Fast range measurement of spot scanning proton beams using a volumetric liquid scintillator detector

被引:16
|
作者
Hui, Cheuk Kai [1 ]
Robertson, Daniel [1 ]
Alsanea, Fahed [2 ]
Beddar, Sam [1 ,2 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas Houston, Grad Sch Biomed Sci, Houston, TX 77030 USA
来源
关键词
proton beam quality assurance; proton beam range measurement; liquid scintillator detector;
D O I
10.1088/2057-1976/1/2/025204
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Accurate confirmation and verification of the range of spot scanning proton beams is crucial for correct dose delivery. Current methods to measure proton beam range using ionization chambers are either time-consuming or result in measurements with poor spatial resolution. The large-volume liquid scintillator detector allows real-time measurements of the entire dose profile of a spot scanning proton beam. Thus, liquid scintillator detectors are an ideal tool for measuring the proton beam range for commissioning and quality assurance. However, optical artefacts may decrease the accuracy of measuring the proton beam range within the scintillator tank. The purpose of the current study was to (1) develop a geometric calibration system to accurately calculate physical distances within the liquid scintillator detector, taking into account optical artefacts; and (2) assess the accuracy, consistency, and robustness of proton beam range measurement using the liquid scintillator detector with our geometric calibration system. The range of the proton beam was measured with the calibrated liquid scintillator system and was compared to the nominal range. Measurements were made on three different days to evaluate the setup robustness from day to day, and three sets of measurements were made for each day to evaluate the consistency from delivery to delivery. All proton beam ranges measured using the liquid scintillator system were within half a millimeter of the nominal range. The delivery-to-delivery standard deviation of the range measurement was 0.04 mm, and the day-to-day standard deviation was 0.10 mm. In addition to the accuracy and robustness demonstrated by these results when our geometric calibration system was used, the liquid scintillator system allowed the range of all 94 proton beams to be measured in just two deliveries, making the liquid scintillator detector a perfect tool for range measurement of spot scanning proton beams.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Performance characterization of a 3D liquid scintillation detector for discrete spot scanning proton beam systems
    Darne, Chinmay D.
    Alsanea, Fahed
    Robertson, Daniel G.
    Sahoo, Narayan
    Beddar, Sam
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (14): : 5652 - 5667
  • [32] A novel and fast method for proton range verification using a step wedge and 2D scintillator
    Shen, Jiajian
    Allred, Bryce C.
    Robertson, Daniel G.
    Liu, Wei
    Sio, Terence T.
    Remmes, Nicholas B.
    Keole, Sameer R.
    Bues, Martin
    MEDICAL PHYSICS, 2017, 44 (09) : 4409 - 4414
  • [33] Measurement and simulation of neutron response function of organic liquid scintillator detector
    Gohil, M.
    Banerjee, K.
    Bhattacharya, S.
    Bhattacharya, C.
    Kundu, S.
    Rana, T. K.
    Mukherjee, G.
    Meena, J. K.
    Pandey, R.
    Pai, H.
    Ghosh, T. K.
    Dey, A.
    Mukhopadhyay, S.
    Pandit, D.
    Pal, S.
    Banerjee, S. R.
    Bandhopadhyay, T.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 664 (01): : 304 - 309
  • [34] Range shift verification in spot scanning proton therapy using gamma electron vertex imaging
    Kim, Sung Hun
    Jeong, Jong Hwi
    Ku, Youngmo
    Lee, Se Byeong
    Shin, Dongho
    Lim, Young Kyung
    Kim, Haksoo
    Kim, Chan Hyeong
    MEDICAL PHYSICS, 2024, 51 (03) : 1985 - 1996
  • [35] Commission Range Shifter On a Spot Scanning Proton System Using Raystation Treatment Planning System
    Ding, X.
    Rosen, L.
    Wu, H.
    MEDICAL PHYSICS, 2015, 42 (06) : 3375 - 3375
  • [36] FAST TIMING MEASUREMENT USING AN LaBr3 (Ce) SCINTILLATOR DETECTOR ARRAY COUPLED WITH GAMMASPHERE
    Rudigier, M.
    Lalkovski, S.
    Gamba, E. R.
    Bruce, A. M.
    Podolyak, Zs.
    Regan, P. H.
    Carpenter, M.
    Zhu, S.
    Ayangeakaa, A. D.
    Anderson, J. T.
    Berry, T.
    Bottoni, S.
    Burrows, I.
    Carroll, R.
    Copp, P.
    Cullen, D.
    Daniel, T.
    Fraile, L.
    Carmona Gallardo, M.
    Grant, A.
    Greene, J. P.
    Guegi, L. A.
    Hartley, D.
    Ilieva, R.
    Ilieva, S.
    Janssens, R. V. F.
    Kondev, F. G.
    Kroell, T.
    Lane, G. J.
    Lauritsen, T.
    Lazarus, I.
    Lotay, G.
    Martinez, G. Fernandez
    Pucknell, V.
    Reed, M.
    Rohrer, J.
    Sethi, J.
    Seweryniak, D.
    Shand, C. M.
    Simpson, J.
    Smolen, M.
    Stefanova, E.
    Vedia, V.
    Yordanov, O.
    ACTA PHYSICA POLONICA B, 2017, 48 (03): : 351 - 357
  • [37] Investigation of three-dimensional localisation of radioactive sources using a fast organic liquid scintillator detector
    Gamage, K. A. A.
    Joyce, M. J.
    Taylor, G. C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2013, 707 : 123 - 126
  • [38] Ionization quenching correction for 3D scintillator detectors for spot scanning proton therapy
    Alsanea, F.
    Beddar, S.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S701 - S701
  • [39] Development of an easy-to-handle range measurement tool using a plastic scintillator for proton beam therapy
    Fukushima, Yasuhiro
    Hamada, Minoru
    Nishio, Teiji
    Maruyama, Koichi
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (22): : 5927 - 5936
  • [40] Proton and Carbon Ion Beam Spot Size Measurement Using 5 Different Detector Types
    Witt, Matthias
    Weber, Uli
    Adeberg, Sebastian
    Baumann, Kilian-Simon
    Zink, Klemens
    INTERNATIONAL JOURNAL OF PARTICLE THERAPY, 2025, 15