THE DOSIMETRIC VERIFICATION OF A PENCIL BEAM BASED TREATMENT PLANNING SYSTEM

被引:66
|
作者
KNOOS, T
CEBERG, C
WEBER, L
NILSSON, P
机构
[1] Dept. of Radiat. Phys., Univ. Hospital, Lund
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1994年 / 39卷 / 10期
关键词
D O I
10.1088/0031-9155/39/10/007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new three-dimensional treatment planning system (?Ps) based on convolution/superposition algorithms (TMS-Radix from HELAX AB, Uppsala, Sweden) was recently installed at the University Hospital in Lund. The purpose of the present study was to design a quality assurance and acceptance testing programme to meet the specific characteristics of this convolution model. The model is based on parametrization of a non-measurable quantity-the polyenergetic pencil beam. However, the verification of the treatment planning model is still dependent on numerous comparisons of measured depth-doses and dose profiles. The test programme was divided in two basic parts: (i) model implementation and beam data consistency and (ii) model performance and limitations in special situations. The first part was scheduled for all photon beam qualities available before they could be used for clinical treatment planning. The second part was performed for selected energies only. The results indicate clearly that the model is well suited for clinical three-dimensional dose planning and that the TPS handles data as expected. For example, calculated depth-doses for open and wedge beams at depths larger than the depth of dose maximum and profiles for open beams shows a very good agreement with measurements. However, depth-dose deviations at shallow depths, especially for high energies, were found. Monitor units calculated by the system were accurate for most fields except for very large fields, where deviations of several per cent were found.
引用
收藏
页码:1609 / 1628
页数:20
相关论文
共 50 条
  • [41] Effects of spot parameters in pencil beam scanning treatment planning
    Kraan, Aafke Christine
    Depauw, Nicolas
    Clasie, Ben
    Giunta, Marina
    Madden, Tom
    Kooy, Hanne M.
    MEDICAL PHYSICS, 2018, 45 (01) : 60 - 73
  • [42] Validation and Application of a Commercial Proton Monte Carlo Pencil Beam Scanning Treatment Planning System
    Huang, S.
    Maslowski, A.
    Hiltunen, P.
    Lin, L.
    MEDICAL PHYSICS, 2019, 46 (06) : E589 - E589
  • [43] First Quality Assurance Implementation of Treatment Planning System for Pencil Beam Scanning Proton Radiotherapy
    Xiong, W.
    Yu, G.
    Kang, M.
    Huang, S.
    Lin, H.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [44] PERFORMANCE EVALUATION OF THE CMS MODULEX TREATMENT PLANNING SYSTEM FOR PENCIL BEAM ELECTRON DOSE COMPUTATIONS
    KEYS, RA
    HARMS, WB
    WONG, JW
    PURDY, JA
    MEDICAL PHYSICS, 1985, 12 (04) : 533 - 533
  • [45] Convergent Pencil Beam Modeling with Monte Carlo Transport Codes for Synchrotron Based Intensity Modulated Proton Beam System Model in Eclipse Treatment Planning System
    Moskvin, V.
    Zhao, L.
    Farr, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E266 - E267
  • [46] Using beam profile inflection point in process of treatment planning system verification
    Wendykier, Jacek
    Grzadziel, Aleksandra
    Bekman, Barbara
    Bieniasiewicz, Marcin
    Bekman, Adam
    Wendykier, Piotr
    Wozniak, Bozena
    Reudelsdorf, Marta
    Slosarek, Krzysztof
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2021, 26 (04) : 553 - 562
  • [47] Development of Optical Fiber Based Measurement System for the Verification of Entrance Dose Map in Pencil Beam Scanning Proton Beam
    Son, Jaeman
    Lee, Se Byeong
    Lim, Youngkyung
    Park, Sung Yong
    Cho, Kwanho
    Yoon, Myonggeun
    Shin, Dongho
    SENSORS, 2018, 18 (01):
  • [48] Dosimetric assessment of the PRESAGE dosimeter for a proton pencil beam
    Wuu, C-S
    Xu, Y.
    Qian, X.
    Adamovics, J.
    Cascio, E.
    Lu, H-M
    7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [49] Dosimetric verification of open asymmetric photon fields calculated with a treatment planning system based on dose-to-energy-fluence concepts
    Hurkmans, C
    Knoos, T
    Nilsson, P
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (08): : 1277 - 1290
  • [50] Dosimetric verification of radiotherapy treatment planning systems in Serbia: national audit
    Laza Rutonjski
    Borislava Petrović
    Milutin Baucal
    Milan Teodorović
    Ozren Čudić
    Eduard Gershkevitsh
    Joanna Izewska
    Radiation Oncology, 7