Undersensing of ventricular fibrillation in a noncommitted nonthoracotomy cardioverter defibrillator system

被引:6
|
作者
Peralta, AO [1 ]
John, RM [1 ]
Venditti, FJ [1 ]
Martin, DT [1 ]
机构
[1] LAHEY CLIN MED CTR,CARDIAC ELECTROPHYSIOL LAB,BURLINGTON,MA 01805
来源
关键词
undersensing; defibrillation; implantable cardioverter defibrillators;
D O I
10.1111/j.1540-8159.1997.tb03878.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Evaluation of the impact of undersensing on VF detection time and the relationship of undersensing to the programmed shock energy. Background: Failure to reconfirm on ongoing arrhythmia due to undersensing by a noncommitted ICD might prolong the time to therapy. Methods: We measured initial detection times and redetection times at predischarge and at 2 and 6 months in 29 patients (22 men, mean age 60 years) with a noncommitted nonthoracotomy ICD. Telemetry data and output markers were used to analyze each induction. Results: Undersensing leading io failure to reconfirm was present in 44 (11.1%) of 398 episodes of sustained VF and prolonged significantly the median initial detection time from 2.3 seconds (25th and 75th percentiles: 2 and 2.6 s, respectively) to 5.45 seconds (4.3 and 7.35 s, P < 0.0001). One episode required external defibrillation after reconfirmation failure occurred during charging; the total detection time prior to shock was 46 seconds. In a subset of 87 episodes with failed first shocks, the initial detection time was 2.3 seconds (2.1 and 2.8 s) and the redetection time 3 seconds (2.5 and 4.7 s, P < 0.0001). The presence of undersensing prolonged the redetection from 2.6 seconds (2.35 and 3.1 s) to 5.4 seconds (4.53 and 7.35 s, P < 0.0001). Undersensing was more prevalent during the redetection period (P = 0.004) and in episodes of sustained VF in which the first shock energy was higher than 15 J (19.7% vs 5.8%, P < 0.0001). Conclusions: In this automatic defibrillator system, undersensing occurs in 11% of the sustained VF inductions and prolongs detection time significantly. Redetection is longer than initial detection mostly due to the presence of undersensing, the frequency of which is proportional to the programmed energy. The clinical significance of this finding is unknown.
引用
收藏
页码:610 / 618
页数:9
相关论文
共 50 条
  • [41] IMPLANTATION OF A HYBRID NONTHORACOTOMY IMPLANTABLE DEFIBRILLATOR SYSTEM
    REDD, RM
    BISSETT, JK
    TOBLER, HG
    CLINICAL RESEARCH, 1994, 42 (03): : A362 - A362
  • [42] IMPLANTABLE CARDIOVERTER DEFIBRILLATOR IMPLANTED BY NONTHORACOTOMY APPROACH - INITIAL CLINICAL-EXPERIENCE WITH THE REDESIGNED TRANSVENOUS LEAD SYSTEM
    MOORE, SL
    MALONEY, JD
    EDEL, TB
    MCGUINN, WP
    GOHN, D
    MORANT, VA
    TROHMAN, R
    SIMMONS, TW
    WILKOFF, BL
    CASTLE, LW
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11): : 1865 - 1869
  • [43] Sinus Rhythm R-wave Amplitude Does Not Predict Undersensing of Ventricular Fibrillation by Implantable Cardioverter-Defibrillators
    Ruetz, Linda L.
    Koehler, Jodi L.
    Jackson, Troy E.
    Belk, Paul
    CIRCULATION, 2009, 120 (18) : S650 - S650
  • [44] INTRACARDIAC EMERGENCY DEFIBRILLATION FOR REFRACTORY VENTRICULAR-FIBRILLATION DURING IMPLANTATION OF CARDIOVERTER DEFIBRILLATORS WITH NONTHORACOTOMY LEAD SYSTEMS
    PITSCHNER, HF
    NEUZNER, J
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (02): : 291 - 296
  • [45] Noise detection during bradycardia pacing with a hybrid nonthoracotomy implantable cardioverter defibrillator system: Incidence and clinical significance
    Rosenthal, ME
    Paskman, C
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (07): : 1380 - 1386
  • [46] IMPLANTATION OF TRANSVENOUS NONTHORACOTOMY CARDIOVERTER-DEFIBRILLATOR SYSTEMS IN PATIENTS WITH PERMANENT ENDOCARDIAL PACEMAKERS
    BROOKS, R
    GARAN, H
    MCGOVERN, BA
    RUSKIN, JN
    AMERICAN HEART JOURNAL, 1995, 129 (01) : 45 - 53
  • [47] SUBSTANTIAL IMPROVEMENT IN REDETECTION OF VENTRICULAR-FIBRILLATION WITH A NEW NONTHORACOTOMY LEAD SYSTEM
    JUNG, W
    MANZ, M
    PFEIFFER, D
    TEBBENJOHANNS, J
    MOOSDORF, R
    KIRCHHOFF, PG
    LUDERITZ, B
    CIRCULATION, 1992, 86 (04) : 791 - 791
  • [48] Pacing induced ventricular fibrillation in internal cardioverter defibrillator patients: A new form of proarrhythmia
    Vlay, LC
    Vlay, SC
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01): : 132 - 133
  • [49] Clinical characteristics and ventricular tachyarrhythmia events in patients with an implantable cardioverter defibrillator and atrial fibrillation
    Ammar, S.
    Streitner, F.
    Schoene, N.
    Veltmann, C.
    Schimpf, R.
    Wolpert, C.
    Borggrefe, M.
    Kuschyk, J.
    EUROPEAN HEART JOURNAL, 2009, 30 : 281 - 282
  • [50] Cognitive impairment after appropriate implantable cardioverter-defibrillator therapy for ventricular fibrillation
    Halas, Katarzyna
    Krzyzanowski, Krystian
    Krzyzanowska, Ewa
    Smurzynski, Pawel
    Ryczek, Robert
    Michalkiewicz, Dariusz
    Orski, Zbigniew
    Makowski, Karol
    Wierzbowski, Robert
    Gielerak, Grzegorz
    KARDIOLOGIA POLSKA, 2014, 72 (02) : 134 - 139