An integrated Monte Carlo dosimetric verification system for radiotherapy treatment planning

被引:20
|
作者
Yamamoto, T.
Mizowaki, T.
Miyabe, Y.
Takegawa, H.
Narita, Y.
Yano, S.
Nagata, Y.
Teshima, T.
Hiraoka, M.
机构
[1] Osaka Univ, Grad Sch Med, Dept Med Phys & Engn, Suita, Osaka 5650871, Japan
[2] Kyoto Univ, Grad Sch Med, Dept Radiat Oncol & Image Appl Therapy, Kyoto 6068507, Japan
[3] Kyoto Univ, Grad Sch Engn, Dept Nucl Engn, Sakyo Ku, Kyoto 6068501, Japan
[4] Kyoto Univ Hosp, Clin Radiol Serv Div, Sakyo Ku, Kyoto 6068507, Japan
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2007年 / 52卷 / 07期
关键词
D O I
10.1088/0031-9155/52/7/014
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An integrated Monte Carlo (MC) dose calculation system, MCRTV (Monte Carlo for radiotherapy treatment plan verification), has been developed for clinical treatment plan verification, especially for routine quality assurance (QA) of intensity-modulated radiotherapy (IMRT) plans. The MCRTV system consists of the EGS4/ PRESTA MC codes originally written for particle transport through the accelerator, the multileaf collimator (MLC), and the patient/phantom, which run on a 28-CPU Linux cluster, and the associated software developed for the clinical implementation. MCRTV has an interface with a commercial treatment planning system (TPS) ( Eclipse, Varian Medical Systems, Palo Alto, CA, USA) and reads the information needed for MC computation transferred in DICOM-RT format. The key features of MCRTV have been presented in detail in this paper. The phase- space data of our 15 MV photon beam from a Varian Clinac 2300C/ D have been developed and several benchmarks have been performed under homogeneous and several inhomogeneous conditions (including water, aluminium, lung and bone media). The MC results agreed with the ionization chamber measurements to within 1% and 2% for homogeneous and inhomogeneous conditions, respectively. The MC calculation for a clinical prostate IMRT treatment plan validated the implementation of the beams and the patient/phantom configuration in MCRTV.
引用
收藏
页码:1991 / 2008
页数:18
相关论文
共 50 条
  • [1] Dosimetric verification of clinical radiotherapy treatment planning system
    Kolarevic, Goran
    Jaros, Drazan
    Marosevic, Goran
    Ignjatic, Dejan
    Mirjanic, Dragoljub
    [J]. VOJNOSANITETSKI PREGLED, 2022, 79 (02) : 115 - 124
  • [2] Dosimetric verification of a Monte Carlo treatment planning system for MR-guided proton therapy
    Fuchs, H.
    Padilla-Cabal, F.
    Oborn, B.
    Traneus, E.
    Georg, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1544 - S1545
  • [3] Dosimetric verification of breast IMRT treatment planning using Monte Carlo method
    Gagne, I.
    Zavgorodni, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S224 - S225
  • [4] Radiotherapy treatment planning with Monte Carlo on a distributed system
    Chauvie, S
    Scielzo, G
    [J]. 2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 1765 - 1769
  • [5] Monte Carlo simulation and dosimetric verification of radiotherapy beam modifiers
    Spezi, E
    Lewis, DG
    Smith, CW
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (11): : 3007 - 3029
  • [6] Dosimetric evaluation of a monte carlo IMRT treatment planning system with the MIMiC
    Rassiah-Szegedi, P.
    Fuss, M.
    Sheikh-Bagheri, D.
    Szegedi, M.
    Papanikolaoul, N.
    Stathakis, S.
    Salter, B.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2441 - 2442
  • [7] Dosimetric verification of IMRT treatment planning using Monte Carlo simulations for prostate cancer
    Yang, J
    Li, J
    Chen, L
    Price, R
    McNeeley, S
    Qin, L
    Wang, L
    Xiong, W
    Ma, CM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (05): : 869 - 878
  • [8] Monte Carlo for Radiotherapy Treatment Planning
    Ma, C.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [9] Dosimetric verification of a commercial Monte Carlo treatment planning system (VMC++) for a 9 MeV electron beam
    Schiapparelli, P.
    Zefiro, D.
    Taccini, G.
    [J]. MEDICAL PHYSICS, 2009, 36 (05) : 1759 - 1767
  • [10] Monte Carlo as a tool for treatment planning verification
    Sánchez-Doblado, F
    Leal, A
    Perucha, A
    Arráns, R
    Rincón, A
    Sánchez-Nieto, B
    Carrasco, E
    Roselló, J
    Errazquin, L
    Sánchez-Calzado, JA
    [J]. PHYSICA MEDICA, 2001, 17 : 84 - 86