High-Pitch 128-Slice Dual-Source CT for the Assessment of Coronary Stents in a Phantom Model

被引:17
|
作者
Donati, Olivio F. [1 ]
Burg, Matthias C. [2 ]
Desbiolles, Lotus [1 ]
Karlo, Christoph [1 ]
Stolzmann, Paul [1 ,3 ]
Bunck, Alexander [2 ]
Baumueller, Stephan [1 ]
Marincek, Borut [1 ]
Seifarth, Harald [2 ]
Alkadhi, Hatem [1 ,3 ]
Leschka, Sebastian [1 ]
Maintz, David [2 ]
机构
[1] Univ Zurich Hosp, Inst Diagnost Radiol, CH-8091 Zurich, Switzerland
[2] Univ Munster, Dept Clin Radiol, D-4400 Munster, Germany
[3] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Cardiac MR PET CT Program, Boston, MA USA
关键词
Stent; computed tomography; high pitch; coronary artery disease; dual source; PACLITAXEL-ELUTING STENT; IN-VITRO EVALUATION; BARE-METAL STENT; COMPUTED-TOMOGRAPHY; ANGIOGRAPHY; PATENCY; RESTENOSIS;
D O I
10.1016/j.acra.2010.06.017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives: To evaluate the quality of stent lumen delineation using dual-source computed tomography (DSCT) in the standard-pitch mode (SP) as compared to the high-pitch mode (HP) in a phantom study. Materials and Methods: Forty different coronary stents placed in plastic tubes filled with contrast agent were imaged with a second generation DSCT system in a SP (pitch 0.23) and HP (pitch 3.4) mode in orientations of 0, 450, and 90 relative to the z-axis. Two observers independently measured the in-stent lumen and the attenuation values in the center of the stents. The artificial lumen narrowing (ALN) was calculated using the measured in-stent lumen and the nominal diameter of the plastic tube. Results: lnterobserver correlation was excellent for in-stent lumen (0.86) and attenuation measurements (0.91). There was no significant difference neither for ALN (SP: 54.7-62.8%; HP: 55.8-64.0%) nor attenuation (SP: 356-395 Hounsfield units [HU]; HP: 352-384 HU) between SP and HP mode. For both modes, the orientation of the stent relative to the z-axis significantly affected ALN and attenuation (each P < .001). CT volume dose index was significantly lower using HP mode as compared to SP mode (P < .001). Conclusion: The HP mode in DSCT provides visualization of the coronary in-stent lumen comparable to that measured in SP mode while reducing applied radiation dose in a stationary phantom model.
引用
收藏
页码:1366 / 1374
页数:9
相关论文
共 50 条
  • [1] Accuracy of 128-slice dual-source CT using high-pitch spiral mode for the assessment of coronary stents
    Yang, X.
    Chen, C. Y. D.
    Yang, Y. J. J.
    He, H. B.
    Jing, Wang
    [J]. EUROPEAN HEART JOURNAL, 2013, 34 : 844 - 844
  • [2] Accuracy of 128-slice dual-source CT using high-pitch spiral mode for the assessment of coronary stents: First in vivo experience
    Yang Xia
    Yang Junjie
    Zhou Ying
    He Bai
    Wang Qi
    Jin Qinhua
    Chen Yundai
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (04) : 617 - 622
  • [3] Coronary artery stent imaging with 128-slice dual-source CT using high-pitch spiral acquisition in a cardiac phantom: comparison with the sequential and low-pitch spiral mode
    Wolf, Florian
    Leschka, Sebastian
    Loewe, Christian
    Homolka, Peter
    Plank, Christina
    Schernthaner, Ruediger
    Bercaczy, Dominik
    Goetti, Robert
    Lammer, Johannes
    Friedrich, Guy
    Marincek, Borut
    Alkadhi, Hatem
    Feuchtner, Gudrun
    [J]. EUROPEAN RADIOLOGY, 2010, 20 (09) : 2084 - 2091
  • [4] DIAGNOSTIC ACCURACY OF 128-SLICE DUAL-SOURCE CT USING HIGH-PITCH SPIRAL MODE IN ASSESSMENT OF CORONARY ARTERY STENT IMAGING COMPARISON WITH INVASIVE CORONARY ANGIOGRAPHY
    Yang Xia
    Chen Yundai
    [J]. HEART, 2012, 98 : E290 - E290
  • [5] Coronary artery stent imaging with 128-slice dual-source CT using high-pitch spiral acquisition in a cardiac phantom: comparison with the sequential and low-pitch spiral mode
    Florian Wolf
    Sebastian Leschka
    Christian Loewe
    Peter Homolka
    Christina Plank
    Ruediger Schernthaner
    Dominik Bercaczy
    Robert Goetti
    Johannes Lammer
    Guy Friedrich
    Borut Marincek
    Hatem Alkadhi
    Gudrun Feuchtner
    [J]. European Radiology, 2010, 20 : 2084 - 2091
  • [6] Benefits of High-Pitch 128-Slice Dual-Source Computed Tomography for Planning of Transcatheter Aortic Valve Implantation
    Plank, Fabian
    Friedrich, Guy
    Bartel, Thomas
    Mueller, Silvana
    Bonaros, Nikolaos
    Heinz, Anneliese
    Klauser, Andrea
    Cartes-Zumelzu, Fabiola
    Grimm, Michael
    Feuchtner, Gudrun
    [J]. ANNALS OF THORACIC SURGERY, 2012, 94 (06): : 1961 - 1966
  • [7] Low-dose, 128-slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode
    Alkadhi, Hatem
    Stolzmann, Paul
    Desbiolles, Lotus
    Baumueller, Stephan
    Goetti, Robert
    Plass, Andre
    Scheffel, Hans
    Feuchtner, Gudrun
    Falk, Volkmar
    Marincek, Borut
    Leschka, Sebastian
    [J]. HEART, 2010, 96 (12) : 933 - 938
  • [8] Low Dose High-Pitch Spiral Acquisition 128-Slice Dual-Source Computed Tomography for the Evaluation of Coronary Artery Bypass Graft Patency
    Goetti, Robert
    Leschka, Sebastian
    Baumueller, Stephan
    Plass, Andre
    Wieser, Monika
    Desbiolles, Lotus
    Stolzmann, Paul
    Falk, Volkmar
    Marincek, Borut
    Alkadhi, Hatem
    Feuchtner, Gudrun
    [J]. INVESTIGATIVE RADIOLOGY, 2010, 45 (06) : 324 - 330
  • [9] Dual Left Anterior Descending Coronary Artery Diagnosed with 128-slice Dual-source CT
    Duarte, Ricardo
    Miranda, Diana
    Costa, Joao C.
    Oliveira, Joao
    Teixeira, Dinis
    Fernandez-Perez, Gabriel C.
    [J]. HEART LUNG AND CIRCULATION, 2011, 20 (09): : 605 - 605
  • [10] High-Pitch Spiral Coronary Angiography by 128-Slice Dual-Source Computed Tomography (CT) Realizes a Sensational Low Radiation Exposure Without Reduced Image Quality
    Utanohara, Yuko
    Iguchi, Nobuo
    Yamashita, Mitsumi
    Umemura, Jun
    Sumiyoshi, Tetsuya
    [J]. CIRCULATION, 2010, 122 (21)