NASAL PULSE OXIMETRY OVERESTIMATES OXYGEN-SATURATION

被引:12
|
作者
ROSENBERG, J [1 ]
PEDERSEN, MH [1 ]
机构
[1] UNIV COPENHAGEN,HVIDOVRE HOSP,DEPT ANAESTHESIOL,DK-2650 HVIDOVRE,DENMARK
关键词
Equipment; pulse oximeter; Oxygen; hypoxaemia;
D O I
10.1111/j.1365-2044.1990.tb14892.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Ten surgical patients were monitored with nasal and finger pulse oximetry (Nellcor N‐200) for five study periods with alternating mouth and nasal breathing and switching of cables and sensors. Nasal pulse oximetry was found to overestimate arterial oxygen saturation by 4.7 (SD 1.4%) (bias and precision). Copyright © 1990, Wiley Blackwell. All rights reserved
引用
收藏
页码:1070 / 1071
页数:2
相关论文
共 50 条
  • [1] Pulse Oximetry Overestimates Oxygen Saturation in COPD
    Amalakanti, Sridhar
    Pentakota, Mohan Rao
    [J]. RESPIRATORY CARE, 2016, 61 (04) : 423 - 427
  • [2] EXPLANATION FOR THE OVERESTIMATION OF OXYGEN-SATURATION SEEN USING NASAL PULSE OXIMETRY
    BROOME, IJ
    REILLY, CS
    [J]. ANAESTHESIA, 1991, 46 (05) : 424 - 424
  • [3] MONITORING OF ARTERIAL OXYGEN-SATURATION BY PULSE OXIMETRY
    STRIEBEL, HW
    STEINHOFF, U
    KRETZ, FJ
    [J]. ANASTHESIOLOGIE & INTENSIVMEDIZIN, 1988, 29 (01): : 8 - 16
  • [4] TIME TO EQUILIBRATION OF OXYGEN-SATURATION USING PULSE OXIMETRY
    GRUBER, P
    KWIATKOWSKI, T
    SILVERMAN, R
    FLASTER, E
    AUERBACH, C
    [J]. ACADEMIC EMERGENCY MEDICINE, 1995, 2 (09) : 810 - 815
  • [5] PREDICTING CHANGES IN OXYGEN-SATURATION USING PULSE OXIMETRY
    SAPSFORD, DJ
    JONES, JG
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1988, 61 (04) : P497 - P497
  • [6] FETAL OXYGEN-SATURATION MEASUREMENT BY TRANSMISSION PULSE OXIMETRY
    BUSCHMANN, J
    RALL, G
    KNITZA, R
    [J]. LANCET, 1992, 339 (8793): : 615 - 615
  • [7] CONTINUOUS MONITORING OF FETAL OXYGEN-SATURATION BY PULSE OXIMETRY
    LUTTKUS, A
    FENGLER, TW
    FRIEDMANN, W
    DUDENHAUSEN, JW
    [J]. OBSTETRICS AND GYNECOLOGY, 1995, 85 (02): : 183 - 186
  • [8] MONITORING OF ARTERIAL OXYGEN-SATURATION BY PULSE OXIMETRY - COMMENT
    SALEM, CJ
    JANDA, A
    BERGER, M
    BRANDSTETTER, A
    [J]. ANASTHESIOLOGIE & INTENSIVMEDIZIN, 1988, 29 (10): : 293 - 293
  • [10] PULSE OXIMETRY IN METHEMOGLOBINEMIA - FAILURE TO DETECT LOW OXYGEN-SATURATION
    RIEDER, HU
    FREI, FJ
    ZBINDEN, AM
    THOMSON, DA
    [J]. ANAESTHESIA, 1989, 44 (04) : 326 - 327