Advanced Design of Birdcage RF Coil for Various Absorption Regions at 3T MRI System

被引:0
|
作者
Lee, Jung-Woo [1 ]
Choe, Bo-Young [1 ]
Yoon, Sung-Ik [1 ]
Choi, Chi-Bong [1 ]
Huh, Soon-Nyoung [2 ]
机构
[1] Catholic Univ Korea, Kangnam St Marys Hosp, Coll Med, Dept Biomed Engn, 505 Banpo Dong, Seoul 137040, South Korea
[2] Seoul Natl Univ, Coll Med, Dept Radiat Oncol, Seoul 110744, South Korea
来源
关键词
Birdcage Resonator; 3; T; Signal-to-Noise Ratio (SNR);
D O I
暂无
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Purpose: The purpose of this study was to design and build an optimized birdcage resonator configuration with a low pass filter, which would facilitate the acquisition of high-resolution 3D-images of small animals at 3T MRI system. Methods and Materials: The birdcage resonator with 12-element structures was built, in order to ensure B1 homogeneity over the image volume and maximum filling factor, and hence to maximize the signal to noise ratio (SNR) and resolution of the 3-dimensional images. The diameter and length of each element of a birdcage resonator were as follows: (1) diameter 13 cm, length 22 cm, (2) diameter 15 cm, length 22 cm, (3) diameter 17 cm, length 25 cm. Spin echo pulse sequence and fast spin echo pulse sequence were employed in obtaining MR images. The quality of the manufactured birdcage resonators was evaluated on the basis of the return loss following matching and tuning process. Results: The experimental MR images of phantoms by the various manufactured birdcage resonators were obtained to compare the SNR in accordance with the size of objects. The size of an object to that of coil was identified by parameters that were estimated from the images of a phantom. First, the diameter of the birdcage resonator was 15cm, and the ratio of the tangerine to the birdcage resonator accounted for approximately 27%. The Q factor was 53.2 and the SNR was 150.7. Second, at the same birdcage resonator, the ratio of the orange was approximately 53%. The SNR and the Q parameter was 212.8 and 91.2, respectively. Conclusion: The present study demonstrated that if birdcage resonators have the same forms, SNR could be different depending on the size of an object, especially when the size of an object to that of coil is approximately 40 similar to 80%, the former is bigger than the latter. Therefore, when the size of an object to be observed is smaller than that of coil, the coil should be manufactured in accordance with the size of an object in order to obtain much more excellent images.
引用
收藏
页码:48 / 60
页数:13
相关论文
共 50 条
  • [1] Numerical Design of RF Birdcage Coil without Lumped Elements for MRI 3T System
    Basari
    Priatna, Arwan
    Rahardjo, Eko Tjipto
    Zulkifli, Fitri Yuli
    2015 INTERNATIONAL CONFERENCE QUALITY IN RESEARCH (QIR), 2015, : 90 - 93
  • [2] Impacts of RF Shimming on MRI Induced Heating of Implantable Medical Lead in 3T Birdcage Coil
    Zeng, Qi
    Wang, Qingyan
    Zheng, Jianfeng
    Kainz, Wolfgang
    Chen, Ji
    2017 IEEE MTT-S INTERNATIONAL MICROWAVE SYMPOSIUM (IMS), 2017, : 1235 - 1237
  • [3] A Passive and Conformal Magnetic Metasurface for 3T MRI Birdcage Coil
    Rotundo, Sabrina
    Brizi, Danilo
    Carluccio, Giuseppe
    Lakshmanan, Karthik
    Collins, Christopher M.
    Lattanzi, Riccardo
    Monorchio, Agostino
    2023 INTERNATIONAL CONFERENCE ON ELECTROMAGNETICS IN ADVANCED APPLICATIONS, ICEAA, 2023, : 222 - 224
  • [4] Birdcage Type NMR Receiver Coil Sensor with Integrated Detuning Circuit for 3T MRI System
    Ahmad, Sheikh Faisal
    Kim, Young Cheol
    Choi, Ick Chang
    Kim, Hyun Deok
    2015 IEEE SENSORS, 2015, : 1777 - 1780
  • [5] Impacts of RF Shimming on Local SAR Caused by MRI 3T Birdcage Coil near Femoral Plate Implants
    Zeng, Qi
    Guo, Ran
    Zheng, Jianfeng
    Chen, Ji
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1005 - 1006
  • [6] Simulation Comparison of Birdcage Coil and Metamaterial Liner for MRI at 3T and 4.7T
    Maunder, Adam
    De Zanche, Nicola
    Iyer, Ashwin K.
    2020 50TH EUROPEAN MICROWAVE CONFERENCE (EUMC), 2020,
  • [7] Simulation Comparison of Birdcage Coil and Metamaterial Liner for MRI at 3T and 4.7T
    Maunder, Adam
    De Zanche, Nicola
    Iyer, Ashwin K.
    2020 50TH EUROPEAN MICROWAVE CONFERENCE (EUMC), 2020, : 1067 - 1070
  • [8] Simulation Comparison of Birdcage Coil and Metamaterial Liner for MRI at 3T and 4.7T
    Maunder, Adam
    De Zanche, Nicola
    Iyer, Ashwin K.
    2020 50TH EUROPEAN MICROWAVE CONFERENCE (EUMC), 2020,
  • [9] Numerical Design of Dual Resonant Phased Array RF Coil for MRI 3T and 7T System
    Basari
    Yuliyanti, Sri
    Rahardjo, Eko Tjipto
    Zulkifli, Fitri Yuli
    2015 2ND INTERNATIONAL CONFERENCE ON INFORMATION TECHNOLOGY, COMPUTER, AND ELECTRICAL ENGINEERING (ICITACEE), 2015, : 275 - 278
  • [10] Recent Progress in Birdcage RF Coil Technology for MRI System
    Ahmad, Sheikh Faisal
    Kim, Young Cheol
    Choi, Ick Chang
    Kim, Hyun Deok
    DIAGNOSTICS, 2020, 10 (12)