Dose reduction from loss of backscatter in superficial x-ray radiation therapy with the Pantak SXT 150 unit

被引:0
|
作者
B. J. Healy
S. Sylvander
K. N. Nitschke
机构
[1] Princess Alexandra Hospital,Southern Zone Radiation Oncology Service
[2] Radiation Oncology Queensland,Mater Centre
关键词
superficial x-ray therapy; backscatter; lead shielding;
D O I
暂无
中图分类号
学科分类号
摘要
Loss of backscatter has been measured for a Pantak Therapax SXT 150 superficial x-ray therapy unit. Ionisation chamber measurements at the surface were performed in solid water with and without underlying lead to confirm the applicability of previous published measurements. Measurements of loss of backscatter with and without lead allowed the effect of underlying lead to be separated from the effect of loss of backscatter. It was found that the difference between the effect of lead and that of air on surface dose is minimal for solid water phantom thickness exceeding 3 cm. Differences up to 3% in dose reduction at the surface were found between this work and previous published results with different x-ray units.
引用
收藏
页码:49 / 55
页数:6
相关论文
共 50 条
  • [1] Dose reduction from loss of backscatter in superficial x-ray radiation therapy with the Pantak SXT 150 unit
    Healy, B. J.
    Sylvander, S.
    Nitschke, K. N.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2008, 31 (01) : 49 - 55
  • [2] A model-based algorithm to correct for the loss of backscatter in superficial X-ray radiation therapy
    Harwood, Jeffrey R.
    Nelli, Flavio E.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 65 : 157 - 166
  • [3] Surface Dose Reduction from Bone Interface in Superficial X-Ray Radiation Therapy: A Monte Carlo Study
    Chow, J.
    Owrangi, A.
    Grigorov, G.
    MEDICAL PHYSICS, 2011, 38 (06) : 3649 - +
  • [4] Stray radiation dose reduction using a backscatter shield and a rectangular collimator for a handheld intraoral X-ray unit
    Otaka, Yusei
    Harata, Yasuo
    Izawa, Maki
    Shiba, Noriyoshi
    Iwawaki, Atsushi
    Ishii, Takeru
    Ochiai, Akihiro
    Takahashi, Nobutoshi
    Saka, Hideki
    Kito, Shinji
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (19) : 2349 - 2355
  • [5] NEW SUPERFICIAL X-RAY THERAPY UNIT
    LILLICRAP, SC
    WILSON, BC
    EDSER, E
    STANLEY, D
    BRITISH JOURNAL OF RADIOLOGY, 1974, 47 (555): : 193 - 195
  • [6] Performance characteristics of the Pantak Therapax-150 superficial X-ray treatment machine: Measurements and calculations.
    Natto S.S.A.
    Australasian Physical and Engineering Sciences in Medicine, 2002, 25 (04): : 162 - 167
  • [7] Experimental measurements and Monte Carlo modelling of the XSTRAHL 150 superficial X-ray therapy unit
    Al-Ghorabie, Fayez
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2015, 14 (01) : 43 - 55
  • [8] A NEW TECHNIQUE FOR RADIATION SHIELDING IN SUPERFICIAL X-RAY THERAPY
    BAILY, B
    COE, MA
    HEARNDEN, TM
    BRITISH JOURNAL OF RADIOLOGY, 1981, 54 (645): : 805 - 807
  • [9] COMPARISON OF F FACTORS DERIVED FROM HVL AND X-RAY-SPECTRUM MEASUREMENTS FOR A SUPERFICIAL THERAPY X-RAY UNIT
    LEVY, L
    WAGGENER, R
    ZANCA, P
    MCDAVID, D
    PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (06): : 883 - &
  • [10] Measurement of depth dose curves for radiation of RTA X-ray unit
    Vasil'ev, V.N.
    Kokontsev, A.A.
    Meditsinskaya Tekhnika, 1997, (05): : 11 - 14