THE EFFECT OF HENNA PASTE ON OXYGEN-SATURATION READING OBTAINED BY PULSE OXIMETRY

被引:0
|
作者
ALMAJED, SA [1 ]
HARAKATI, MS [1 ]
机构
[1] KING SAUD UNIV, KING KHALID UNIV HOSP, COLL MED, DEPT MED 38, RIYADH 11472, SAUDI ARABIA
来源
TROPICAL AND GEOGRAPHICAL MEDICINE | 1994年 / 46卷 / 01期
关键词
HENNA; PULSE OXIMETRY; OXYGEN SATURATION; SAUDI ARABIA;
D O I
暂无
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
A controlled study was designed to examine the effect of red and black henna on the measurement of oxygen saturation by pulse oximetry. Fifty adult normal female volunteers had their left thumb coloured with red henna (40) or black henna (10). The uncoloured right thumb was used as a control. All measurements were done under room temperature, and the same pulse oximeter was used in ah volunteers. In the red henna group there was no difference on oximetry reading between right and left thumb. In contrast, the thumb coloured with black henna gave no reading when compared to the uncoloured right thumb, On the basis of this study, pulse oximetry is not limited by red henna while black henna has a potential of causing major error in the measurement of oxygen saturation by pulse oximetry, In these cases we recommend to use ear oximetry for accurate measurement of oxygen saturation.
引用
收藏
页码:38 / 39
页数:2
相关论文
共 50 条
  • [1] Effects of henna dye on oxygen saturation reading using pulse oximetry
    Samman, YS
    Rahimi, JS
    Wali, SO
    Krayem, AB
    Abdelaziz, MM
    SAUDI MEDICAL JOURNAL, 2006, 27 (02) : 268 - 269
  • [2] THE EFFECT OF CARBOXYHEMOGLOBIN ON PULSE-OXIMETRY OXYGEN-SATURATION MEASUREMENTS
    PANACEK, EA
    JOURILLES, NJ
    SHERMAN, BW
    RUTHERFORD, WF
    CLINICAL RESEARCH, 1991, 39 (02): : A421 - A421
  • [3] MONITORING OF ARTERIAL OXYGEN-SATURATION BY PULSE OXIMETRY
    STRIEBEL, HW
    STEINHOFF, U
    KRETZ, FJ
    ANASTHESIOLOGIE & INTENSIVMEDIZIN, 1988, 29 (01): : 8 - 16
  • [4] NASAL PULSE OXIMETRY OVERESTIMATES OXYGEN-SATURATION
    ROSENBERG, J
    PEDERSEN, MH
    ANAESTHESIA, 1990, 45 (12) : 1070 - 1071
  • [5] PREDICTING CHANGES IN OXYGEN-SATURATION USING PULSE OXIMETRY
    SAPSFORD, DJ
    JONES, JG
    BRITISH JOURNAL OF ANAESTHESIA, 1988, 61 (04) : P497 - P497
  • [6] TIME TO EQUILIBRATION OF OXYGEN-SATURATION USING PULSE OXIMETRY
    GRUBER, P
    KWIATKOWSKI, T
    SILVERMAN, R
    FLASTER, E
    AUERBACH, C
    ACADEMIC EMERGENCY MEDICINE, 1995, 2 (09) : 810 - 815
  • [7] FETAL OXYGEN-SATURATION MEASUREMENT BY TRANSMISSION PULSE OXIMETRY
    BUSCHMANN, J
    RALL, G
    KNITZA, R
    LANCET, 1992, 339 (8793): : 615 - 615
  • [8] MONITORING OF ARTERIAL OXYGEN-SATURATION BY PULSE OXIMETRY - COMMENT
    SALEM, CJ
    JANDA, A
    BERGER, M
    BRANDSTETTER, A
    ANASTHESIOLOGIE & INTENSIVMEDIZIN, 1988, 29 (10): : 293 - 293
  • [9] CONTINUOUS MONITORING OF FETAL OXYGEN-SATURATION BY PULSE OXIMETRY
    LUTTKUS, A
    FENGLER, TW
    FRIEDMANN, W
    DUDENHAUSEN, JW
    OBSTETRICS AND GYNECOLOGY, 1995, 85 (02): : 183 - 186
  • [10] PULSE OXIMETRY IN METHEMOGLOBINEMIA - FAILURE TO DETECT LOW OXYGEN-SATURATION
    RIEDER, HU
    FREI, FJ
    ZBINDEN, AM
    THOMSON, DA
    ANAESTHESIA, 1989, 44 (04) : 326 - 327