Dual-source CT in step-and-shoot mode: Noninvasive coronary angiography with low radiation dose

被引:203
|
作者
Stolzmann, Paul [1 ]
Leschka, Sebastian [1 ]
Scheffel, Hans [1 ]
Krauss, Tobias [1 ]
Desbiolles, Lotus [1 ]
Plass, Andre [2 ]
Genoni, Michele [2 ]
Flohr, Thomas G. [3 ]
Wildermuth, Simon [4 ]
Marincek, Borut [1 ]
Alkadhi, Hatem [1 ]
机构
[1] Univ Zurich Hosp, Inst Diagnost Radiol, CH-8091 Zurich, Switzerland
[2] Univ Zurich Hosp, Cardiovasc Surg Clin, CH-8091 Zurich, Switzerland
[3] Siemens Med Solut, Comp Tomog CTE PA, Forchheim, Germany
[4] Kantonsspital St Gallen, Inst Radiol, St Gallen, Switzerland
基金
瑞士国家科学基金会;
关键词
D O I
10.1148/radiol.2483072032
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To prospectively investigate computed tomographic (CT) image quality parameters by using different protocols and to calculate radiation dose estimates for noninvasive coronary angiography performed with dual-source CT in the step-and-shoot (SAS) mode. Materials and Methods: This study was local ethics board approved; written informed consent was obtained from all patients. In the preliminary portion of the study, 40 patients underwent CT coronary angiography in the SAS mode: at 100 kV (protocol A) in 22 patients with a body mass index (BMI) of less than 25 kg/m(2) and at 120 kV (protocol B) in 18 patients with a BMI of 25-30 kg/m(2). Both protocols involved use of an attenuation-based tube current and 1 mL of contrast material per kilogram of body weight. The final portion of the study involved 50 additional patients: 21 patients with a BMI of 25-30 kg/cm(2) assigned to protocol B and 29 patients with a BMI of less than 25 kg/cm(2) assigned to protocol C, which was performed with 100 kV, an attenuation-based tube current, and a reduced contrast material dose of 0.8 mL/kg. Image quality was independently assessed. Attenuation in the aorta and coronary arteries and image noise were measured. Radiation dose was estimated. Results: Mean image noise was similar with protocols A and B. Mean attenuation in the aorta and coronary arteries with protocol A (444 HU) was significantly (P < .001) higher than that with protocol B (358 HU). The reduced contrast material dose in protocol C yielded attenuation similar to that with protocol B. Diagnostic image quality was achieved with all protocols in 1237 (97.9%) of 1264 coronary segments. No significant differences in image quality between the 100- and 120-kV protocols were found. Mean heart rate had a significant effect on motion artifacts (area under receiver operating characteristic curve [AUC] = 0.818; 95% confidence interval [CI]: 0.723, 0.892; P < .001), whereas heart rate variability had a significant effect on stair-step artifacts (AUC = 0.79; 95% CI: 0.687, 0.865; P < .001). The mean estimated effective dose was 1.2 mSv +/- 0.2 for protocols A and C and 2.6 mSv +/- 0.5 for protocol B. Conclusion: Dual-source SAS-mode CT coronary angiography yielded diagnostic image quality for 97.9% of coronary segments at a low radiation dose. (C) RSNA, 2008.
引用
收藏
页码:71 / 80
页数:10
相关论文
共 50 条
  • [1] Low-dose, 128-slice, dual-source CT coronary angiography: accuracy and radiation dose of the high-pitch and the step-and-shoot mode
    Alkadhi, Hatem
    Stolzmann, Paul
    Desbiolles, Lotus
    Baumueller, Stephan
    Goetti, Robert
    Plass, Andre
    Scheffel, Hans
    Feuchtner, Gudrun
    Falk, Volkmar
    Marincek, Borut
    Leschka, Sebastian
    [J]. HEART, 2010, 96 (12) : 933 - 938
  • [2] Low-dose CT coronary angiography in the step-and-shoot mode: diagnostic performance
    Scheffel, H.
    Alkadhi, H.
    Leschka, S.
    Plass, A.
    Desbiolles, L.
    Guber, I.
    Krauss, T.
    Gruenenfelder, J.
    Genoni, M.
    Luescher, T. F.
    Marincek, B.
    Stolzmann, P.
    [J]. HEART, 2008, 94 (09) : 1132 - 1137
  • [3] Low-Dose Scan Protocols in Dual-Source CT Coronary Angiography
    Li, Min
    Sun, Gang
    [J]. RADIOLOGY, 2012, 263 (03) : 937 - 938
  • [4] Radiation dose for thoracic and coronary step-and-shoot CT using a 128-slice dual-source machine in infants and small children with congenital heart disease
    Jean-François Paul
    Adela Rohnean
    Eric Elfassy
    Anne Sigal-Cinqualbre
    [J]. Pediatric Radiology, 2011, 41 : 244 - 249
  • [5] Radiation dose for thoracic and coronary step-and-shoot CT using a 128-slice dual-source machine in infants and small children with congenital heart disease
    Paul, Jean-Francois
    Rohnean, Adela
    Elfassy, Eric
    Sigal-Cinqualbre, Anne
    [J]. PEDIATRIC RADIOLOGY, 2011, 41 (02) : 244 - 249
  • [6] Radiation dose of dual-source CT is much higher compared to invasive coronary angiography
    Duchonova, R.
    Adla, T.
    Veselka, J.
    Neuwirth, J.
    Martinkovicova, L.
    Suchanek, V.
    Hajek, P.
    Maly, M.
    Zemanek, D.
    Blasko, P.
    [J]. CIRCULATION, 2008, 118 (12) : E291 - E292
  • [7] Low-Dose Scan Protocols in Dual-Source CT Coronary Angiography Response
    Neefjes, Lisan A.
    Mollet, Nico R.
    [J]. RADIOLOGY, 2012, 263 (03) : 938 - 938
  • [8] Low dose dual-source CT angiography of the thoracic aorta
    Farrelly, Cormac
    Davarpanah, Amir
    Keeling, Aoife N.
    Sheehan, John
    Ragin, Ann
    Yaghmai, Vahid
    Carr, James C.
    [J]. INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2011, 27 (07): : 1025 - 1034
  • [9] Contrast medium administration with a body surface area protocol in step-and-shoot coronary computed tomography angiography with dual-source scanners
    Jin, Liang
    Gao, Yiyi
    Sun, Yingli
    Li, Cheng
    Gao, Pan
    Zhao, Wei
    Li, Ming
    [J]. SCIENTIFIC REPORTS, 2020, 10 (01)
  • [10] Low dose dual-source CT angiography of the thoracic aorta
    Cormac Farrelly
    Amir Davarpanah
    Aoife N. Keeling
    John Sheehan
    Ann Ragin
    Vahid Yaghmai
    James C. Carr