Low-dose, 128-slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode

被引:147
|
作者
Alkadhi, Hatem [1 ,2 ]
Stolzmann, Paul [1 ]
Desbiolles, Lotus [1 ]
Baumueller, Stephan [1 ]
Goetti, Robert [1 ]
Plass, Andre [3 ]
Scheffel, Hans [1 ]
Feuchtner, Gudrun [1 ]
Falk, Volkmar [3 ]
Marincek, Borut [1 ]
Leschka, Sebastian [1 ]
机构
[1] Univ Zurich Hosp, Inst Diagnost Radiol, CH-8091 Zurich, Switzerland
[2] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Cardiac MR PET CT Program, Boston, MA USA
[3] Univ Zurich Hosp, Ctr Cardiovasc, CH-8091 Zurich, Switzerland
关键词
COMPUTED-TOMOGRAPHY; DIAGNOSIS;
D O I
10.1136/hrt.2009.189100
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective To compare the diagnostic accuracy and radiation doses of two low-dose protocols for coronary artery imaging with second-generation, dual-source CT in comparison with catheter angiography (CA). Design, setting and patients Prospective, single-centre study conducted in a referral centre enrolling 100 patients with low-to-intermediate risk and suspicion of coronary artery disease. All patients underwent contrast-enhanced, 128-slice, dual-source CT coronary angiography and CA. Patients were randomly assigned to two different low-dose CT protocols (each 100 kV/320 mA): in group A (n = 50), CT was performed using the prospectively electrocardiography (ECG)-gated step-and-shoot (SAS) mode; in group B (n = 50), CT was performed using the prospectively ECG-gated high-pitch mode (pitch 3.4). The image quality and presence or absence of significant coronary stenosis in all coronary segments were evaluated by two blinded and independent observers. CA served as the standard of reference. Results Sixty-one significant stenoses were found in group A, and 69 in group B. There was no significant difference in age (group A, 62+/-8yrs; group B, 63+/-8yrs; p = 0.72), body mass index (group A, 26.4+/-3.1kg/m2; group B, 25.9+/-2.8kg/m2; p = 0.41) and heart rate (HR) (group A, 58+/-8bpm; group B, 56+/-10bpm; p = 0.66) between the groups. Diagnostic image quality was obtained in 98.6% (651/660) of segments in group A and in 98.9% (642/649) in group B, with no significant differences between groups. Sensitivity, specificity and positive and negative predictive values were 94%, 91%, 85% and 97% per-patient in group A, and 93%, 94%, 89% and 97% per-patient in group B (no significant differences). The effective radiation dose in group B (0.9+/-0.1 mSv) was significantly (p<0.01) lower than that in group A (1.4+/-0.4 mSv). Conclusions Both the high-pitch and the SAS mode for low-dose CT coronary angiography provide high accuracy for the assessment of significant coronary stenoses, while the high-pitch mode further significantly lowers the radiation dose.
引用
收藏
页码:933 / 938
页数:6
相关论文
共 50 条
  • [1] Dual-source CT in step-and-shoot mode: Noninvasive coronary angiography with low radiation dose
    Stolzmann, Paul
    Leschka, Sebastian
    Scheffel, Hans
    Krauss, Tobias
    Desbiolles, Lotus
    Plass, Andre
    Genoni, Michele
    Flohr, Thomas G.
    Wildermuth, Simon
    Marincek, Borut
    Alkadhi, Hatem
    [J]. RADIOLOGY, 2008, 249 (01) : 71 - 80
  • [2] Accuracy of 128-slice dual-source CT using high-pitch spiral mode for the assessment of coronary stents
    Yang, X.
    Chen, C. Y. D.
    Yang, Y. J. J.
    He, H. B.
    Jing, Wang
    [J]. EUROPEAN HEART JOURNAL, 2013, 34 : 844 - 844
  • [3] 128-slice dual-source CT coronary angiography with prospectively electrocardiography-triggered high-pitch spiral mode: radiation dose, image quality, and diagnostic acceptability
    Matsubara, Kosuke
    Sakuda, Keita
    Nunome, Haruka
    Takata, Tadanori
    Koshida, Kichiro
    Gabata, Toshifumi
    [J]. ACTA RADIOLOGICA, 2016, 57 (01) : 25 - 32
  • [4] Radiation dose for thoracic and coronary step-and-shoot CT using a 128-slice dual-source machine in infants and small children with congenital heart disease
    Jean-François Paul
    Adela Rohnean
    Eric Elfassy
    Anne Sigal-Cinqualbre
    [J]. Pediatric Radiology, 2011, 41 : 244 - 249
  • [5] Radiation dose for thoracic and coronary step-and-shoot CT using a 128-slice dual-source machine in infants and small children with congenital heart disease
    Paul, Jean-Francois
    Rohnean, Adela
    Elfassy, Eric
    Sigal-Cinqualbre, Anne
    [J]. PEDIATRIC RADIOLOGY, 2011, 41 (02) : 244 - 249
  • [6] Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease
    Pei Nie
    Ximing Wang
    Zhaoping Cheng
    Xiaopeng Ji
    Yanhua Duan
    Jiuhong Chen
    [J]. European Radiology, 2012, 22 : 2057 - 2066
  • [7] High-Pitch 128-Slice Dual-Source CT for the Assessment of Coronary Stents in a Phantom Model
    Donati, Olivio F.
    Burg, Matthias C.
    Desbiolles, Lotus
    Karlo, Christoph
    Stolzmann, Paul
    Bunck, Alexander
    Baumueller, Stephan
    Marincek, Borut
    Seifarth, Harald
    Alkadhi, Hatem
    Leschka, Sebastian
    Maintz, David
    [J]. ACADEMIC RADIOLOGY, 2010, 17 (11) : 1366 - 1374
  • [8] Accuracy, image quality and radiation dose comparison of high-pitch spiral and sequential acquisition on 128-slice dual-source CT angiography in children with congenital heart disease
    Nie, Pei
    Wang, Ximing
    Cheng, Zhaoping
    Ji, Xiaopeng
    Duan, Yanhua
    Chen, Jiuhong
    [J]. EUROPEAN RADIOLOGY, 2012, 22 (10) : 2057 - 2066
  • [9] Accuracy of 128-slice dual-source CT using high-pitch spiral mode for the assessment of coronary stents: First in vivo experience
    Yang Xia
    Yang Junjie
    Zhou Ying
    He Bai
    Wang Qi
    Jin Qinhua
    Chen Yundai
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2013, 82 (04) : 617 - 622
  • [10] DIAGNOSTIC ACCURACY OF 128-SLICE DUAL-SOURCE CT USING HIGH-PITCH SPIRAL MODE IN ASSESSMENT OF CORONARY ARTERY STENT IMAGING COMPARISON WITH INVASIVE CORONARY ANGIOGRAPHY
    Yang Xia
    Chen Yundai
    [J]. HEART, 2012, 98 : E290 - E290