Undersensing of ventricular fibrillation in a noncommitted nonthoracotomy cardioverter defibrillator system

被引:6
|
作者
Peralta, AO [1 ]
John, RM [1 ]
Venditti, FJ [1 ]
Martin, DT [1 ]
机构
[1] LAHEY CLIN MED CTR,CARDIAC ELECTROPHYSIOL LAB,BURLINGTON,MA 01805
来源
关键词
undersensing; defibrillation; implantable cardioverter defibrillators;
D O I
10.1111/j.1540-8159.1997.tb03878.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Evaluation of the impact of undersensing on VF detection time and the relationship of undersensing to the programmed shock energy. Background: Failure to reconfirm on ongoing arrhythmia due to undersensing by a noncommitted ICD might prolong the time to therapy. Methods: We measured initial detection times and redetection times at predischarge and at 2 and 6 months in 29 patients (22 men, mean age 60 years) with a noncommitted nonthoracotomy ICD. Telemetry data and output markers were used to analyze each induction. Results: Undersensing leading io failure to reconfirm was present in 44 (11.1%) of 398 episodes of sustained VF and prolonged significantly the median initial detection time from 2.3 seconds (25th and 75th percentiles: 2 and 2.6 s, respectively) to 5.45 seconds (4.3 and 7.35 s, P < 0.0001). One episode required external defibrillation after reconfirmation failure occurred during charging; the total detection time prior to shock was 46 seconds. In a subset of 87 episodes with failed first shocks, the initial detection time was 2.3 seconds (2.1 and 2.8 s) and the redetection time 3 seconds (2.5 and 4.7 s, P < 0.0001). The presence of undersensing prolonged the redetection from 2.6 seconds (2.35 and 3.1 s) to 5.4 seconds (4.53 and 7.35 s, P < 0.0001). Undersensing was more prevalent during the redetection period (P = 0.004) and in episodes of sustained VF in which the first shock energy was higher than 15 J (19.7% vs 5.8%, P < 0.0001). Conclusions: In this automatic defibrillator system, undersensing occurs in 11% of the sustained VF inductions and prolongs detection time significantly. Redetection is longer than initial detection mostly due to the presence of undersensing, the frequency of which is proportional to the programmed energy. The clinical significance of this finding is unknown.
引用
收藏
页码:610 / 618
页数:9
相关论文
共 50 条
  • [1] Undersensing of ventricular fibrillation by a biventricular implantable cardioverter-defibrillator: What is the cause and the troubleshooting?
    Kien Vo Le
    Okamura, Hideo
    Nakajima, Kenzaburo
    Noda, Takashi
    Kusano, Kengo
    JOURNAL OF ARRHYTHMIA, 2019, 35 (02) : 276 - 278
  • [2] Ventricular fibrillation undersensing due to air entrapment in a patient implanted with a subcutaneous cardioverter defibrillator
    Adduci, Carmen
    Spadoni, Loris
    Palano, Francesca
    Francia, Pietro
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2019, 30 (08) : 1373 - 1374
  • [3] FAILURE OF AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR TO REDETECT VENTRICULAR-FIBRILLATION IN PATIENTS WITH A NONTHORACOTOMY LEAD SYSTEM
    JUNG, W
    MANZ, M
    MOOSDORF, R
    LUDERITZ, B
    CIRCULATION, 1992, 86 (04) : 1217 - 1222
  • [4] Selective Left Ventricular Sensing Lead Implantation to Overcome Undersensing of Ventricular Fibrillation During Implantable Cardioverter Defibrillator Implantation
    Steinberg, Christian
    Philippon, Francois
    O'Hara, Gilles
    Molin, Franck
    CANADIAN JOURNAL OF CARDIOLOGY, 2013, 29 (06) : 751.e3 - 751.e6
  • [5] SUBMAMMARY IMPLANTATION OF A CARDIOVERTER-DEFIBRILLATOR WITH A NONTHORACOTOMY LEAD SYSTEM
    KOLETTIS, TM
    SAXENA, A
    KROL, RB
    SAKSENA, S
    AMERICAN HEART JOURNAL, 1993, 126 (05) : 1222 - 1223
  • [6] Nonthoracotomy cardioverter defibrillator implantation in infants
    Schreiber, C
    Eicken, A
    RESUSCITATION, 2006, 69 (02) : 350 - 350
  • [7] Sinus rhythm R-wave amplitude as a predictor of ventricular fibrillation undersensing in patients with implantable cardioverter-defibrillator
    Ruetz, Linda L.
    Koehler, Jodi L.
    Brown, Mark L.
    Jackson, Troy E.
    Belk, Paul
    Swerdlow, Charles D.
    HEART RHYTHM, 2015, 12 (12) : 2411 - 2418
  • [8] Compatibility of a nonthoracotomy lead system with a biphasic implantable cardioverter-defibrillator
    Porterfield, JG
    Porterfield, LM
    Levine, JH
    Luceri, RM
    Hsia, HH
    AbiSamra, F
    Alvarez, L
    Buxton, AE
    Camm, J
    Cutler, JE
    Fogoros, RN
    Hsia, HH
    Janosik, DL
    Kelly, P
    Levine, JH
    Luceri, RM
    Mann, D
    Meesmann, M
    McCowan, R
    Nathan, A
    Porterfield, JG
    Rothbart, ST
    Steinberg, JS
    Torres, VI
    Trappe, HJ
    Winters, SL
    AMERICAN JOURNAL OF CARDIOLOGY, 1996, 77 (08): : 586 - 590
  • [9] IMPLANTATION OF A CARDIOVERTER DEFIBRILLATOR IN THE SUBPECTORAL REGION COMBINED WITH A NONTHORACOTOMY LEAD SYSTEM
    HAMMEL, D
    BLOCK, M
    BORGGREFE, M
    KONERTZ, W
    BREITHARDT, G
    SCHELD, HH
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (04): : 367 - 368
  • [10] A prospective, randomized evaluation of a nonthoracotomy implantable cardioverter defibrillator lead system
    Baker, JH
    Epstein, AE
    VoshageStahl, L
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01): : 72 - 78