FAILURE OF AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR TO REDETECT VENTRICULAR-FIBRILLATION IN PATIENTS WITH A NONTHORACOTOMY LEAD SYSTEM

被引:56
|
作者
JUNG, W [1 ]
MANZ, M [1 ]
MOOSDORF, R [1 ]
LUDERITZ, B [1 ]
机构
[1] UNIV BONN,DEPT CARDIOVASC SURG,W-5300 BONN 1,GERMANY
关键词
IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR; LEAD SYSTEMS; NONTHORACOTOMY; SUDDEN CARDIAC DEATH; ELECTROGRAM; ENDOCARDIAL;
D O I
10.1161/01.CIR.86.4.1217
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Shock delivery of an implantable defibrillator may cause a change in the amplitude of endocardial electrograms and impair the detection of ventricular fibrillation. Thus, the effects of shock discharges on the amplitude of endocardial electrograms were evaluated in five patients undergoing implantation of a cardioverter-defibrillator in combination with a new nonthoracotomy lead system. Methods and Results. At implant, bipolar endocardial electrograms were recorded before each shock application, during ventricular fibrillation, during redetection of ventricular fibrillation in case the applied shock was ineffective, and at intervals of 5, 10, 20, 30, 60, and 120 seconds after each shock delivery. The amplitude of the endocardial electrograms decreased from 10.5+/-3.8 mV during sinus rhythm to 6.3+/-1.9 mV during initial ventricular fibrillation and declined to 2.2+/-1.3 mV during redetection of ventricular fibrillation. After successful termination, the following bipolar electrograms could be obtained at the predetermined intervals: 1.9+/-1.2 mV, 3.1+/-1.8 mV, 4.5+/-1.9 mV, 6.5+/-2.9 mV, 9.5+/-3.3 mV, and 10.4+/-3.8 mV. At predischarge testing, failure of redetection of ventricular fibrillation could be documented in two patients, requiring rescue external defibrillation in both cases to restore sinus rhythm. Conclusions. These findings demonstrate that the implantable cardioverter-defibrillator did not ensure reliable redetection of ventricular fibrillation in patients using the implanted nonthoracotomy lead system. Thus, the potential risk of sudden cardiac death may persist in these patients despite defibrillator therapy.
引用
收藏
页码:1217 / 1222
页数:6
相关论文
共 50 条
  • [11] Undersensing of ventricular fibrillation in a noncommitted nonthoracotomy cardioverter defibrillator system
    Peralta, AO
    John, RM
    Venditti, FJ
    Martin, DT
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (03): : 610 - 618
  • [12] IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR THERAPY IN 300 PATIENTS WITH CORONARY-ARTERY-DISEASE PRESENTING EXCLUSIVELY WITH VENTRICULAR-FIBRILLATION
    LESSMEIER, TJ
    LEHMANN, MH
    STEINMAN, RT
    FROMM, BS
    AKHTAR, M
    CALKINS, H
    DIMARCO, JP
    EPSTEIN, AE
    ESTES, NAM
    FOGOROS, RN
    MARCHLINSKI, FE
    WILBER, D
    AMERICAN HEART JOURNAL, 1994, 128 (02) : 211 - 218
  • [13] LOW INCIDENCE OF LEAD RELATED COMPLICATIONS ASSOCIATED WITH NONTHORACOTOMY IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR SYSTEMS
    FAHY, GJ
    KLEMAN, JM
    WILKOFF, BL
    MORANT, VA
    PINSKI, SL
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (01): : 172 - 178
  • [14] A prospective, randomized evaluation of a nonthoracotomy implantable cardioverter defibrillator lead system
    Baker, JH
    Epstein, AE
    VoshageStahl, L
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01): : 72 - 78
  • [15] Impact of atrial fibrillation in patients with implantable cardioverter-defibrillator
    Madrid, A
    Rebollo, JG
    Peña, G
    Rodriguez, A
    Cano, L
    Mejia, MA
    Moro, C
    Camino, A
    EUROPEAN HEART JOURNAL, 2001, 22 : 476 - 476
  • [16] FAILURE OF THE AUTOMATIC IMPLANTABLE DEFIBRILLATOR TO DETECT VENTRICULAR-FIBRILLATION
    BARDY, GH
    IVEY, TD
    STEWART, R
    GRAHAM, EL
    GREENE, HL
    AMERICAN JOURNAL OF CARDIOLOGY, 1986, 58 (11): : 1107 - 1108
  • [17] Implantable cardioverter-defibrillator lead performance
    Maisel, William H.
    Kramer, Daniel B.
    CIRCULATION, 2008, 117 (21) : 2721 - 2723
  • [18] Implantable Cardioverter-Defibrillator Lead Failure in Children and Young Adults
    Janson, Christopher M.
    Patel, Akash R.
    Bonney, William J.
    Smoots, Karen
    Shah, Maully J.
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2014, 63 (02) : 133 - 140
  • [19] NONTHORACOTOMY LEAD SYSTEM FOR IMPLANTABLE DEFIBRILLATOR
    BLAKEMAN, BP
    SULLIVAN, HJ
    MONTOYA, A
    CALANDRA, D
    WILBER, D
    OLSHANSKY, B
    KALL, J
    KOPP, D
    PIFARRE, R
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1993, 106 (06): : 1040 - 1047
  • [20] Failure of implantable cardioverter-defibrillator leads: A matter of lead size?
    Rordorf, Roberto
    Poggio, Luca
    Savastano, Simone
    Vicentini, Alessandro
    Petracci, Barbara
    Chieffo, Enrico
    Klersy, Catherine
    Landolina, Maurizio
    HEART RHYTHM, 2013, 10 (02) : 184 - 190