摘要
A new three-dimensional treatment planning system (?Ps) based on convolution/superposition algorithms (TMS-Radix from HELAX AB, Uppsala, Sweden) was recently installed at the University Hospital in Lund. The purpose of the present study was to design a quality assurance and acceptance testing programme to meet the specific characteristics of this convolution model. The model is based on parametrization of a non-measurable quantity-the polyenergetic pencil beam. However, the verification of the treatment planning model is still dependent on numerous comparisons of measured depth-doses and dose profiles. The test programme was divided in two basic parts: (i) model implementation and beam data consistency and (ii) model performance and limitations in special situations. The first part was scheduled for all photon beam qualities available before they could be used for clinical treatment planning. The second part was performed for selected energies only. The results indicate clearly that the model is well suited for clinical three-dimensional dose planning and that the TPS handles data as expected. For example, calculated depth-doses for open and wedge beams at depths larger than the depth of dose maximum and profiles for open beams shows a very good agreement with measurements. However, depth-dose deviations at shallow depths, especially for high energies, were found. Monitor units calculated by the system were accurate for most fields except for very large fields, where deviations of several per cent were found.
引用
收藏
页码:1609 / 1628
页数:20
相关论文
共 50 条
  • [1] Dosimetric verification of treatment planning system superficial dose calculations for proton beam
    Rah, Jeong-Eun
    Shin, Dongho
    Park, Jeong-hoon
    Hwang, Ui-Jung
    Oh, Do Hoon
    Park, Sung Yong
    RADIATION MEASUREMENTS, 2013, 55 : 60 - 63
  • [2] Dosimetric verification of clinical radiotherapy treatment planning system
    Kolarevic, Goran
    Jaros, Drazan
    Marosevic, Goran
    Ignjatic, Dejan
    Mirjanic, Dragoljub
    VOJNOSANITETSKI PREGLED, 2022, 79 (02) : 115 - 124
  • [3] Dosimetric verification of a clinical inverse treatment planning system
    Xing, L
    Findley, D
    Ma, L
    Ma, CM
    Forster, K
    Boyer, AL
    RADIOLOGY, 1998, 209P : 397 - 397
  • [4] Dosimetric verification of a commercial inverse treatment planning system
    Lei, X
    Curran, B
    Hill, R
    Holmes, T
    Ma, LJ
    Forster, KM
    Boyer, AL
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (02): : 463 - 478
  • [5] Verification of a pencil beam based treatment planning system:: output factors for open photon beams shaped with MLC or blocks
    Hansson, H
    Björk, P
    Knöös, T
    Nilsson, P
    PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (09): : N201 - N207
  • [6] CLINICAL COMMISSIONING AND DOSIMETRIC VERIFICATION OF THE RAYSTATION TREATMENT PLANNING SYSTEM
    Tugrul, Taylan
    EAST EUROPEAN JOURNAL OF PHYSICS, 2021, (04): : 114 - 119
  • [7] Test procedures for verification of an electron pencil beam algorithm implemented for treatment planning
    Blomquist, M
    Karlsson, M
    Karlsson, M
    RADIOTHERAPY AND ONCOLOGY, 1996, 39 (03) : 271 - 286
  • [8] Dosimetric comparison of treatment planning systems using collapsed cone convolution and pencil beam algorithms
    Elcim, Yelda
    Dirican, Bahar
    Yavas, Omer
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2016, 15 (04) : 364 - 377
  • [9] An integrated Monte Carlo dosimetric verification system for radiotherapy treatment planning
    Yamamoto, T.
    Mizowaki, T.
    Miyabe, Y.
    Takegawa, H.
    Narita, Y.
    Yano, S.
    Nagata, Y.
    Teshima, T.
    Hiraoka, M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (07): : 1991 - 2008
  • [10] Dosimetric verification of a commercial Monte Carlo treatment planning system (VMC++) for a 9 MeV electron beam
    Schiapparelli, P.
    Zefiro, D.
    Taccini, G.
    MEDICAL PHYSICS, 2009, 36 (05) : 1759 - 1767