FAILURE OF AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR TO REDETECT VENTRICULAR-FIBRILLATION IN PATIENTS WITH A NONTHORACOTOMY LEAD SYSTEM

被引:56
|
作者
JUNG, W [1 ]
MANZ, M [1 ]
MOOSDORF, R [1 ]
LUDERITZ, B [1 ]
机构
[1] UNIV BONN,DEPT CARDIOVASC SURG,W-5300 BONN 1,GERMANY
关键词
IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR; LEAD SYSTEMS; NONTHORACOTOMY; SUDDEN CARDIAC DEATH; ELECTROGRAM; ENDOCARDIAL;
D O I
10.1161/01.CIR.86.4.1217
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Shock delivery of an implantable defibrillator may cause a change in the amplitude of endocardial electrograms and impair the detection of ventricular fibrillation. Thus, the effects of shock discharges on the amplitude of endocardial electrograms were evaluated in five patients undergoing implantation of a cardioverter-defibrillator in combination with a new nonthoracotomy lead system. Methods and Results. At implant, bipolar endocardial electrograms were recorded before each shock application, during ventricular fibrillation, during redetection of ventricular fibrillation in case the applied shock was ineffective, and at intervals of 5, 10, 20, 30, 60, and 120 seconds after each shock delivery. The amplitude of the endocardial electrograms decreased from 10.5+/-3.8 mV during sinus rhythm to 6.3+/-1.9 mV during initial ventricular fibrillation and declined to 2.2+/-1.3 mV during redetection of ventricular fibrillation. After successful termination, the following bipolar electrograms could be obtained at the predetermined intervals: 1.9+/-1.2 mV, 3.1+/-1.8 mV, 4.5+/-1.9 mV, 6.5+/-2.9 mV, 9.5+/-3.3 mV, and 10.4+/-3.8 mV. At predischarge testing, failure of redetection of ventricular fibrillation could be documented in two patients, requiring rescue external defibrillation in both cases to restore sinus rhythm. Conclusions. These findings demonstrate that the implantable cardioverter-defibrillator did not ensure reliable redetection of ventricular fibrillation in patients using the implanted nonthoracotomy lead system. Thus, the potential risk of sudden cardiac death may persist in these patients despite defibrillator therapy.
引用
收藏
页码:1217 / 1222
页数:6
相关论文
共 50 条
  • [1] Compatibility of a nonthoracotomy lead system with a biphasic implantable cardioverter-defibrillator
    Porterfield, JG
    Porterfield, LM
    Levine, JH
    Luceri, RM
    Hsia, HH
    AbiSamra, F
    Alvarez, L
    Buxton, AE
    Camm, J
    Cutler, JE
    Fogoros, RN
    Hsia, HH
    Janosik, DL
    Kelly, P
    Levine, JH
    Luceri, RM
    Mann, D
    Meesmann, M
    McCowan, R
    Nathan, A
    Porterfield, JG
    Rothbart, ST
    Steinberg, JS
    Torres, VI
    Trappe, HJ
    Winters, SL
    AMERICAN JOURNAL OF CARDIOLOGY, 1996, 77 (08): : 586 - 590
  • [2] Failure to detect ventricular fibrillation in a patient with an implantable cardioverter-defibrillator
    Antolinos-Perez, Maria J.
    Garcia-Alberola, Arcadio
    Valdes-Chavarri, Mariano
    REVISTA ESPANOLA DE CARDIOLOGIA, 2006, 59 (05): : 517 - 518
  • [3] SUBMAMMARY IMPLANTATION OF A CARDIOVERTER-DEFIBRILLATOR WITH A NONTHORACOTOMY LEAD SYSTEM
    KOLETTIS, TM
    SAXENA, A
    KROL, RB
    SAKSENA, S
    AMERICAN HEART JOURNAL, 1993, 126 (05) : 1222 - 1223
  • [4] A PERMANENT TRANSVENOUS LEAD SYSTEM FOR AN IMPLANTABLE PACEMAKER CARDIOVERTER-DEFIBRILLATOR - NONTHORACOTOMY APPROACH TO IMPLANTATION
    YEE, R
    KLEIN, GJ
    LEITCH, JW
    GUIRAUDON, GM
    GUIRAUDON, CM
    JONES, DL
    NORRIS, C
    CIRCULATION, 1992, 85 (01) : 196 - 204
  • [5] Failure to detect ventricular fibrillation in a patient with an implantable cardioverter-defibrillator [3]
    Antolinos-Pérez, María J.
    García-Alberola, Arcadio
    Redondo-Bermejo, Belén
    Valdés-Chávarri, Mariano
    Technische Uberwachung, 2006, 47 (06): : 517 - 518
  • [6] Pacemaker/implantable cardioverter-defibrillator problem: Right ventricular lead missensing in a biventricular implantable cardioverter-defibrillator
    Otten, Richard F.
    Foreman, Lynne D.
    Groh, William J.
    HEART RHYTHM, 2008, 5 (01) : 158 - 159
  • [7] Prediction of atrial fibrillation in patients with an implantable cardioverter-defibrillator and heart failure
    Bertini, Matteo
    Borleffs, C. Jan Willem
    Delgado, Victoria
    Ng, Arnold C. T.
    Piers, Sebastiaan R. D.
    Shanks, Miriam
    Antoni, M. Louisa
    Biffi, Mauro
    Boriani, Giuseppe
    Schalij, Martin J.
    Bax, Jeroen J.
    Van de Veire, Nico R. L.
    EUROPEAN JOURNAL OF HEART FAILURE, 2010, 12 (10) : 1101 - 1110
  • [8] Management of the Patient With Implantable Cardioverter-Defibrillator Lead Failure
    Kalahasty, Gautham
    Ellenbogen, Kenneth A.
    CIRCULATION, 2011, 123 (12) : 1352 - 1354
  • [9] Lead failure in an entirely subcutaneous implantable cardioverter-defibrillator
    Kolb, Christof
    Weigand, Severin
    Schaarschmidt, Claudia
    Lennerz, Carsten
    EUROPACE, 2020, 22 (02): : 183 - 183
  • [10] An algorithm to predict implantable cardioverter-defibrillator lead failure
    Gunderson, BD
    Patel, AS
    Bounds, CA
    Shepard, RK
    Wood, MA
    Ellenbogen, KA
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 44 (09) : 1898 - 1902