]) and 3 mGy (4.61% [2.75%-6.51%] versus 9.97% [5.17%-14.41%]), all p < 0.001. Conclusions: Calcium quantification using low-dose PCD-CT with high-pitch scanning is feasible and accurate, and superior to EID-CT.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] High-Pitch Multienergy Coronary CT Angiography in Dual-Source Photon-Counting Detector CT Scanner at Low Iodinated Contrast Dose
    Rajiah, Prabhakar Shantha
    Dunning, Chelsea A. S.
    Rajendran, Kishore
    Tandon, Yasmeen K.
    Ahmed, Zaki
    Larson, Nicholas B.
    Collins, Jeremy D.
    Thorne, Jamison
    Williamson, Eric
    Fletcher, Joel G.
    McCollough, Cynthia
    Leng, Shuai
    [J]. INVESTIGATIVE RADIOLOGY, 2023, 58 (09) : 681 - 690
  • [2] Improved coronary calcium detection and quantification with low-dose full field-of-view photon-counting CT: a phantom study
    N. R. van der Werf
    P. A. Rodesch
    S. Si-Mohamed
    R. W. van Hamersvelt
    M. J. W. Greuter
    T. Leiner
    L. Boussel
    M. J. Willemink
    P. Douek
    [J]. European Radiology, 2022, 32 : 3447 - 3457
  • [3] Improved coronary calcium detection and quantification with low-dose full field-of-view photon-counting CT: a phantom study
    van der Werf, N. R.
    Rodesch, P. A.
    Si-Mohamed, S.
    van Hamersvelt, R. W.
    Greuter, M. J. W.
    Leiner, T.
    Boussel, L.
    Willemink, M. J.
    Douek, P.
    [J]. EUROPEAN RADIOLOGY, 2022, 32 (05) : 3447 - 3457
  • [4] Image Quality and Radiation Dose of CTPA With Iodine Maps: A Prospective Randomized Study of High-Pitch Mode Photon-Counting Detector CT Versus Energy-Integrating Detector CT
    Pannenbecker, Pauline
    Heidenreich, Julius F.
    Grunz, Jan-Peter
    Huflage, Henner
    Gruschwitz, Philipp
    Patzer, Theresa S.
    Feldle, Philipp
    Bley, Thorsten A.
    Petritsch, Bernhard
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2024, 222 (02)
  • [5] Detectability and Volumetric Accuracy of Pulmonary Nodules in Low-Dose Photon-Counting Detector Computed Tomography: An Anthropomorphic Phantom Study
    Hop, Joost F.
    Walstra, Anna N. H.
    Pelgrim, Gert-Jan
    Xie, Xueqian
    Panneman, Noor A.
    Schurink, Niels W.
    Faby, Sebastian
    van Straten, Marcel
    de Bock, Geertruida H.
    Vliegenthart, Rozemarijn
    Greuter, Marcel J. W.
    [J]. DIAGNOSTICS, 2023, 13 (22)
  • [6] Optimization of the Reconstruction Settings for Low-Dose Ultra-High-Resolution Photon-Counting Detector CT of the Lungs
    Graafen, Dirk
    Halfmann, Moritz C.
    Emrich, Tilman
    Yang, Yang
    Kreuter, Michael
    Dueber, Christoph
    Kloeckner, Roman
    Mueller, Lukas
    Jorg, Tobias
    [J]. DIAGNOSTICS, 2023, 13 (23)
  • [7] Low-dose lung cancer screening with photon-counting CT: a feasibility study
    Symons, Rolf
    Cork, Tyler E.
    Sahbaee, Pooyan
    Fuld, Matthew K.
    Kappler, Steffen
    Folio, Les R.
    Bluemke, David A.
    Pourmorteza, Amir
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (01): : 202 - 213
  • [8] Quantum Iterative Reconstruction for Low-Dose Ultra-High-Resolution Photon-Counting Detector CT of the Lung
    Sartoretti, Thomas
    Racine, Damien
    Mergen, Victor
    Jungblut, Lisa
    Monnin, Pascal
    Flohr, Thomas G.
    Martini, Katharina
    Frauenfelder, Thomas
    Alkadhi, Hatem
    Euler, Andre
    [J]. DIAGNOSTICS, 2022, 12 (02)
  • [9] Myocardial CT perfusion with a high-pitch low-dose protocol
    Hausleiter, J.
    Gramer, B.
    Meyer, T.
    Bischoff, B.
    Hadamitzky, M.
    Spiegel, S.
    Ardelean, C.
    Haschka, M.
    Martinoff, S.
    Schoemig, A.
    [J]. EUROPEAN HEART JOURNAL, 2010, 31 : 581 - 581
  • [10] Low-dose performance of a whole-body research photon-counting CT scanner
    Yu, Zhicong
    Leng, Shuai
    Kappler, Steffen
    Hahn, Katharina
    Li, Zhoubo
    Halaweish, Ahmed F.
    Henning, Andre
    Ritman, Erik L.
    McCollough, Cynthia H.
    [J]. MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783